관전압에 따른 적정 격자비에 관한 실험
Experiment with Respect to Proper Grid Ratio According to Tube Voltage
0
조회
0
다운로드
3
인용
0
참고문헌
키워드
초록
The physical properties of the grid, including the CNR and SNR of the image, were measuredin order to properly select the grid ratio, which affects the contrast and quality of an imageand greatly affects the exposure dose of the patient. The following are the conclusions arrived ataccording to the results of the study. The contrast improvement factor (K) increased as the gridratio increased, and the increased amount was highest at 8 : 1. The tube voltage increased up to 80kVp which resulted in maximum contrast improvement factor. Further increase of tube voltagetended to decrease the contrast improvement factor. The selectivity (Σ) increased as the grid ratioincreased, but such tendency slowed down when the ratio was 12 : 1. The selectivity increased asthe tube voltage increased, with the largest volume recorded at 80 kVp, whereas selectivity ratherdecreased when the tube voltage increased further. Bucky factor (B) continued to increase as thegrid ratio increased; however, such increase tendency slowed down from 80 kVp. The exposurefactor decreased as the tube voltage increased. SNR decreased as the grid ratio increased, and itbecame largest at the 6 : 1 grid. It was shown that SNR decreased as the tube voltage increased. CNR increased up to 10 : 1 grid as the grid ratio increased, but it decreased at 12 : 1 grid. It wasshown that there was no effect to the CNR as the tube voltage increased. According to the result ofthe physical evaluation on the conventional theory that the selection of grid depends on tube voltage,the grid ratio was suitable for 8 : 1 or 10 : 1 at all tube voltages. However, it is necessary to carefullydetermine the patients’ exposure dose. For SNR, 6 : 1 is good for all tube voltages but for CNR,10 : 1 is considered good for all tube voltages. Therefore, it is recommended to select a grid in considerationof the tube voltage used and exposure dose of the patients.
Abstract
The physical properties of the grid, including the CNR and SNR of the image, were measuredin order to properly select the grid ratio, which affects the contrast and quality of an imageand greatly affects the exposure dose of the patient. The following are the conclusions arrived ataccording to the results of the study. The contrast improvement factor (K) increased as the gridratio increased, and the increased amount was highest at 8 : 1. The tube voltage increased up to 80kVp which resulted in maximum contrast improvement factor. Further increase of tube voltagetended to decrease the contrast improvement factor. The selectivity (Σ) increased as the grid ratioincreased, but such tendency slowed down when the ratio was 12 : 1. The selectivity increased asthe tube voltage increased, with the largest volume recorded at 80 kVp, whereas selectivity ratherdecreased when the tube voltage increased further. Bucky factor (B) continued to increase as thegrid ratio increased; however, such increase tendency slowed down from 80 kVp. The exposurefactor decreased as the tube voltage increased. SNR decreased as the grid ratio increased, and itbecame largest at the 6 : 1 grid. It was shown that SNR decreased as the tube voltage increased. CNR increased up to 10 : 1 grid as the grid ratio increased, but it decreased at 12 : 1 grid. It wasshown that there was no effect to the CNR as the tube voltage increased. According to the result ofthe physical evaluation on the conventional theory that the selection of grid depends on tube voltage,the grid ratio was suitable for 8 : 1 or 10 : 1 at all tube voltages. However, it is necessary to carefullydetermine the patients’ exposure dose. For SNR, 6 : 1 is good for all tube voltages but for CNR,10 : 1 is considered good for all tube voltages. Therefore, it is recommended to select a grid in considerationof the tube voltage used and exposure dose of the patients.
참고문헌 (13)
- Stewart CB, "Radiologic Science for Technologists", pp.290~295 (2004), Mosby
- 백철하, "디지털 방사선영상 시스템에서 산란선이 영상 품질에 미치는 영향", 방사선기술과학, 40(1), pp.71~77 (2017), 대한방사선과학회[DOI]
- 정명진, "우리나라의 흉부엑스선검사에서의 환자선량 권고량", 대한영상의학회지, 62(6), pp.523~528 (2010), 대한영상의학회
- 정성훈, "소아 흉부검사 시 격자 사용에 따른 영상 화질 평가", 방사선기술과학, 40(3), pp.371~376 (2017), 대한방사선과학회[DOI]
- Technical Information for Pediatric radiography, "Food and Drug Safety Evaluation Center", NiFDS, 17~27 (2010)
- John MB, "Grid and slot scan scatter reduction in mammography : comparison by using Monte Carlo techniques", Radiology, 222(2), pp.519~527 (2002)
- 김정수, "영상의학검사 일반촬영 분야의 촬영기법에 대한 분석", 방사선기술과학, 37(3), pp.169~175 (2014), 대한방사선과학회
- Kim YH, "Methods for measurement of entrance surface dose and roles of radiation technologist", J. Radiol. Sci. Technol, 28(3), pp.173~191 (2005)
- 이후민, "X선 진단장치의 고압정류방식이 대조도 향상에 미치는 영향", 방사선기술과학, 37(3), pp.187~193 (2014), 대한방사선과학회
- 이인자, "FPD System에서 상․하지 촬영 시 격자에 따른 환자 선량 및 화질 평가", 방사선기술과학, 33(4), pp.341~348 (2010), 대한방사선과학회
- Richard RC, "Principles of Radiographic Imaging", pp.255~269 (2006), Delmar
- Suh JH, "Investigation of scattered radiation on images effect in digital radiography", J. Korean Soc. Radiol, 8(1), pp.77~79 (2014)
- 유영은, "디지털 무선 검출기를 이용한 이동형 X선검사에서 영상품질 저하의 요인분석", 방사선기술과학, 37(2), pp.93~100 (2014), 대한방사선과학회