Objectives: The purpose of this study is to evaluate the accuracy of the digital models from Cone beam computed tomography (CBCT) scans of orthodontic diagnostic models by comparing the Digital Imaging & Communication in Medicine (DICOM) file to Stereolithography (STL) conversion data obtained by various voxel sized CBCT machines with the digital model obtained by a 3D optical extraoral scanner. Materials and Methods: Three orthodontic diagnostic models were scanned with a 3D optical scanner and stored as STL file data. The three models were scanned with the 5 modes which were Object scan (100 μm), 300, 400 μm of RAYSCAN α+ and 300, 390 μm of Alphard 3030. The DICOM files were converted to STL file data. For each orthodontic model, two digital models obtained by CBCT and optical scanner were paired, and GOM inspect software was used to superimpose and obtain the deviation between two digital models. One way ANOVA test was performed to determine whether there is a statistical difference in the mean value of deviations between digital models by the 5 protocols. Results: In three diagnostic models, the mean values of the deviations between the digital model of the Object scan (100 μm) and the digital model of 3D optical scanner were the smallest in comparison with other image protocols and there was a statistically significant difference (p<0.001). The mean values of the deviations were in the range of 0.0967 ~0.1022 mm. On the other hand, deviations of STL data from 400 μm of RAYSCAN α+ showed the largest mean value from 0.4007 to 0.4534 mm. Conclusions: The STL file data of the Object scan (100 μm) of RAYSCAN α+ is more accurate than those of large voxel sized protocols of CBCT. It shows the availability as an orthodontic digital model.
Abstract
Objectives: The purpose of this study is to evaluate the accuracy of the digital models from Cone beam computed tomography (CBCT) scans of orthodontic diagnostic models by comparing the Digital Imaging & Communication in Medicine (DICOM) file to Stereolithography (STL) conversion data obtained by various voxel sized CBCT machines with the digital model obtained by a 3D optical extraoral scanner. Materials and Methods: Three orthodontic diagnostic models were scanned with a 3D optical scanner and stored as STL file data. The three models were scanned with the 5 modes which were Object scan (100 μm), 300, 400 μm of RAYSCAN α+ and 300, 390 μm of Alphard 3030. The DICOM files were converted to STL file data. For each orthodontic model, two digital models obtained by CBCT and optical scanner were paired, and GOM inspect software was used to superimpose and obtain the deviation between two digital models. One way ANOVA test was performed to determine whether there is a statistical difference in the mean value of deviations between digital models by the 5 protocols. Results: In three diagnostic models, the mean values of the deviations between the digital model of the Object scan (100 μm) and the digital model of 3D optical scanner were the smallest in comparison with other image protocols and there was a statistically significant difference (p<0.001). The mean values of the deviations were in the range of 0.0967 ~0.1022 mm. On the other hand, deviations of STL data from 400 μm of RAYSCAN α+ showed the largest mean value from 0.4007 to 0.4534 mm. Conclusions: The STL file data of the Object scan (100 μm) of RAYSCAN α+ is more accurate than those of large voxel sized protocols of CBCT. It shows the availability as an orthodontic digital model.
참고문헌 (18)
- De Luca Canto G, "Intra-arch dimensional measurement validity of laserscanned digital dental models compared with the original plaster models: a systematic review", Orthod Craniofac Res, 18(2), pp.65~76 (2015)
- Swennen GR, "A cone-beam computed tomography triple scan procedure to obtain a three-dimensional augmented virtual skull model appropriate for orthognathic surgery planning", J Craniofac Surg, 20(2), pp.297~307 (2009)
- Sousa MVS, "Accuracy and reproducibility of 3-dimensional digital model measurements", Am J Orthod Dentofacial Orthop, 142(2), pp.269~273 (2012)
- Hassan WNW, "Assessing agreement in measurements of orthodontic study models: Digital caliper on plaster models vs 3-dimensional software on models scanned by structured-light scanner", Am J Orthod Dentofacial Orthop, 150(5), pp.886~895 (2016)
- 김종은, "임플란트 가이드 수술을 위한 Planning 방식에 대한 고찰", 대한치과의사협회지, 54(2), pp.108~122 (2016), 대한치과의사협회[DOI]
- 김재홍, "인상 스캐닝 방법에 의해 제작된 디지털 치과 모형의 체적안정성 평가", 치위생과학회지, 14(1), pp.15~21 (2014), 한국치위생과학회
- 김려운, "치과 디지털 인상의 이해와 적용", 대한치과재료학회지, 41(4), pp.253~261 (2014), 대한치과재료학회[DOI]
- Vogel AB, "Optical 3D scans for orthodontic diagnostics performed on full-arch impressions", J Orofac Orthop, 76(6), pp.493~507 (2015)
- Ji-won Anh, "A comparison of the precision of three-dimensional images acquired by 2 digital intraoral scanners: effects of tooth irregularity and scanning direction", 대한치과교정학회지, 46(1), pp.3~12 (2016), 대한치과교정학회
- Kang S-H, "Dental image replacement on cone beam computed tomography with threedimensional optical scanning of a dental cast, occlusal bite, or bite tray impression", Int J Oral Maxillofac Surg, 43(10), pp.1293~1301 (2014)
- Wesemann C, "Accuracy and efficiency of full-arch digitalization and 3D printing: A comparison between desktop model scanners, an intraoral scanner, a CBCT model scan, and stereolithographic 3D printing", Quintessence Int (2016)
- de Waard O, "Reproducibility and accuracy of linear measurements on dental models derived from cone-beam computed tomography compared with digital dental casts", Am J Orthod Dentofacial Orthop, 146(3), pp.328~336 (2014)
- DeLong R, "Accuracy of a system for creating 3D computer models of dental arches", J Dent Res, 82(6), pp.438~442 (2003)
- 이경탁, "치과용 스캐너로 채득한 디지털 모형의 반복성 및 재현성에 관한 3차원적 평가", 대한치과기공학회지, 34(3), pp.213~220 (2012), 대한치과기공학회
- Zhao S, "X-ray CT metal artifact reduction using wavelets: an application for imaging total hip prostheses", IEEE Trans Med Imaging, 19(12), pp.1238~1247 (2000)
- Rangel FA, "Accuracy and Reliability of a Novel Method for Fusion of Digital Dental Casts and Cone Beam Computed Tomography Scans", PLoS One, 8(3), pp.e59130~ (2013)
- Widmann G, "Errors and error management in image-guided craniomaxillofacial surgery", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 107(5), pp.701~715 (2009)