본문 바로가기
KERD
논문KCI 등재

치과용 세라믹 임플란트 지대주의 기계적 특성 및 절단면 평가

Mechanical Properties and Cross-sectional Surface Evaluation of Dental Ceramic Abutment

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류기계공학

초록

The purpose of this study is to assess the mechanical properties of the ceramic abutment with washer. In this study, ceramic abutment were used, tested with 30 ° compression load, shear fatigue, adaptation accuracy test(rotation angle, contact interval), removal torque force test, torsional breaking force test. The 30° compression load was 729 N, the shear fatigue load was 275 N, adaptation accuracy test of rotation angle was within 3°, contact interval within 10 μm, and removal torque force test value is 18.88 N·㎝, torsional breaking force test value is 35.52 N·㎝. Ceramic abutment with a washer fitted have sufficient mechanical strength and may be substituted for titanium abutment.

Abstract

The purpose of this study is to assess the mechanical properties of the ceramic abutment with washer. In this study, ceramic abutment were used, tested with 30 ° compression load, shear fatigue, adaptation accuracy test(rotation angle, contact interval), removal torque force test, torsional breaking force test. The 30° compression load was 729 N, the shear fatigue load was 275 N, adaptation accuracy test of rotation angle was within 3°, contact interval within 10 μm, and removal torque force test value is 18.88 N·㎝, torsional breaking force test value is 35.52 N·㎝. Ceramic abutment with a washer fitted have sufficient mechanical strength and may be substituted for titanium abutment.

참고문헌 (17)

  1. 송영균, "세라믹과 지르코니아의 골유착에 관한 고찰", 구강회복응용과학지, 28(3), pp.319~326 (2012), 대한턱관절교합학회
  2. 정용훈, "치과 임플란트 고정체의 여러 가지 제조공정과정에 따른 표면특성", 한국표면공학회지, 43(1), pp.17~24 (2010), 한국표면공학회
  3. 신성애, "지르코니아 임플란트 지대주의 기계적 강도에 관한 연구", 구강회복응용과학지, 25(4), pp.349~360 (2009), 대한턱관절교합학회
  4. 김태정, "레이저 처리후 임프란트 표면 변화에 관한 연구", Journal of Periodontal & Implant Science, 32(3), pp.615~630 (2002), 대한치주과학회
  5. 김태수, "파절된 임플란트 나사의 제거: 증례보고", 구강회복응용과학지, 31(1), pp.60~66 (2015), 대한턱관절교합학회[DOI]
  6. 배병룡, "임플랜트 지대주 종류가 나사풀림력과 연결부의 안정성에 미치는 영향", 구강회복응용과학지, 26(2), pp.97~109 (2010), 대한턱관절교합학회
  7. J Bickford, "An introduction of the design and behavior of bolted joints", pp.515~564 (1995), Marcel Dekker
  8. A Versluis, "Numerical analysis of a dental implant system preload with a washer", Int J Oral Maxillofac Implants, 14, pp.337~341 (1999)
  9. TW Korioth, "Effect of washers on reverse torque displacement of dental implant gold retaining screws", J Prosthet Dent, 82, pp.312~316 (1999)
  10. 황갑운, "Ti-6Al-4V 합금 미니 임플란트에 마찰고정한 경사 지대주의 피로특성", 대한치과재료학회지, 35(4), pp.359~368 (2008), 대한치과재료학회
  11. WC Marthin, "Implant abutment screw retations and preloads for four different screw matrials and surfaces", J Prosthet Dent, 86, pp.24~32 (2001)
  12. DG Gratton, "Micromotion and dynamic fatigue properties of the dental implantabutment interface", J Prosthet Dent, 85, pp.47~52 (2001)
  13. 신현모, "연결부 형태가 다른 세가지 임플랜트 시스템에서 지대주 디자인 및 조임회전력의 크기가 나사 풀림회전력에 미치는 영향", 대한치과재료학회지, 35(2), pp.121~132 (2008), 대한치과재료학회
  14. 정다운, "Effects of TiN Coating on the Fatigue Fracture of Dental Implant System with Various Cyclic Loads", 한국표면공학회지, 48(6), pp.283~291 (2015), 한국표면공학회[DOI]
  15. A Versluis, "Numerical analysis of a dental implant system preloaded with a washer", Int J Oral Maxillofac Implants, 14, pp.337~341 (1999)
  16. TW Korioth, "Effect of washers on reverse torque displacement of dental implant gold retaining screws", J Prosthet Dent, 82, pp.312~316 (1999)
  17. O Robert, "Standard handbook of fastening and joining", ~ (1997)