비화재보를 위한 X-ray tube 내 열 유동해석에 관한 연구
A study on thermal fluid analysis in X-ray tube for non-fire alarm
0
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Currently, Korea is an aging society, and it is expected to enter a super-aging society in about 4 years. Accordingly, many X-ray technologies are being developed. In X-rays, 99% of X-rays are converted into heat energy and 1% into light energy (X-rays). 99% of the thermal energy raises the temperature of the anode and its surroundings, and the cooling system is an important factor as overheating can affect the deterioration of X-ray quality and shortened lifespan. There is a method of forced air cooling using natural convection. Therefore, in this study, when X-rays were taken 5 times, Flow analysis was performed on heat removal according to temperature rise and cooling time for the heat generated at the anode of the X-ray tube (input power 60kW, 75kW, 90kW). Based on one-shot, the most rapid temperature rise section increased by more than 57% to 0.03 seconds, A constant temperature rises from 0.03 seconds to 0.1 seconds, It is judged that the temperature rises by about 8.2% or more at one time. After one-shot cooling, the cooling drops sharply from about 60% to 0.03 seconds, It is judged that the temperature has cooled by more than 86% compared to the temperature before shooting. One-shot is cooled by more than 86% with cooling time after 0.1 seconds, As the input power of the anode increases, the cooling temperature gradually increases. Since the tungsten of the anode target inside the X-ray tube may be damaged by thermal shock caused by a rapid temperature rise, an improvement method for removing thermal energy is required when using a high-input power supply.
Abstract
Currently, Korea is an aging society, and it is expected to enter a super-aging society in about 4 years. Accordingly, many X-ray technologies are being developed. In X-rays, 99% of X-rays are converted into heat energy and 1% into light energy (X-rays). 99% of the thermal energy raises the temperature of the anode and its surroundings, and the cooling system is an important factor as overheating can affect the deterioration of X-ray quality and shortened lifespan. There is a method of forced air cooling using natural convection. Therefore, in this study, when X-rays were taken 5 times, Flow analysis was performed on heat removal according to temperature rise and cooling time for the heat generated at the anode of the X-ray tube (input power 60kW, 75kW, 90kW). Based on one-shot, the most rapid temperature rise section increased by more than 57% to 0.03 seconds, A constant temperature rises from 0.03 seconds to 0.1 seconds, It is judged that the temperature rises by about 8.2% or more at one time. After one-shot cooling, the cooling drops sharply from about 60% to 0.03 seconds, It is judged that the temperature has cooled by more than 86% compared to the temperature before shooting. One-shot is cooled by more than 86% with cooling time after 0.1 seconds, As the input power of the anode increases, the cooling temperature gradually increases. Since the tungsten of the anode target inside the X-ray tube may be damaged by thermal shock caused by a rapid temperature rise, an improvement method for removing thermal energy is required when using a high-input power supply.
참고문헌 (5)
- 이석문, "Thermal Characteristics of Rotating Anode X-ray Tube with Emissivity in Aging Process for Digital Radiography", Applied Science and Convergence Technology, 24(5), pp.125~131 (2015), 한국진공학회[DOI]
- 이석문, "Compact Anode Design with the Heat Capacity Performance in Rotating Anode X-ray Tube for Digital Radiography", Applied Science and Convergence Technology, 24(5), pp.136~141 (2015), 한국진공학회[DOI]
- 전성오, "냉이온수기 냉각시스템에 관한 열유동 해석", 한국유체기계학회 논문집, 14(3), pp.33~38 (2011), 한국유체기계학회
- 손창우, "유동해석을 통한 유체제어벨브 시스템의 내부 유동 특성 분석", 한국산학기술학회논문지, 19(6), pp.160~166 (2018), 한국산학기술학회[DOI]
- 이준엽, "텅스텐의 열충격 손상에 대한 초음파 특성 평가", 동력시스템공학회지, 22(4), pp.5~10 (2018), 한국동력기계공학회[DOI]