초록

To estimate the measurement uncertainty for industrial platinum resistance thermometers(IPRTs) made with 3-wire connection, the immersion temperature profile was investigated using a liquid bath. Two types of IPRTs having lead wires made of silver and nickel were constructed and the immersion profiles were measured at temperatures from −50℃ to 250℃ using 3-wire and 4-wire method. As immersion depth and temperature increased, the resistances measured by 3-wire method increased linearly but not for 4-wire method. To calibrate a 3-wire IPRT, the immersion effect must be accounted for. We propose a linear equation to assess correctly the measurement uncertainty.

Abstract

To estimate the measurement uncertainty for industrial platinum resistance thermometers(IPRTs) made with 3-wire connection, the immersion temperature profile was investigated using a liquid bath. Two types of IPRTs having lead wires made of silver and nickel were constructed and the immersion profiles were measured at temperatures from −50℃ to 250℃ using 3-wire and 4-wire method. As immersion depth and temperature increased, the resistances measured by 3-wire method increased linearly but not for 4-wire method. To calibrate a 3-wire IPRT, the immersion effect must be accounted for. We propose a linear equation to assess correctly the measurement uncertainty.

참고문헌 (7)

  1. "Techniques for approximating the international temperature scale of 1990, BIPM", ~ (2005)
  2. P.Marcarino, "Proposal:new reference functions for industrial platinum resistance thermometers", TEMPMEKO 2004, 통권 : 2, pp.807~812 (2004)
  3. T.Weckstrom, "Pt-100 thermometers and the ITS- 90", TEMPMEKO 2004, 통권 : 2, pp.813~816 (2004)
  4. "산업용 저항온도계 표준교정절차서, KASTO 03- 50106-077, 한국계량측정협회", ~ (2003)
  5. "Smithells Metals Reference Book, 8th edition" (2004), Elsevier Butterworth-Heinemann
  6. H. Preston-Thomas, "The International Temperature Scale of 1990 (ITS-90)", Metrologia, 27, pp.3~10 (1990)
  7. "측정불확도 표현지침, KRISS-98-096-SP, 한국표준 과학연구원", ~ (1998)