Dose Customized Apron Micro Functional Design Using Convergence Shielding Sheet
1
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Radiation shielding clothing for medical institutions is used based on lead equivalent of 0.25 mmPb. However, this study intends to study the shielding suit that can guarantee the user's activity while considering the sensitivity of each part of the body. By manufacturing based on eco-friendly shielding material, it was attempted to solve the weight problem and environmental problem of existing lead aprons, and to present the same shielding performance as lead equivalent in thickness. The fabric of the produced shielding sheet was manufactured through a calendar process that adjusts the thickness of the shielding sheet from lead equivalent 0.12 mmPb to 0.32 mmPb. In addition, the usability evaluation of the manufactured shielding clothes was conducted for the subjects who were workers in medical institutions. As a result, the activity became easier and the weight was reduced by 0.26 kg. In the future, it is thought that it is necessary to improve the shielding suit design considering the activity.
Abstract
Radiation shielding clothing for medical institutions is used based on lead equivalent of 0.25 mmPb. However, this study intends to study the shielding suit that can guarantee the user's activity while considering the sensitivity of each part of the body. By manufacturing based on eco-friendly shielding material, it was attempted to solve the weight problem and environmental problem of existing lead aprons, and to present the same shielding performance as lead equivalent in thickness. The fabric of the produced shielding sheet was manufactured through a calendar process that adjusts the thickness of the shielding sheet from lead equivalent 0.12 mmPb to 0.32 mmPb. In addition, the usability evaluation of the manufactured shielding clothes was conducted for the subjects who were workers in medical institutions. As a result, the activity became easier and the weight was reduced by 0.26 kg. In the future, it is thought that it is necessary to improve the shielding suit design considering the activity.
참고문헌 (20)
- K. Yue, "A new lead-free radiation shielding material for radiotherapy", Radiation Protection Dosimetry, 133(4), pp.256~260 (2009)[DOI]
- 김선칠, "방사선 차폐시트의 적층 구조와 섬유 코팅의 융합적인 현상이 인장강도에 미치는 영향", 한국융합학회논문지, 11(6), pp.83~88 (2020), 한국융합학회[DOI]
- Ha Lu, "Lightweight Lead Aprons: The Emperor's New Clothes in the Angiography Suite?", European Journal of Vascular and Endovascular Surgery, 57(5), pp.730~739 (2019), Elsevier BV[DOI]
- Shrawan Kumar, "Perceived physical stress at work and musculoskeletal discomfort in X-ray technologists", Ergonomics, 47(2), pp.189~201 (2007), Informa UK Limited[DOI]
- H. M. Soylu, "Gamma radiation shielding efficiency of a new lead-free composite material", Journal of Radioanalytical and Nuclear Chemistry, 305(2), pp.529~534 (2015), Springer Science and Business Media LLC[DOI]
- Testing method of lead equivalent for X-ray protective devices, "Korean Standards Association", KSA, ~, pp.KSA 4025 (2017)
- 조용인, "PET", CT 업무 환경에서 선원 취급 시 Apron의 실효성 평가, 2015, pp.38(3), 237~244[DOI]
- D. Gupta., "Functional clothing-Definition and classification", Indian Journal of Fibre &Textile Research, 36(4), pp.321~326 (2011)
- A. Majumdar, "Modelling, optimization and decision making techniques in designing of functional clothing", Indian Journal of Fibre & Textile Research, 36(4), pp.398~409 (2011)
- P. Rajaraman, "Human individual radiation sensitivity and prospects for prediction", Annals of the ICRP, 47(3-4), pp.126~141 (2018), SAGE Publications[DOI]
- W. Y. Park, "Correlation Between the Parameters of Radiosensitivity in Human Cancer Cell Lines", The Korean Society for Radiation Oncology, 16(2), pp.99~106 (1998)
- 김태희, "The radiation safety education and the pain physicians’efforts to reduce radiation exposure", The Korean Journal of Pain, 30(2), pp.104~115 (2017), 대한통증학회
- C. S. Lee, "Consideration of the ICRP 2006 revised tissue weighting factors on age-dependent values of the effective dose for external photons", Physics in Medicine and BioloThe 2007 Recommendations of the Internatgy, 52(1), pp.41~58 (2007)[DOI]
- "ICRP Publication 103, ional Commission on Radiological Protection", ~ (2007)
- 장덕현, "리커트형 척도의 중간점은 필요한가?: 중간점 있는 척도와 없는 척도 간 측정 결과 비교", 조사연구, 18(4), pp.1~24 (2017), 한국조사연구학회[DOI]
- C. H. LAWSHE, "A QUANTITATIVE APPROACH TO CONTENT VALIDITY", Personnel Psychology, 28(4), pp.563~575 (1975), Wiley[DOI]
- J. Nielsen, "A mathematical model of the finding of usability problems", Proceedings of the Interact '93 & Chi '93 Conference : Human Factors in Computing Systems, 5(1), pp.206~213 (1993)
- B. Sullivan., "Designing Web Usability : The Practice of Simplicity", Technical Communication, 47(3), pp.411~413 (2000)[DOI]
- L. J. Cronbach, "Essentials of Psychological Testing" (1949), Harper e Brothers
- Nadin Jamal AbuAlRoos, "Tungsten-based material as promising new lead-free gamma radiation shielding material in nuclear medicine", Physica Medica, 78, pp.48~57 (2020), Elsevier BV[DOI]