Purpose: The gonad shielding is used to minimize the impact of the exposure to gonadswhen Abdomen, Pevis and Hip X-ray inspections are conducted on radiation impressionable pediatricpatients. By the way, the gonad is palpable difficult and impossible to check visually becauseit’s a sensitive area, so tests are conducted with the approximate location of shielding, therebyappearing problems of not shielding gonads accurately. Accordingly, this study produced shieldsby age and gender of pediatric patients and studied the method of positioning shields with ASIS asa reference point without palpable sensitive areas, and tried to evaluate its usability. Materials andmethods: The study surveyed 30 pediatric patients by gender and age, who came and got inspectedin Department of Radiology, our hospital from February 2012 to January 2014 and obtained thevalue of tolerance by measuring the average size of the pelvis using the distance measurementfunction of Infinitt Piview with the images stored in the PACS and producing shields by age andgender of pediatric patients and specifying the areas at random for the comparative analysis ofpre- and post-using. It calculated the technology statistics (mean±SD) with the value of tolerancemeasured the length using SPSS 12.0 statistical program. Results: As for boys, differences in thetolerance range of pre- and post-using shields were 2.69 mm in case of 1 year old, 2.58 mm in2 years, 2.37 mm in 3 years, 2.815 mm in 4~5 years, 2.043 mm in 7~10 years, and as for girls,1.92 mm in 1~2 years, 1.75 mm in 3~4 years, 2.52 mm in 5~6 years and 1.93 mm in 7~10. After analyzing the pre- and post-using shields for all of boys and girls, there were statisticallysignificant differences (P<0.050). Conclusions: It is considered that we can minimize the exposureto gonads and get a better video for diagnosis in testing high biological impressionable pediatric,if we use shields correctly with ASIS as a reference point considering its shape and size by age and gender in Abdomen, Pevis and Hip X-ray inspections.
Abstract
Purpose: The gonad shielding is used to minimize the impact of the exposure to gonadswhen Abdomen, Pevis and Hip X-ray inspections are conducted on radiation impressionable pediatricpatients. By the way, the gonad is palpable difficult and impossible to check visually becauseit’s a sensitive area, so tests are conducted with the approximate location of shielding, therebyappearing problems of not shielding gonads accurately. Accordingly, this study produced shieldsby age and gender of pediatric patients and studied the method of positioning shields with ASIS asa reference point without palpable sensitive areas, and tried to evaluate its usability. Materials andmethods: The study surveyed 30 pediatric patients by gender and age, who came and got inspectedin Department of Radiology, our hospital from February 2012 to January 2014 and obtained thevalue of tolerance by measuring the average size of the pelvis using the distance measurementfunction of Infinitt Piview with the images stored in the PACS and producing shields by age andgender of pediatric patients and specifying the areas at random for the comparative analysis ofpre- and post-using. It calculated the technology statistics (mean±SD) with the value of tolerancemeasured the length using SPSS 12.0 statistical program. Results: As for boys, differences in thetolerance range of pre- and post-using shields were 2.69 mm in case of 1 year old, 2.58 mm in2 years, 2.37 mm in 3 years, 2.815 mm in 4~5 years, 2.043 mm in 7~10 years, and as for girls,1.92 mm in 1~2 years, 1.75 mm in 3~4 years, 2.52 mm in 5~6 years and 1.93 mm in 7~10. After analyzing the pre- and post-using shields for all of boys and girls, there were statisticallysignificant differences (P<0.050). Conclusions: It is considered that we can minimize the exposureto gonads and get a better video for diagnosis in testing high biological impressionable pediatric,if we use shields correctly with ASIS as a reference point considering its shape and size by age and gender in Abdomen, Pevis and Hip X-ray inspections
참고문헌 (8)
- 김정수, "소아암 환자의 양성자치료 시 DIPS 촬영의 피폭선량에 관한 연구", 석사, 방사선학, pp.2010
- 박창현, "소아 골반 검사 시 생식선 차폐체 사용 유무에 따른 선량 고찰", 제46차 전국방사선사 종합학술대회 (2011), 대한영상의학기술학회
- "소아 방사선 권고사항", International Atomic Energy Agency (IAEA), ~
- "소아 방사선 촬영을 위한 기술정보", 식품의약품 안전청, 1~2 (2010)
- Bardo DM, "Location of the ovaries in girls from newborn to 18 years of age : reconsidering ovarian shielding", Pediatr Radiol, 39(3), pp.253~259 (2009)
- Fawcett SL, "The use of gonad shielding in paediatric hip and pelvis radiographs", BJR, 82(977), pp.363~370 (2009)
- Radiation Health Newsletter, "식품의약품안전청 통권 52호, Vol. 14", No. 3, pp.2008, ~
- "UNSCEAR 2000 Report Vol. I Source and Effcets of Ionization Radiation", AnnexD Medical Radiation Exposure, ~