CT실에서 무연보드 차폐체의 X선 차폐능력 평가

Evaluation of the X-ray Shielding Ability of Lead Free Board Shielding in the CT Room

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

This study compared the X-ray shielding abilities of the shields using Computed Radiography(CR) System after manufacturing a lead-free boards using gypsum and BaSO4, an eco-friendly X-ray shielding material. Total six lead-free boards were manufactured with BaSO4 concentrations of 25%, 50% and thickness of 10mm, 15mm, 20mm respectively, and additional thickness of 1.0 mm, 1.5 mm, 2.0 mm leads were prepared. In the experiment, Nine shields were placed on the Image Plate and placed in a Computed Tomography(CT) Room where CT scans were performed for 2 weeks. After that, the X-ray image of the shields were obtained through CR Reader, and Pixel Value(PV) were measured to evaluated the X-ray shielding abilities of the lead-free shields. The criterion for evaluating the shields was determined by comparing PV of lead-free board to that of the 1.5 mm thickness lead used in the CT rooms. As a result of the experiment, the PV of the lead-free boards within 25 % of the BaSO4 concentration and within 10 mm of the thickness were not enough to be used as X-ray shields in the CT Room because they did not reach the PV of the 1.5 mm thickness lead. BaSO4 concentration of 50 % at 20 mm thickness showed PV of 1.5 mm lead thickness or more indicating that it has an X-ray shielding ability to replace lead in the CT room

Abstract

This study compared the X-ray shielding abilities of the shields using Computed Radiography(CR) System after manufacturing a lead-free boards using gypsum and BaSO4, an eco-friendly X-ray shielding material. Total six lead-free boards were manufactured with BaSO4 concentrations of 25%, 50% and thickness of 10mm, 15mm, 20mm respectively, and additional thickness of 1.0 mm, 1.5 mm, 2.0 mm leads were prepared. In the experiment, Nine shields were placed on the Image Plate and placed in a Computed Tomography(CT) Room where CT scans were performed for 2 weeks. After that, the X-ray image of the shields were obtained through CR Reader, and Pixel Value(PV) were measured to evaluated the X-ray shielding abilities of the lead-free shields. The criterion for evaluating the shields was determined by comparing PV of lead-free board to that of the 1.5 mm thickness lead used in the CT rooms. As a result of the experiment, the PV of the lead-free boards within 25 % of the BaSO4 concentration and within 10 mm of the thickness were not enough to be used as X-ray shields in the CT Room because they did not reach the PV of the 1.5 mm thickness lead. BaSO4 concentration of 50 % at 20 mm thickness showed PV of 1.5 mm lead thickness or more indicating that it has an X-ray shielding ability to replace lead in the CT room

참고문헌 (19)

  1. Valentin J, "Radiation and your patient : A guide for medical practitioners", Annals of the ICRP (2001)[DOI]
  2. Kim HC, "Evaluation and management of lead exposure", Annals of Occupational and Environmental Medicine, 7(30), pp.1~9 (2015)[DOI]
  3. Allonneau A, "Lead contamination among Paris fire brigade firefighters who fought the Nortre Dame Cathedral fire in Paris", International Journal of Hygiene and Environmental Health, 233(7), pp.7~76 (2021)[DOI]
  4. Raj K, "Lead pollution : Impact on environmental and human health and approach for a sustainable solution", Environmental Chemistry and Ecotoxicology, 5, pp.79~85 (2023)[DOI]
  5. Kim SC, "Medical radiation shielding effect by composition of barium compounds", Annals of Nuclear Energy, 47, pp.1~5 (2021)[DOI]
  6. Kim SJ, "Developement and performance evaluati on of eco-friendly X-ray shields using BaSO4", Journal of Radiation Industry, 14(10), pp.13~18 (2020)
  7. "Current status of national medical radiation use 2020~2022" (2024)[URL]
  8. Obchowski N, "Receiver operating characteristic curves and their use in radiology", Radiology, 229(1), pp.3~8 (2003)[DOI]
  9. Lee HM, "Effect of back scatter radiaton on CR image", Korean Journal of Digital Imaging in Medicine, 2(1), pp.68~72 (1996)
  10. Pierluigi C, "Real time dosimetry with radiochromic film", Scientific Reports, 9(5307), pp.1~11 (2019)[DOI]
  11. Katoh Y, "Shielding evaluation of lead-free boards for diagnostic X-ray", Japanese Society of Radiological Technology, 66(12), pp.1555~1560 (2010)[DOI]
  12. Esen Y, "An investigation of X-ray and radio isotope energy absorption of heavyweight concretes containing barite", Bulletin of Material Science, 34(1), pp.169~175 (2011)[DOI]
  13. Fukuda A, "Estimation of primary radiation output for wide-beam computed tomography scanner", Journal of Applied Clinical Medical Physics, 20(6), pp.152~159 (2019)[DOI]
  14. Kim KJ, "A Study on the shielding element using Monte Carlo simulation", Journal of Radiological Science and Technology, 40(2), pp.187~195 (2011)[DOI]
  15. Jeong HW, "Development and performance comparison of silicon mixed shielding material", Journal of Radiological Science and Technology, 46(3), pp.187~195 (2023)[DOI]
  16. "A study on guidance for test and methods of radiation protection" (2012)[URL]
  17. Structure of shielding design for medical X-ray image facilities, "Harrison JD", ICRP, ~, pp.ICRP Report 147 (2012)
  18. Kim BE, "Clinical application of eco-friendly lead-free board as a lead replacement in medical radiation protection facilities", master, pp.2023
  19. "Protective devices against medical X-radiation (KS C IEC 61331-3)" (2022)[URL]