본문 바로가기
KERD
논문KCI 등재

광학 전산화단층촬영 스캐너 사용을 위한 중합체 겔과 BANG-TM 겔 선량계의 특성 비교

Comparison of Polymer and BANG-TM Gel Dosimeters to Use a Optical Computed Tomography Scanner

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

The purpose of this study was to compare the basic radiological characteristics of in-house polymer gel and commercially-available gel (BANGTM) gel dosimeters with a spectro-photometer to use a optical computed tomography (CT) scanner. We investigated the radiological characteristics including dose linearity, absorbance spectrum, dose rate dependency and inter-and intra-reproducibility at wavelengths of 590, 600 and 630 nm. The optimal dose linearities of two gel dosimeters showed R2 value of 0.939 and 0.948 at wavelengths of 590 nm and 600 nm, respectively. For two polymer gel dosimeters, there is no peak sensitivity within the range of all wavelengths in absorbance spectrum. For in-house gel dosimeter, the dose rate dependency were within 5% for all wavelengths except for the dose rate of 100 MU/min. For BANG-TM gel dosimeter, the dose rate dependency showed an error range of ±5% for all wavelengths. The inter-and intra-reproducibility of two gel dosimeters were within the range of 2.5%. We have confirmed that the two gel dosimeters was appropriate for use with a optical CT scanner.

Abstract

The purpose of this study was to compare the basic radiological characteristics of in-house polymer gel and commercially-available gel (BANGTM) gel dosimeters with a spectro-photometer to use a optical computed tomography (CT) scanner. We investigated the radiological characteristics including dose linearity, absorbance spectrum, dose rate dependency and inter-and intra-reproducibility at wavelengths of 590, 600 and 630 nm. The optimal dose linearities of two gel dosimeters showed R2 value of 0.939 and 0.948 at wavelengths of 590 nm and 600 nm, respectively. For two polymer gel dosimeters, there is no peak sensitivity within the range of all wavelengths in absorbance spectrum. For in-house gel dosimeter, the dose rate dependency were within 5% for all wavelengths except for the dose rate of 100 MU/min. For BANG-TM gel dosimeter, the dose rate dependency showed an error range of ±5% for all wavelengths. The inter-and intra-reproducibility of two gel dosimeters were within the range of 2.5%. We have confirmed that the two gel dosimeters was appropriate for use with a optical CT scanner.

참고문헌 (25)

  1. Miften M, "Tolerance limits and methodologies for IMRT measurement-based verification QA : Recommendations of AAPM Task Group No. 218", Med Phys, 45(4), pp.e53~e83 (2018)[DOI]
  2. Intensity Modulated Radiation Therapy Collaborative Working Group, "Intensity-modulated radiotherapy : Current status and issues of interest", Int J Radiat Oncol Biol Phys, 51(4), pp.880~914 (2001)
  3. 조병철, "Intensity-modulated radiation therapy: a review with a physics perspective", Radiation oncology journal, 36(1), pp.1~10 (2018), 대한방사선종양학회
  4. Babic S, "Three-dimensional dose verification for intensity-modulated radiation therapy in the radiological physics centre headand-neck phantom using optical computed tomography scans of ferrous xylenol-orange gel dosimeters", Int J Radiat Oncol Biol Phys, 70(4), pp.1281~1291 (2008)
  5. Oldham M, "3D dosimetry by optical-CT scanning", J Phys Conf Ser, 56, pp.58~71 (2006)
  6. Chang KH, "Development of a 3D optical scanner for evaluating patient-specific dose distributions", Physica Medica, 31(5), pp.553~559 (2015)
  7. Mein S, "Development of a 3D remote dosimetry protocol compatible with MRgIMRT", Med Phys, 44(11), pp.6018~6028 (2017)
  8. Soliman YS, "Fricke gel dosimeter as a tool in quality assurance of the radiotherapy treatment plans", Appl Radiat Isot, 120, pp.126~132 (2017)
  9. De Deene Y, "Validation of MR-based polymer gel dosimetry as a preclinical three-dimensional verification tool in conformal radiotherapy", Magn Reson Med, 43(1), pp.116~125 (2000)
  10. Guo PY, "Characterization of a new radiochromic three-dimensional dosimeter", Med Phys, 33(5), pp.1338~1345 (2006)
  11. Kyung Hwan Chang, "Basic Radiological Characteristics of a Non-Scattering Gel Dosimeter for 3D Dosimetry", Journal of the Korean Physical Society, 69(11), pp.1694~1699 (2016), 한국물리학회
  12. Valente M, "Fricke gel dosimeter with improved sensitivity for low-dose-level measurements", J Appl Clin Med Phys, 17(4), pp.402~417 (2016)
  13. De Deene Y, "A basic study of some normoxic polymer gel dosimeters", Phys Med Biol, 47(19), pp.3441~3463 (2002)
  14. Haraldsson P, "Dose response evaluation of a low-density normoxic polymer gel dosimeter using MRI", Phys Med Biol, 51(4), pp.919~928 (2006)
  15. Jirasek A, "Polymer gel dosimeters with enhanced sensitivity for use in x-ray CT polymer gel dosimetry", Phys Med Biol, 55(18), pp.5269~5281 (2010)
  16. Jirasek AI, "Effects of crosslinker fraction in polymer gel dosimeters using FT Raman spectroscopy", Phys Med Biol, 46(7), pp.1949~1961 (2001)
  17. 이석, "Analysis of the Dosimetric Characteristics of Normoxic Polymer Gel by Magnetic Resonance Images", Journal of the Korean Physical Society, 56(3), pp.874~879 (2010), 한국물리학회
  18. Murakami Y, "Evaluation of the basic properties of the BANGkit gel dosimeter", Phys Med Biol, 52(8), pp.2301~2311 (2007)
  19. Sakhalkar HS, "A comprehensive evaluation of the PRESAGE", optical-CT 3D dosimetry system, 2009, pp.36(1), 71~82
  20. Campbell WG, "A prototype fan-beam optical CT scanner for 3D dosimetry", Med Phys, 40(6), pp.061712~ (2013)
  21. Papadakis AE, "A new optical-CT apparatus for 3-D radiotherapy dosimetry : Is free space scanning feasible?", IEEE Trans Med Imaging, 29(5), pp.1204~1212 (2010)
  22. Kumar DS, "Investigation on Tissue Equivalent Normoxic Polymer Gel Dosimeter using In-house Laser CT scanning system", J. Phys. Conf. Ser, 250, pp.012~40 (2010)
  23. 장경환, "Spectrophotometric Determination of the Optimal Wavelength for a Polymer-gel Dosimeter", Journal of the Korean Physical Society, 62(8), pp.1194~1198 (2013), 한국물리학회
  24. Bero MA, "Radiochromic gel dosemeter for three-dimensional dosimetry", Radiation Physics and Chemistry, 61(3), pp.433~435 (2001)
  25. Solc J, "New radiochromic gel for 3D dosimetry based on Turnbull blue : Basic properties", Phys Med Biol, 54(17), pp.5095~5107 (2009)

관련 논문AI추천

1차원 광섬유 팬텀선량계를 이용한 치료용 광자선의 피부 및 선량보강영역에서 상대선량 측정문진수, 박장연, 서정기, 유욱재, 이봉수, 장경원, 조영호한국센서학회 · 센서학회지2011 · 제20권 제2호GafChromic RTQA Film을 이용한 광역학적 치료용 레이저의 선질 측정이병구, 임현수, 네클라 커나르대한의용생체공학회 · 의공학회지2013 · 제34권 제2호몬테카를로 시뮬레이션을 이용한 양성자 조사에 따른 Polymer Gel 내부의 선량 분포 특성 평가김기섭, 김정민, 김현지, 박민석, 박세영, 윤용수, 정해조, 최인석대한방사선과학회 · Journal of Radiological Science and Technology2013 · 제36권 제2호액체 섬광체를 이용한 100MeV 양성자 빔의 선량 분포 평가김성환대한방사선과학회 · Journal of Radiological Science and Technology2017 · 제40권 제4호고 에너지 X-선 조사에 의한 광섬유 방사선량계와 이온 전리함의 심부선량 백분율 측정 및 비교문주현, 박병기, 서정기, 유욱재, 이봉수, 장경원, 조동현, 조영호, 허지연한국센서학회 · 센서학회지2009 · 제18권 제2호Investigation of radiological properties and water equivalence of polymer gel dosimeters in the energy range 0.1–10 keVBüyükyıldız Mehmet, Kaur Parminder, Levet Aytaç, Parsaei Sara, Rafiei Mustafa Mohammad, Thakur Sonika한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제6호

이 간행물의 다른 논문 (7건)