B3 선량계 기반 투과깊이선량 분포의 해석적 방법을 통한 10 MeV 전자가속기 공정의 균일성 평가
Evaluation of Uniformity of 10 MeV Electron Beam Processing through Analytical Method of Penetration Depth Distribution Based on B3 Dosimeter
2
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
The demand for industrial using through accelerators is increasing. There are variousmethods to measure the absorbed dose of the sample irradiated with electron beam, but amongthese, a comparatively easy and simple method is used to evaluate the B3 film dosimeter developedby the Riso Institute of Denmark. Accordingly, the B3 dosimeter was used to check the PenetrationDepth Distribution (PDD) for each medium and determine the optimal dose according to theexperiment. After verifying the linearity of the dose according to the conveyor speed in order tosecure the reliability of the results and confirm the stability of the operation, the experiment wasconducted based on ISO / ASTTM 51649, which is known as a guideline for dose evaluation in thefield of radiation processing. The PDD graph was respectively checked on the Polyethylene (PE),Polyethylene terephthalate (PET), Acrylic, Polyvinyl chloride (PVC) and Aluminum plate withdensity of 0.722, 1.206, 1.289, 1.611, and 2.667 g·cm-3. The optimal depth was reached at about34 cm for low density materials and about 17 to 24 cm for high density. It was verified throughexperiments that it was dominant in density, but it was found that in the case of acrylic, about 5 cmof thickness was required to reach the optimum depth compared to PET with low density. Throughthis, in addition to physical factors such as density, the mechanism according to the binding state ofthe polymer material should be identified and will be studied in the future work.
Abstract
The demand for industrial using through accelerators is increasing. There are variousmethods to measure the absorbed dose of the sample irradiated with electron beam, but amongthese, a comparatively easy and simple method is used to evaluate the B3 film dosimeter developedby the Riso Institute of Denmark. Accordingly, the B3 dosimeter was used to check the PenetrationDepth Distribution (PDD) for each medium and determine the optimal dose according to theexperiment. After verifying the linearity of the dose according to the conveyor speed in order tosecure the reliability of the results and confirm the stability of the operation, the experiment wasconducted based on ISO / ASTTM 51649, which is known as a guideline for dose evaluation in thefield of radiation processing. The PDD graph was respectively checked on the Polyethylene (PE),Polyethylene terephthalate (PET), Acrylic, Polyvinyl chloride (PVC) and Aluminum plate withdensity of 0.722, 1.206, 1.289, 1.611, and 2.667 g·cm-3. The optimal depth was reached at about34 cm for low density materials and about 17 to 24 cm for high density. It was verified throughexperiments that it was dominant in density, but it was found that in the case of acrylic, about 5 cmof thickness was required to reach the optimum depth compared to PET with low density. Throughthis, in addition to physical factors such as density, the mechanism according to the binding state ofthe polymer material should be identified and will be studied in the future work.
참고문헌 (12)
- Benny PG, "Comparison of graphite calorimeter dosimetry system with Monte Carlo simulation at an industrial electron beam accelerator", Nucl Instrum. Methods Phys. Res. A, 703, pp.98~101 (2013)
- 변명우, "식품 및 공중보건산업에서 방사선 조사기술 이용 국내·외 현황", Food Science and Preservation, 10(1), pp.106~123 (2003), 한국식품저장유통학회
- Chai JS, "The Status and Prospects of Accelerators in Radiation Industries", J. Radiat. Ind, 6(1), pp.101~110 (2012)
- Choi JH, "Radiation Processing of Polymeric Materials", Polym. Sci. Technol, 18(3), pp.253~258 (2007)
- 하태성, "국제표준에 따른 10MeV급 전자빔 조사시설의 흡수선량 품질보증에 관한 연구", 방사선기술과학, 33(4), pp.387~394 (2010), 대한방사선과학회
- Han DH, "Industrial Application of Electron Beam Accelerator. News & Information for Chemical Engineers", Korean J. Chem. Eng., 21(4), pp.484~490 (2003)
- Hisada K, "Industralization of fabric processing through electron-beam grafting", J. Fiber Sci. Technol, 65(3), pp.373~376 (2009)
- ISO, "Standard Practice for Dosimetry in an Electron Beam Facility for Radiation Processing at Energies Between 300 keV and 25 MeV", ~ (2015)
- Lee BC, "Industrial Use of Electronic Acclerator, Physics & high technology", J. Korean Phys. Soc, 15(3), pp.21~28 (2006)
- 이윤종, "Performance Qualifications for the Beam in a 10 MeV Electron Accelerator", 방사선산업학회지, 13(4), pp.333~338 (2019), (사)한국방사선산업학회[DOI]
- 임희진, "연구용 선형가속기의 전자총 가열 전류에 따른 전자선의 에너지 인자 측정과 출력 측정 연구", 의학물리, 27(1), pp.25~30 (2016), 한국의학물리학회
- Miller A, "A new radiochromic thin-film dosimeter system", Radiat Phys Chem, 31, pp.491~496 (1988)