Automated blood culture systems are widely used in clinical microbiology laboratories to minimize the workload of laboratory personnel and permit fast turnaround times. However, sometimes false positive signals occur due to leukocytosis, presence of fastidious bacteria, or unexplained causes. We experienced false positive signal in a patient with hyperleukocytosis for the first time since the automated blood culture system was introduced in our hospital over 20 years ago. We present two case reports with literature review and describe the procedure for dealing with false-positive cases in our hospital.
Abstract
Automated blood culture systems are widely used in clinical microbiology laboratories to minimize the workload of laboratory personnel and permit fast turnaround times. However, sometimes false positive signals occur due to leukocytosis, presence of fastidious bacteria, or unexplained causes. We experienced false positive signal in a patient with hyperleukocytosis for the first time since the automated blood culture system was introduced in our hospital over 20 years ago. We present two case reports with literature review and describe the procedure for dealing with false-positive cases in our hospital.
참고문헌 (7)
- Opota O, "Blood culture-based diagnosis of bacteraemia : state of the art", Clin Microbiol Infect, 21, pp.313~322 (2015)
- Ruiz-Giardin JM, "Diagnosis of bacteraemia and growth times", Int J Infect Dis, 41, pp.6~10 (2015)
- Qian Q, "Direct identification of bacteria from positive blood cultures by amplification and sequencing of the 16s rRNA gene : evaluation of BACTEC 9240 instrument true-positive and false-positive results", J Clin Microbiol, 39, pp.3578~3582 (2001)
- Khan M, "Increased peripheral leukemia blasts leading to false-positive blood culture", Blood Cells Mol Dis, 64, pp.8~9 (2017)
- Ebihara Y, "False-positive blood culture results in patients with hematologic malignancies", J Infect Chemother, 25, pp.404~406 (2019)
- Martinez RM, "An infrequent cause of falsepositive blood cultures", Clin Microbiol Newsl, 15, pp.7~8 (1993)
- Ziegler R, "Controlled clinical laboratory comparison of two supplemented aerobic and anaerobic media used in automated blood culture systems to detect bloodstream infections", J Clin Microbiol, 36, pp.657~661 (1998)