Although variable resistance displacement transducer is widely applied for its technical simplicity, electrical pointcontact limits measurement resolution (precision). The present study introduced roler shaped structure for electrical lineproduct. When the displacement was varied with steps of 5 m, 1 m, and down to 0.1 m, the correlation coefficientbetween resistance and displacement decreased lower than 0.9 in the commercialized product while the proto-typetransducer maintained values higher than 0.99. Therefore, the measurement resolution was enhanced by approximately10 times. Since manufacturing also seemed easier, the present results would enable more accurate and les expensivemanufacturing of variable resistance displacement transducer.

참고문헌 (7)

  1. "“기본 센서와 이론” 의공학 교육연구회 역편", pp.61~121 (1993)
  2. "Transducers for biomedical measurements", pp.114~189 (1974)
  3. "“정밀기계가공을 위한 미세변위 계측·분석시스템”", 13(1), pp.139~145 (2002)
  4. "“소형공기 챔버를 센서소자로 사용하는 새로운 호식기류 계측 기술”", 13(2), pp.79~84 (2004)
  5. "“선형 저항체를 이용하여 연속적인 호식기류를 측정하는 장치”" (2005), 대한민국, 발명특허 제10-2005-0087501호
  6. "“Principles of biomechanics”In: Biomedical Engineering Principles", pp.342~358 (2005), Taylor & Francis, Boca Raton
  7. "“이중이차코일을 이용한 LVDT 센서의 구현”", 15(1) (2004)

관련 논문AI추천