The apron is one of the essential protectors to reduce the exposure dose of radiological technologists. This study is to provide a guideline for purchasing the aprons with excellent performance and to help reducing the exposure dose by measuring the shielding ration and uniformity of aprons according to lead equivalent and form types. The shielding ratio of aprons were measured by using radiation generator and dosimeter. Exposure conditions were 81 kVp, 25 mAs, source to image receptor distance (SID) 100 cm and field of view (FOV) 17"×17". Exposure areas for front type and around type aprons were divided into 9 areas and for 2 pieces type aprons were divided into 3 areas of top and 4 areas of skirt. The uniformity of aprons were measured by using fluoroscopy and Image J. The 4 regions of interest (ROI) were set into acquired images and measured uniformity by measuring the standard deviation of pixel intensity in ROIs. In continuous shielding ration measurement of aprons according to exposure area, there was not statistical significance (P>0.05). In ANOVA test of aprons, there was statistical significance (P<0.01). In the results of sheilding ratio, although the aprons had equal lead equivalent, there were difference in shielding ratio from 83.59% to 98.15%. In the results of uniformity, the front type aprons with equal lead equivalent indicated the similar uniformity. However, the around type and 2 pieces type apron with equal lead equivalent indicated the different uniformity each other, from 1.8 to 22.2. If the performance evaluation in this study were conducted regularly before and after purchase the aprons, the exposure does to patients and radiological technologists could be reduced.
Abstract
The apron is one of the essential protectors to reduce the exposure dose of radiological technologists. This study is to provide a guideline for purchasing the aprons with excellent performance and to help reducing the exposure dose by measuring the shielding ration and uniformity of aprons according to lead equivalent and form types. The shielding ratio of aprons were measured by using radiation generator and dosimeter. Exposure conditions were 81 kVp, 25 mAs, source to image receptor distance (SID) 100 cm and field of view (FOV) 17"×17". Exposure areas for front type and around type aprons were divided into 9 areas and for 2 pieces type aprons were divided into 3 areas of top and 4 areas of skirt. The uniformity of aprons were measured by using fluoroscopy and Image J. The 4 regions of interest (ROI) were set into acquired images and measured uniformity by measuring the standard deviation of pixel intensity in ROIs. In continuous shielding ration measurement of aprons according to exposure area, there was not statistical significance (P>0.05). In ANOVA test of aprons, there was statistical significance (P<0.01). In the results of sheilding ratio, although the aprons had equal lead equivalent, there were difference in shielding ratio from 83.59% to 98.15%. In the results of uniformity, the front type aprons with equal lead equivalent indicated the similar uniformity. However, the around type and 2 pieces type apron with equal lead equivalent indicated the different uniformity each other, from 1.8 to 22.2. If the performance evaluation in this study were conducted regularly before and after purchase the aprons, the exposure does to patients and radiological technologists could be reduced.
참고문헌 (12)
- 김기원, "복부 전후 방향 검사의 자동노출제어 사용 시 선량 비교 연구", 방사선기술과학, 38(3), pp.205~211 (2015), 대한방사선과학회[DOI]
- 김영근, "방사선 방호용 에이프런의 경량화와 차폐능 개선", 대한방사선과학회, 26(1), pp.45~50 (2003)
- 김유현, "진단방사선검사에서 환자피폭선량에 관한 연구", 방사선기술 과학회지, 28(3), pp.241~248 (2005)
- 송종남, "Over Tube Type의 투시촬영장치를 이용한 제조사별, 납당량별 엑스선방어 앞치마의 Protection 성능 평가에 관한 연구", 한국방사선학회논문지, 5(3), pp.135~141 (2011), 한국방사선학회
- 유세종, "방사선방어용앞치마(Apron)의 차폐율과 균일성 측정을 통한 성능평가 연구", 대한안전경영과학회지, 17(1), pp.103~109 (2015), 대한안전경영과학회[DOI]
- 추성실, "방사선의 의학적 이용에 따른 방사선 안전관리현황", 대한방사선방어의학회지, 15(1), pp.39~46 (1990)
- "한국산업규격(KS). 1990. X선 방호용품류의 납당량 시험 방법, A 4025", ~ (1990)
- "ICRP 103. 2007. Recommendations of the International Commision on Radiological Protection", ICRP Publication, ~ (2007)
- KFDA., "Interventional radiology radiation protection guidelines", KFDA, ~ (2006)
- Michel R, "Implementation of an X-ray radiation protective equipment inspection program", Health Phys, 82, pp.51~53 (2000)
- Orhan O, "How protective are the lead aprons we use against ionizing radiation", Diagn. Interv. Radiol, 18(2), pp.147~152 (2012)
- Stam W, "Inspection of lead aprons: A practical rejection model", Health Phys, 95(2), pp.133~136 (2008)