본문 바로가기
KERD
논문KCI 등재

보상용 적외선 센서를 사용한 비분산 적외선 이산화탄소 센서의 온도특성

Temperature Dependency of Non-dispersive Infrared Carbon Dioxide Gas Sensor by using Infrared Sensor for Compensation

1

조회

0

다운로드

3

인용

0

참고문헌

키워드

주제분류공학일반

초록

NDIR CO2 gas sensor was built with ASIC implemented thermopile sensor which included temperature sensor and unique elliptical waveguide structures in this paper. The temperature dependency of dual infrared sensor module (CO2 and reference IR sensors) has been characterized and its output voltage characteristics according to the temperature and gas concentration were proposed for the first time. NDIR CO2 gas and reference IR sensors showed linear output voltages according to the variation of ambient temperatures from 243 K to 333 K and their slopes were 14.2 mV/K and 8.8 mV/K, respectively. The output voltages of temperature sensor also presented a linear dependency according to the ambient temperature and could be described with . The output voltage ratio between CO2 and reference IR sensors revealed irrelevant to the changes of ambient temperatures and gave a constant value around 1.6255 with standard deviation 0.008 at 0 ppm. The output voltage of CO2 gas sensor at zero ppm CO2 gas consisted of two components; one is caused by the HPB (half pass-band) of IR filter and the other is attributed to the part of CO2 absorption wavelength. The characteristics of output voltages of CO2 gas sensor could be accurately modeled with three parameters which are dependent upon the ambient temperatures and represented small average error less than 1.5% with 5% standard deviation.

Abstract

NDIR CO2 gas sensor was built with ASIC implemented thermopile sensor which included temperature sensor and unique elliptical waveguide structures in this paper. The temperature dependency of dual infrared sensor module (CO2 and reference IR sensors) has been characterized and its output voltage characteristics according to the temperature and gas concentration were proposed for the first time. NDIR CO2 gas and reference IR sensors showed linear output voltages according to the variation of ambient temperatures from 243 K to 333 K and their slopes were 14.2 mV/K and 8.8 mV/K, respectively. The output voltages of temperature sensor also presented a linear dependency according to the ambient temperature and could be described with . The output voltage ratio between CO2 and reference IR sensors revealed irrelevant to the changes of ambient temperatures and gave a constant value around 1.6255 with standard deviation 0.008 at 0 ppm. The output voltage of CO2 gas sensor at zero ppm CO2 gas consisted of two components; one is caused by the HPB (half pass-band) of IR filter and the other is attributed to the part of CO2 absorption wavelength. The characteristics of output voltages of CO2 gas sensor could be accurately modeled with three parameters which are dependent upon the ambient temperatures and represented small average error less than 1.5% with 5% standard deviation.

참고문헌 (19)

  1. M. S. Zuraimi, "Development and application of a protocol to evaluate impact of duct cleaning on IAQ of office buildings", Buildings and Environment, 56, pp.86~94 (2012)
  2. G. Visco, "Organic carbons and TOC in waters : an overview of the international norm for its measurements", Microchemical Journal, 79, pp.185~191 (2005)
  3. K. Kaneyasu, "A carbon dioxide gas sensor based on solid electrolyte for air quality control", Sensors and Actuators B, 66, pp.56~58 (2000)
  4. N. Kawasaki, "Fast and precise method for HPLC-size exclusion chromatography with UV and TOC(NDIR)detection : Importance of multiplr detectors to evaluate the characteristics of dissolved organic matter", Water research, 45, pp.6240~6248 (2011)
  5. K. Tian, "A permeable membrane capacitance sensor for inogenic gases : Application to the measurement of total organic carbon", Analytica Chimica Acta, 652, pp.245~250 (2009)
  6. L. Lindberg, "Breath alcohol concentration determined with a new analyzer using free exhalation predicts almost precisely the arterial blood alcohol concentration", Forensic Science International, 168, pp.200~207 (2007)
  7. S. Adam, "Detection limit improvement for NDIR ethylene gas detectors using passive approaches", Sensors and Actuators B, 175, pp.246~254 (2012)
  8. R. V. Maikala, "Modified Beer’s law-historical perspectives and relevance in near-infrared monitoring of optical properties of human tissue", Inter. Journal of Industrial Ergonomics, 40, pp.125~134 (2010)
  9. G. Zhang, "A miniature carbon dioxide gas sensor based on infrared absorption", Optics and Lasers in Engineering, 48, pp.1206~1212 (2010)
  10. 장성호, "두 개의 동일한 타원형 구조를 지닌 광 도파관의 특성", 한국가스학회지, 18(2), pp.48~54 (2014), 한국가스학회[DOI]
  11. S. H. Yi, "Characteristics of NDIR alcohol sensor with elliptical optical structures", J. Auto-Vech. Safety Assoc, 7(2), pp.39~43 (2015)
  12. R. Eisberg, "Quantumn Physics of Atoms, Molecules, Solids, Nuclei, and Particles", pp.1~25 (1985), John Wiley &Sons
  13. L. Jun, "Miniature low-power IR monitor for methane detection", Measurement, 44, pp.823~831 (2011)
  14. J. M. Park, "Novel NDIR CO2 sensor with two concave mirrors", Proc. of APCOT 2006 (2006)
  15. J. S. Park, "Nondispersive infrared ray CH4 gas sensor using focused infrared beam structures", Sensors and Materials, 23(3), pp.147~158 (2011)
  16. S. Kim, "Derivation of the centerwavelength shift of narrow-bandpass filter under temperature change", Optics Express, 12(23), pp.5634~5639 (2004)
  17. C. N. Chen, "Fully quantitative characterization of CMOSMEMS polysilicon", titanium thermopile infrared sensors, 2012, pp.161, 892~900
  18. T. Stolberg-Rohr, "Spectral design of temperature-invariant narrow bandpass filters for mid-infrared", Optics Express, 23(1), pp.580~596 (2015)

관련 논문AI추천