Performance Qualifications for the Beam in a 10 MeV Electron Accelerator

2

조회

0

다운로드

1

인용

0

참고문헌

키워드

초록

The Advanced Radiation Technology Institute (ARTI) introduced an electron beamaccelerator capable of continuous radiation processing in 2018. The instrument can be used forempirical research to industrialize already developed radiation technologies. Radiation processinghas minimum absorbed dose requirements and maximum absorbed dose limits. It is important forradiation accelerators to maintain constant values for the device̓s operating parameters to meetthe absorbed dose required by the product (AAPM 1998). Therefore, an inspection (PerformanceQualification) of the accelerator performance is required, and the operation values of all processesneed to be set so that the absorption dose requirements can be met. To maintain dose requirementsduring the process, the processing variables need to be controlled: electronic energy, beam current,and parameters according to material handling (transfer vehicle speed or survey time), beamwidth, and characteristics of the processing process and conditions of investigation. In this study,the performance test of the electron-ray accelerators presented by the international standards ofISO/ASTM was carried out. Assessments include Electron Beam Energy, Average Beam Current,Beam Size (Beam Spot Size), Spoken Size and Homogeneity (Scan Size and Uniformity of Unit), andMoving Dose Assessment (Uniformity). This allowed individual measurements to be obtained withinthe allowable range for each item specified in ISO/ASTM 51649. Periodic inspection procedures willneed to be established to maintain a constant dose for the devices.

Abstract

The Advanced Radiation Technology Institute (ARTI) introduced an electron beamaccelerator capable of continuous radiation processing in 2018. The instrument can be used forempirical research to industrialize already developed radiation technologies. Radiation processinghas minimum absorbed dose requirements and maximum absorbed dose limits. It is important forradiation accelerators to maintain constant values for the device̓s operating parameters to meetthe absorbed dose required by the product (AAPM 1998). Therefore, an inspection (PerformanceQualification) of the accelerator performance is required, and the operation values of all processesneed to be set so that the absorption dose requirements can be met. To maintain dose requirementsduring the process, the processing variables need to be controlled: electronic energy, beam current,and parameters according to material handling (transfer vehicle speed or survey time), beamwidth, and characteristics of the processing process and conditions of investigation. In this study,the performance test of the electron-ray accelerators presented by the international standards ofISO/ASTM was carried out. Assessments include Electron Beam Energy, Average Beam Current,Beam Size (Beam Spot Size), Spoken Size and Homogeneity (Scan Size and Uniformity of Unit), andMoving Dose Assessment (Uniformity). This allowed individual measurements to be obtained withinthe allowable range for each item specified in ISO/ASTM 51649. Periodic inspection procedures willneed to be established to maintain a constant dose for the devices.

참고문헌 (12)

  1. Radiochromic Film Dosimetry, "American Association of Physicists in Medicine", Radiation Therapy Committee, 55~, pp.AAPM Report No. 63 (1998)
  2. Soares CG, "Radiochromic film dosimetry", Radiation Measurements, 41, pp.S100~S116 (2007)
  3. International Atomic Energy Agency, "Gamma Irradiators for Radiation Processing" (2000)
  4. Practice for dosimetry in an electron beam facility for radiation processing at energies between 300 keV and 25 MeV, "ISO", ASTM International, ASTM, pp.2015, ~[DOI]
  5. Lawrence EO, "The production of heavy high-speed ions without the use of high voltages", Phys. Rev, 38, pp.2022~ (1931)
  6. Martins MN, "Electron accelerators : History, applications and perspectives", Radiation Physics and Chemistry, 95, pp.78~85 (2014)
  7. Teixeira MI, "Dosimetric properties of various colored commercial glasses", Applied Radiation and Isotopes, 57, pp.407~413 (2002)
  8. Butson MJ, "Radiochromic film for medical radiation dosimetry", Materials Science and Engineering, R41, pp.61~120 (2003)
  9. Dosimetry for Electron Beam Applications, "Miller A", Risø National Laboratory, ~, pp.Risø-M, No. 2401 (2019)
  10. Chilkulwar RH, "Dosimetric evaluation of an indigenously developed 10 MeV industrial electron beam irradiator", Radiation Measurements, 47, pp.628~633 (2012)
  11. Wideröe R, "Über ein neues Prinzip zur Herstellung hoher Spannungen", Archiv für Elektrotechnik, 21, pp.387~406 (1928)
  12. Zofia PS, "A comparison of the performance characteristics of four film dosimeters in a 10-MeV electron beam", Applied Radiation and Isotopes, 66, pp.346~352 (2008)

관련 논문AI추천

이 간행물의 다른 논문 (12건)