건강인의 백혈구 백분율에 대한 검색
Findings of Leukocyte Differential Count in Healthy Persons
1
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
.
Abstract
To evaluate reference intervals of the leukocyte differential count in healthy persons, we measured WBC and five leukocyte differential count, determined by Coulter Counter model GEN-S in 1,060(M: 495, F: 565) persons. The results were summarized as follows 1. The correlation between Coulter GEN-S and the manual leukocyte differential shows good correlation for neutrophils(r=0.88), lymphocytes(r=0.76), monocytes(r=0.52) and eosinophils(r=0.89). The correlation for basophils (r=0.18) was less acceptable. 2. Reference interval of WBC count : 3.5-9.9 x 103/mm'(Male: 4.1-10.5, Female: 3.6-9.2). 3. Reference interval of neutrophils ; 37.2-7 1.2%(Male: 38.0-7 1.2, Female: 36.5-7 1.3). 4. Reference interval of lymphocytes ; 20.5-52.5 % (Male: 19.9-5 1.1, Female: 21.4-53.4). 5. Reference interval of monocytes ; 3.2-9.6%(Male: 3.2-10.0, Female: 3.0-9.4). 6. Reference interval of eosinophils ; 0.0-5.9%(Male: 0.0-6.5, Female: 0.0-5.4). 7. Reference interval of basophils ; 0.0- 1.4%(Male: 0.0- 1.1, Female: 0.0- 1.7).
