Purpose: This study aimed to measure the test-retest reliability of the Korean Matrix sentencein- noise test in a sound-field testing condition. Methods: Fifty-five listeners with normal hearing participated. The sound-field speech-in-noise test was administered with a front loudspeaker by presenting noise at a fixed level (65 dB SPL) while adaptively adjusting the sentence level based on the listener’s response. The speech reception thresholds (SRTs) corresponding to 50% sentencein- noise intelligibility were obtained using speech-shaped noise (SSN), International Collegium of Rehabilitative Audiology (ICRA) noise, and International Speech Test Signal (ISTS) noise. For the testretest reliability, each subject was tested twice in a 2-week interval. The test-retest reliabilities were judged by the intraclass correlation coefficient (ICC), the minimal detectable change (MDC) values, and Bland-Altman plots. Results: The test-retest reliabilities of the Korean Matrix sentence test were good (ICC, 0.84~0.86) across SSN, ICRA, ISTS noises. The MDC was 0.92 dB SNR for SSN and 2.83 and 2.98 dB SNR for ICRA and ISTS, indicating the random measurement error varied across the types of noise. Conclusion: The Korean Matrix sentence-in-noise test can be used as a reliable tool to assess speech-in-noise intelligibility in sound-field testing condition. Further study across various sessions would strengthen the present finding.

Abstract

Purpose: This study aimed to measure the test-retest reliability of the Korean Matrix sentencein- noise test in a sound-field testing condition. Methods: Fifty-five listeners with normal hearing participated. The sound-field speech-in-noise test was administered with a front loudspeaker by presenting noise at a fixed level (65 dB SPL) while adaptively adjusting the sentence level based on the listener’s response. The speech reception thresholds (SRTs) corresponding to 50% sentencein- noise intelligibility were obtained using speech-shaped noise (SSN), International Collegium of Rehabilitative Audiology (ICRA) noise, and International Speech Test Signal (ISTS) noise. For the testretest reliability, each subject was tested twice in a 2-week interval. The test-retest reliabilities were judged by the intraclass correlation coefficient (ICC), the minimal detectable change (MDC) values, and Bland-Altman plots. Results: The test-retest reliabilities of the Korean Matrix sentence test were good (ICC, 0.84~0.86) across SSN, ICRA, ISTS noises. The MDC was 0.92 dB SNR for SSN and 2.83 and 2.98 dB SNR for ICRA and ISTS, indicating the random measurement error varied across the types of noise. Conclusion: The Korean Matrix sentence-in-noise test can be used as a reliable tool to assess speech-in-noise intelligibility in sound-field testing condition. Further study across various sessions would strengthen the present finding.

참고문헌 (33)

  1. Akeroyd, M. A., "International Collegium of Rehabilitative Audiology (ICRA) recommendations for the construction of multilingual speech tests, ICRA working group on multilingual speech tests", International Journal of Audiology, 54(Suppl 2), pp.17~22 (2015)
  2. Bland, J. M., "Statistical methods for assessing agreement between two methods of clinical measurement", The Lancet, 1(8476), pp.307~310 (1986)
  3. Brand, T., "Efficient adaptive procedures for threshold and concurrent slope estimates for psychophysics and speech intelligibility tests", The Journal of the Acoustical Society of America, 111(6), pp.2801~2810 (2002)
  4. "Practice Guidance: The Acoustics of Sound Field Audiometry in Clinical Audiological Applications" (2019)[URL]
  5. Bronkhorst, A. W., "A clinical test for the assessment of binaural speech perception in noise", Audiology, 29(5), pp.275~285 (1990)
  6. Cooke, M., "A glimpsing model of speech perception in noise", The Journal of the Acoustical Society of America, 119(3), pp.1562~1573 (2006)
  7. Dreschler, W. A., "ICRA noises: Artificial noise signals with speechlike spectral and temporal properties for hearing instrument assessment, International collegium for rehabilitative audiology", Audiology, 40(3), pp.148~157 (2001)
  8. Fogerty, D., "Glimpsing speech in temporally and spectro-temporally modulated noise", The Journal of the Acoustical Society of America, 143(5), pp.3047~3057 (2018)
  9. Giavarina, D., "Understanding Bland Altman analysis", Biochemia Medica, 25(2), pp.141~151 (2015)
  10. Haley, S. M., "Interpreting change scores of tests and measures used in physical therapy", Physical Therapy, 86(5), pp.735~743 (2006)
  11. Holube, I., "Development and analysis of an International Speech Test Signal(ISTS)", International Journal of Audiology, 49(12), pp.891~903 (2010)
  12. Hwang, J. S., "The effect of background noise on immediate free recall of words in younger and older listeners with hearing loss", Journal of the American Academy of Audiology, 31(10), pp.701~707 (2020)
  13. International Organization for Standardization, "Acoustics:Audiometric Test Methods. Part 2: Sound Field Audiometry with Puretone and Narrow-Band Test Signals", International Organization for Standardization, ~ (2009)
  14. International Organization for Standardization, "Acoustics—Audiometric Test Methods—Part 3: Speech Audiometry", International Organization for Standardization, ~ (2012)
  15. Jensen, K. K., "The fluctuating masker benefit for normal-hearing and hearing-impaired listeners with equal audibility at a fixed signal-to-noise ratio", The Journal of the Acoustical Society of America, 145(4), pp.2113~2125 (2019)
  16. 김광민, "Prediction of Software-Based Sentence-in-Noise Recognition Thresholds of Hearing-Impaired Listeners", Audiology and Speech Research, 16(2), pp.140~146 (2020), 한국청각언어재활학회
  17. 김경현, "한국어 Matrix 문장 검사 도구 평가: 목록 간 동질성과 단어 위치의 영향에 대한 검증", Audiology and Speech Research, 14(2), pp.100~107 (2018), 한국청각언어재활학회
  18. 김경현, "측정 절차가 건청인의 한국어 Matrix 소음하 문장인지에 미치는 영향", 대한이비인후-두경부외과학회지, 64(1), pp.7~12 (2021), 대한이비인후과학회[DOI]
  19. Killion, M. C., "Development of a quick speech-in-noise test for measuring signal-to-noise ratio loss in normal-hearing and hearing-impaired listeners", The Journal of the Acoustical Society of America, 116(4), pp.2395~2405 (2004)
  20. 김나경, "고정된 신호대잡음비에서의 소음 하 문장인지도의 신뢰도와 타당도에 관한 연구", Audiology and Speech Research, 13(1), pp.62~69 (2017), 한국청각언어재활학회
  21. 고요셉, "보청기 착용 효과 측정을 위한 음장어음인지평가의 임상적 중요성", 대한이비인후-두경부외과학회지, 64(5), pp.310~320 (2021), 대한이비인후과학회[DOI]
  22. Kollmeier, B., "The multilingual matrix test: Principles, applications, and comparison across languages: A review", International Journal of Audiology, 54(Suppl 2), pp.3~16 (2015)
  23. Landis, J. R., "The measurement of observer agreement for categorical data", Biometrics, 33(1), pp.159~174 (1977)
  24. 이재희, "변동형 소음 하 문장인지 평가 도구의 비교", Audiology and Speech Research, 13(1), pp.9~18 (2017), 한국청각언어재활학회
  25. Lu, W. S., "The minimal detectable change of the simplified stroke rehabilitation assessment of movement measure", Journal of Rehabilitation Medicine, 40(8), pp.615~619 (2008)
  26. Miller, R. E., "Glimpsing speech interrupted by speech-modulated noise", The Journal of the Acoustical Society of America, 143(5), pp.3058~3067 (2018)
  27. Moon, S. K., "The Korean hearing in noise test", International Journal of Audiology, 47(6), pp.375~376 (2008)
  28. Nilsson, M., "Development of the hearing in noise test for the measurement of speech reception thresholds in quiet and in noise", The Journal of the Acoustical Society of America, 95(2), pp.1085~1099 (1994)
  29. Taylor, B., "Speech-in-noise tests: How and why to include them in your basic test battery", The Hearing Journal, 56(1), pp.40~ (2003)
  30. Shrout, P. E., "Intraclass correlations : Uses in assessing rater reliability", Psychological Bulletin, 86(2), pp.420~428 (1979)
  31. Szalma, J. L., "Noise effects on human performance : A meta-analytic synthesis", Psychological Bulletin, 137(4), pp.682~707 (2011)
  32. Wilson, R. H., "Clinical experience with the words-in-noise test on 3430 veterans : Comparisons with pure-tone thresholds and word recognition in quiet", Journal of the American Academy of Audiology, 22(7), pp.405~423 (2011)
  33. Yi, D., "Study on the construction and optimization of the Korean matrix sentence materials", master, pp.2016