디지털 방사선에서 격자 비에 따른 영상변화에 관한 연구

A Study on Image Change according to Grid Ratio in Digital Radiography

1

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류방사선과학

초록

With the introduction of the D.R system, the grid with high removal rate of scattered radiation is selected and used without considering the grid performance evaluation. Despite the high removal rate of scattered radiation belonging to high grid ratio, it is also possible to see the cut-off phenomenon in which the primary ray involved in the formation of image could be removed as well. Thus, the purpose of this study is to provide the basic data for researches on the usefulness of using the grid by presenting the appropriate grid ratio in the D.R system through the measurement of image in grid such as non-grid, 6:1, 8:1, 10:1, 12:1. The equipments used for this study were radiation generator with grid in 12:1, 10:1, 8:1, 6:1, indirect-type detector, and acryl phantom. As the study for image evaluation, this study measured the SNR, PSNR, MSE, and Entropy. In the results of this study, the PSNR was the highest in 6:1 and the lowest in 8:1. The SNR was high in 6:1 and 8:1, and the lowest in 12:1. In case of Entropy, it was high in 8:1 and 10:1, and the lowest in 12:1. Therefore, when the grid is used, it would be more proper to choose the grid in 8:1 or 10:1 with less loss in information content of primary ray rather than the high grid ratio showing the increased patient exposure dose.

Abstract

With the introduction of the D.R system, the grid with high removal rate of scattered radiation is selected and used without considering the grid performance evaluation. Despite the high removal rate of scattered radiation belonging to high grid ratio, it is also possible to see the cut-off phenomenon in which the primary ray involved in the formation of image could be removed as well. Thus, the purpose of this study is to provide the basic data for researches on the usefulness of using the grid by presenting the appropriate grid ratio in the D.R system through the measurement of image in grid such as non-grid, 6:1, 8:1, 10:1, 12:1. The equipments used for this study were radiation generator with grid in 12:1, 10:1, 8:1, 6:1, indirect-type detector, and acryl phantom. As the study for image evaluation, this study measured the SNR, PSNR, MSE, and Entropy. In the results of this study, the PSNR was the highest in 6:1 and the lowest in 8:1. The SNR was high in 6:1 and 8:1, and the lowest in 12:1. In case of Entropy, it was high in 8:1 and 10:1, and the lowest in 12:1. Therefore, when the grid is used, it would be more proper to choose the grid in 8:1 or 10:1 with less loss in information content of primary ray rather than the high grid ratio showing the increased patient exposure dose.

참고문헌 (17)

  1. Seo WJ, "Using Image J program, compared of focusing distance and grid rate", Korean Journal Digital Imaging Med, 14(1), pp.10~15 (2012)
  2. 김동식, "고정 그리드를 사용한 디지털 방사선 영상에서 1차원 필터에 기초한 왜곡의 제거", 전자공학회논문지 - SP, 47(5), pp.117~126 (2010), 대한전자공학회
  3. Lee SJ, "Study on a moirm Artifact in the Use of Carbon Interspaced Antiscatter Grids for Digital Radiography", Journal of the Korean Society of Radiology, 2(4), pp.5~7 (2008)
  4. Seo JH, "Investigation of Scattered Radiation on Images Effect in Digital Radiograph", Journal of the Korean Society of Radiology. Spring Conference Discussion, 8(1), pp.77~79 (2014)
  5. Ko IH, "An experimental study on performance evaluation and use of diagnostic X-ray Grid being used in che-ju province", Korea Radiological Association, 25(1), pp.192~195 (1999)
  6. Kim DS, "Grid Angle Optimization for Grid Artifact Reduction in Digital Radiography", Korean Society of Broadcasting Engineers, 2010(7), pp.439~440 (2010)
  7. 이상호, "Moire artifact 제거 후 그리드 주파수에 따른 미세병변의 영상분석", 방사선기술과학, 41(5), pp.463~469 (2018), 대한방사선과학회[DOI]
  8. Choi SH., "A Study on Dose and Image Quality in D.R System Evaluation According to Use of the Grid", Department of Radiological Science, pp.2012
  9. Kwon DM, "Analog & Digital·PACS medical imaging informatics", pp.60~69 (2021), Daihak Seorim
  10. Nam TH, "Medical imaging practice using matlab", pp.108~112 (2013), Wolsong
  11. Choi SH., "A Study on Dose and Image Quality in DR System Evaluation According to Use of the Grid", Department of Radiological Science, pp.2012
  12. Go SJ, "Introduction to Radiologic Science", 110~111 (2016)
  13. 백철하, "디지털 방사선영상 시스템에서 산란선이 영상 품질에 미치는 영향", 방사선기술과학, 40(1), pp.71~77 (2017), 대한방사선과학회[DOI]
  14. 이기호, "Computed Radiography에서 고정형 그리드와 이동형 그리드 영상의 인식률 비교", 한국방사선학회논문지, 9(7), pp.515~521 (2015), 한국방사선학회[DOI]
  15. Seo WJ, "Using Image J program, compared of focusing distance and grid rate", Journal of the Korean Society for Digital Imaging in Medicine, 14(1), pp.37~39 (2012)
  16. 이인자, "FPD System에서 상․하지 촬영 시 격자에 따른 환자 선량 및 화질 평가", 방사선기술과학, 33(4), pp.341~348 (2010), 대한방사선과학회
  17. Jeung H, "Digital Medical Imaging", pp.174~175 (2003), Jeongmungak

관련 논문AI추천

이 간행물의 다른 논문 (7건)