Bimodal hearing refers to the combination of the cochlear implant in one ear and the hearing aid in the opposite ear. Recently, cochlear implant candidacy criteria have been expanded and the number of people receiving cochlear implants with residual hearing has increased, leading to the raised bimodal population. Providing professional fitting services based on systematic bimodal fitting methods or guidelines are essential for the successful bimodal intervention of the hearing impaired. However, there is a lack of evidence-based bimodal fitting method or guideline clinically applicable. In general, bimodal fitting consists of three areas of fitting; cochlear implant mapping, hearing aid fitting, and bimodal fitting. Among them, establishing evidence-based bimodal hearing aid fitting will be a key factor to improve bimodal fitting services of hearing aid centers. The purpose of this study is to review recent literature related to bimodal hearing aid fitting to establish bimodal hearing aid fitting guidelines for hearing aid professionals. This review involves five bimodal hearing aid fitting issues including frequency response, loudness, dynamic compression, frequency transposition, and hearing aid fitting formula. In addition, it described three recent clinical bimodal hearing aid fitting guidelines provided by manufacturers. Two factors contributing to bimodal hearing aid fitting optimization across the studies are frequency response and loudness balance. Conducting loudness balance in bimodal hearing devices and adjusting gains based on the loudness balance tests are one of the important procedures for bimodal fitting optimization. Hearing aid fitting based on wide-band frequency responses and conventional hearing aid formula is recommended for the initial step of the bimodal fitting.

Abstract

Bimodal hearing refers to the combination of the cochlear implant in one ear and the hearing aid in the opposite ear. Recently, cochlear implant candidacy criteria have been expanded and the number of people receiving cochlear implants with residual hearing has increased, leading to the raised bimodal population. Providing professional fitting services based on systematic bimodal fitting methods or guidelines are essential for the successful bimodal intervention of the hearing impaired. However, there is a lack of evidence-based bimodal fitting method or guideline clinically applicable. In general, bimodal fitting consists of three areas of fitting; cochlear implant mapping, hearing aid fitting, and bimodal fitting. Among them, establishing evidence-based bimodal hearing aid fitting will be a key factor to improve bimodal fitting services of hearing aid centers. The purpose of this study is to review recent literature related to bimodal hearing aid fitting to establish bimodal hearing aid fitting guidelines for hearing aid professionals. This review involves five bimodal hearing aid fitting issues including frequency response, loudness, dynamic compression, frequency transposition, and hearing aid fitting formula. In addition, it described three recent clinical bimodal hearing aid fitting guidelines provided by manufacturers. Two factors contributing to bimodal hearing aid fitting optimization across the studies are frequency response and loudness balance. Conducting loudness balance in bimodal hearing devices and adjusting gains based on the loudness balance tests are one of the important procedures for bimodal fitting optimization. Hearing aid fitting based on wide-band frequency responses and conventional hearing aid formula is recommended for the initial step of the bimodal fitting.

참고문헌 (38)

  1. "Bimodal Fitting Quick Guide" (2014)[URL]
  2. Ching, T. Y., "The effect of auditory experience on speech perception, localization, and functional performance of children who use a cochlear implant and a hearing aid in opposite ears", International Journal of Audiology, 44(12), pp.677~690 (2005)
  3. Ching, T. Y., "Binaural-bimodal fitting or bilateral implantation for managing severe to profound deafness : A review", Trends in Amplification, 11(3), pp.161~192 (2007)
  4. "Bimodal Hearing. A Guide to Fitting" (2012)[URL]
  5. Davidson, L. S., "Evaluation of hearing aid frequency response fittings in pediatric and young adult bimodal recipients", Journal of the American Academy of Audiology, 26(4), pp.393~407 (2015)
  6. Dorman, M. F., "The benefits of combining acoustic and electric stimulation for the recognition of speech, voice and melodies", Audiology and Neurotology, 13(2), pp.105~112 (2008)
  7. Dowell, R., "Cochlear implantation: Optimizing outcomes through evidence-based clinical decisions", International Journal of Audiology, 55(Suppl 2), pp.S1~S2 (2016)
  8. Francart, T., "Psychophysics, fitting, and signal processing for combined hearing aid and cochlear implant stimulation", Ear and Hearing, 34(6), pp.685~700 (2013)
  9. René H. Gifford, "Bimodal Hearing or Bilateral Cochlear Implants? Ask the Patient", Ear and Hearing, 40(3), pp.501~516 (2019), Ovid Technologies (Wolters Kluwer Health)[DOI]
  10. Hua, H., "Cochlear implant combined with a linear frequency transposing hearing aid", Journal of the American Academy of Audiology, 23(9), pp.722~732 (2012)
  11. Keilmann, A. M., "Cochlear implant and hearing aid : A new approach to optimizing the fitting in this bimodal situation", European Archives of Oto-Rhino-Laryngology, 266(12), pp.1879~1884 (2009)
  12. Kong, Y. Y., "Cross-frequency integration for consonant and vowel identification in bimodal hearing", Journal of Speech, Language, and Hearing Research, 54(3), pp.959~980 (2011)
  13. Kong, Y. Y., "Speech and melody recognition in binaurally combined acoustic and electric hearing", The Journal of the Acoustical Society of America, 117(3 Pt 1), pp.1351~1361 (2005)
  14. Lee, J. H., "Bimodal fitting with a cochlear implant and a hearing aid in the opposite ear", Audiology and Speech Research, 1(1), pp.14~18 (2005)
  15. Leigh, J. R., "Evidence-based guidelines for recommending cochlear implantation for postlingually deafened adults", International Journal of Audiology, 55(Suppl 2), pp.S3~S8 (2016)
  16. McDermott, H., "Benefits of combined acoustic and electric hearing for music and pitch perception", Seminars in Hearing, 32(1), pp.103~114 (2011)
  17. McDermott, H., "The use of frequency compression by cochlear implant recipients with postoperative acoustic hearing", Journal of the American Academy of Audiology, 21(6), pp.380~389 (2010)
  18. Messersmith, J. J., "Reduction in highfrequency hearing aid gain can improve performance in patients with contralateral cochlear implant : A pilot study", American Journal of Audiology, 24(4), pp.462~468 (2015)
  19. Neuman, A. C., "Effect of hearing aid bandwidth on speech recognition performance of listeners using a cochlear implant and contralateral hearing aid(bimodal hearing)", Ear and Hearing, 34(5), pp.553~561 (2013)
  20. Nittrouer, S., "Emergent literacy in kindergartners with cochlear implants", Ear and Hearing, 33(6), pp.683~697 (2012)
  21. Oh, S. H., "Top-down processes in simulated electric-acoustic hearing : The effect of linguistic context on bimodal benefit for temporally interrupted speech", Ear and Hearing, 37(5), pp.582~592 (2016)
  22. 오수희, "인공와우를 통한 말지각과 반대편 귀 사용 효과에 대한 최근 연구 고찰", Audiology and Speech Research, 10(2), pp.87~98 (2014), 한국청각언어재활학회
  23. "Bimodal Hearing Aid Fitting Guidelines" (2016)[URL]
  24. Park, L. R., "Effects of frequency compression hearing aids for unilaterally implanted children with acoustically amplified residual hearing in the nonimplanted ear", Ear and Hearing, 33(4), pp.e1~e12 (2012)
  25. Perreau, A. E., "The contribution of a frequency-compression hearing aid to contralateral cochlear implant performance", Journal of the American Academy of Audiology, 24(2), pp.105~120 (2013)
  26. Potts, L. G., "Recognition and localization of speech by adult cochlear implant recipients wearing a digital hearing aid in the nonimplanted ear(bimodal hearing)", Journal of the American Academy of Audiology, 20(6), pp.353~373 (2009)
  27. Schafer, E. C., "A meta-analysis to compare speech recognition in noise with bilateral cochlear implants and bimodal stimulation", International Journal of Audiology, 50(12), pp.871~880 (2011)
  28. Siburt, H. W., "Bimodal programming : A survey of current clinical practice", American Journal of Audiology, 24(2), pp.243~249 (2015)
  29. Sucher, C. M., "Bimodal stimulation: Benefits for music perception and sound quality", Cochlear Implants International, 10(Suppl 1), pp.96~99 (2009)
  30. Duo-Duo Tao, "Bilaterally Combined Electric and Acoustic Hearing in Mandarin-Speaking Listeners: The Population With Poor Residual Hearing", Trends in Hearing, 22, pp.233121651875789~ (2018), SAGE Publications[DOI]
  31. Vermeire, K., "The influence of different speech processor and hearing aid settings on speech perception outcomes in electric acoustic stimulation patients", Ear and Hearing, 29(1), pp.76~86 (2008)
  32. Veugen, L. C., "Matching automatic gain control across devices in bimodal cochlear implant users", Ear and Hearing, 37(3), pp.260~270 (2016)
  33. Veugen, L. C., "Frequency-dependent loudness balancing in bimodal cochlear implant users", Acta Oto-Laryngologica, 136(8), pp.775~781 (2016)
  34. Vroegop, J. L., "How to optimally fit a hearing aid for bimodal cochlear implant users : A systematic review", Ear and Hearing, 39(6), pp.1039~1045 (2018)
  35. Vroegop, J. L., "Comparing two hearing aid fitting algorithms for bimodal cochlear implant users", Ear and Hearing, 40(1), pp.98~106 (2019)
  36. Vroegop, J. L., "Comparing the effect of different hearing aid fitting methods in bimodal cochlear implant users", American Journal of Audiology, 28(1), pp.1~10 (2019)
  37. Warren, S. E., "Bimodal hearing in individuals with severe-to-profound hearing loss: Benefits, challenges, and management", Seminars in Hearing, 39(4), pp.405~413 (2018)
  38. Zhang, T., "Information from the voice fundamental frequency(F0)region accounts for the majority of the benefit when acoustic stimulation is added to electric stimulation", Ear and Hearing, 31(1), pp.63~69 (2010)