Effect of casting procedure on screw loosening of UCLA abutment in two implant-abutment connection systems

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류치의학

초록

STATEMENT OF PROBLEM: The cast abutment has advantages of overcoming angulation problem and esthetic problem. However, when a gold-machined UCLA abutment undergoes casting, the abutment surfaces in contact with the implant may change. PURPOSE: The purpose of this study was to compare the detorque values of prefabricated machined abutments with gold-premachined cast-on UCLA abutments before and after casting in two types of internal implant-abutment connection systems: (1) internal hexagonal joint, (2) internal octagonal joint. Furthermore, the detorque values of two implant-abutment connection systems were compared. MATERIALS AND METHODS: Twenty internal hexagonal implants with an 11-degree taper and twenty internal octagonal implants with an 8-degree taper were acquired. Ten prefabricated titanium abutments and ten gold-premachined UCLA abutments were used for each systems. Each abutment was torqued to 30 N㎝ according to the manufacturer’s instructions and detorque value was recorded. The detorque values were measured once more, after casting with gold alloy for UCLA abutment, and preparation for titanium abutments. Group means were calculated and compared using independent t-test and paired t-test (α=0.05). RESULTS: The results were as follows: 1. The detorque values between titanium abutments and UCLA-type abutments showed significant differences in internal octagonal implants (P〈0.05), not in internal hexagonal implants (P〉0.05). 2. In comparison of internal hexagonal and octagonal implants, the detorque values of titanium abutments had significant differences between two connection systems on the initial analysis (P〈0.05), not on the second analysis (P〉0.05) and the detorque values of UCLA-type abutments were not significantly different between two connection systems (P〉0.05). 3. The detorque values of titanium abutments and UCLA-type abutments decreased significantly on the second analysis than the initial analysis in internal hexagonal implants (P 〈0.05), not in internal octagonal implants (P〉0.05). CONCLUSION: Casting procedures of UCLA-type abutments had no significant effect on screw loosening in internal implant-abutment connection systems, and UCLA-type abutments showed higher detorque values than titanium abutments in internal octagonal implants.

Abstract

STATEMENT OF PROBLEM: The cast abutment has advantages of overcoming angulation problem and esthetic problem. However, when a gold-machined UCLA abutment undergoes casting, the abutment surfaces in contact with the implant may change. PURPOSE: The purpose of this study was to compare the detorque values of prefabricated machined abutments with gold-premachined cast-on UCLA abutments before and after casting in two types of internal implant-abutment connection systems: (1) internal hexagonal joint, (2) internal octagonal joint. Furthermore, the detorque values of two implant-abutment connection systems were compared. MATERIALS AND METHODS: Twenty internal hexagonal implants with an 11-degree taper and twenty internal octagonal implants with an 8-degree taper were acquired. Ten prefabricated titanium abutments and ten gold-premachined UCLA abutments were used for each systems. Each abutment was torqued to 30 N㎝ according to the manufacturer’s instructions and detorque value was recorded. The detorque values were measured once more, after casting with gold alloy for UCLA abutment, and preparation for titanium abutments. Group means were calculated and compared using independent t-test and paired t-test (α=0.05). RESULTS: The results were as follows: 1. The detorque values between titanium abutments and UCLA-type abutments showed significant differences in internal octagonal implants (P〈0.05), not in internal hexagonal implants (P〉0.05). 2. In comparison of internal hexagonal and octagonal implants, the detorque values of titanium abutments had significant differences between two connection systems on the initial analysis (P〈0.05), not on the second analysis (P〉0.05) and the detorque values of UCLA-type abutments were not significantly different between two connection systems (P〉0.05). 3. The detorque values of titanium abutments and UCLA-type abutments decreased significantly on the second analysis than the initial analysis in internal hexagonal implants (P 〈0.05), not in internal octagonal implants (P〉0.05). CONCLUSION: Casting procedures of UCLA-type abutments had no significant effect on screw loosening in internal implant-abutment connection systems, and UCLA-type abutments showed higher detorque values than titanium abutments in internal octagonal implants.

참고문헌 (22)

  1. Branemark PI, "Osseointegrated implants in the treatment of edentulous jaw. Experience from a 10-year period", Scand J Plastic Reconstr Surg Suppl, 16, pp.1~132 (1977)
  2. Adell R, "Long-term follow-up of osseointegrated implants in the treatment of totally edentulous jaws", Int J Oral Maxillofac Implants, 5, pp.347~359 (1990)
  3. Jemt T, "Failures and complications in 127 consecutively placed fixed partial prostheses supported by Branemark implants: from prosthetic treatment to first annual checkup", Int J Oral Maxillofac Implants, 7, pp.40~44 (1992)
  4. Jemt T, "3-year follow up study of early single implant restorations ad modum Branemark", Int J Periodontics Restorative Dent, 10, pp.340~349 (1990)
  5. Scheller H, "A 5-year multicenter study on implant-supported single crown restorations", Int J Oral Maxillofac Implants, 13, pp.212~218 (1998)
  6. Jemt T, "A 3-year follow-up study on single implant treatment", J Dent, 21, pp.203~208 (1993)
  7. McGlumphy EA, "Implant screw mechanics", Dent Clin N Am, 42, pp.71~89 (1998)
  8. Lan LA, "The effect of the use of a countertorque device on the abutment-implant complex", J Prosthet Dent, 81, pp.411~417 (1999)
  9. Lewis S, "The “UCLA” abutment", Int J Oral Maxillofac Implants, 3, pp.183~189 (1988)
  10. Lewis S, "The restoration of improperly inclined osseointegrated implants", Int J Oral Maxillofac Implants, 4, pp.147~152 (1989)
  11. Quirynen M, "The influence of titanium abutment surface roughness on plaque accumulation and gingivitis: short-term observations", Int J Oral Maxillofac Implants, 11, pp.169~178 (1996)
  12. Burguete RL, "Tightening characteristics for screwed joints in osseointegrated dental implants", J Prosthet Dent, 71, pp.592~599 (1994)
  13. Shigley JE, "Standard Handbook of Machine Design" (1986), McGraw-Hill
  14. Norton MR, "Assessment of cold welding properties of the internal conical interface of two commercially available implant systems", J Prosthet Dent, 81, pp.159~166 (1999)
  15. Merz BR, "Mechanics of the implant-abutment connection: an 8-degree taper compared to a butt joint connection", Int J Oral Maxillofac Implants, 15, pp.519~526 (2000)
  16. Cehreli MC, "Dynamic fatigue resistance of implant abutment junction in an internally notched morse-taper oral implant: influence of abutment design", Clin Oral Impl Res, 15, pp.459~465 (2004)
  17. Tan KB, "Implant-abutment screw joint preload of 7 hex-top abutment systems", Int J Oral Maxillofac Implants, 16, pp.367~377 (2001)
  18. Byrne D, "Preloads generated with repeated tightening in three types of screws used in dental implant assemblies", J Prosthodont, 15, pp.164~171 (2006)
  19. Kano SC, "Effect of casting procedures on screw loosening in UCLA-type abutments", J Prosthodont, 15, pp.77~81 (2006)
  20. Carr AB, "Characterization of noble metal implant cylinders: as-received cylinders and cast interfaces with noble metal alloys", J Prosthet Dent, 75, pp.77~85 (1996)
  21. Byrne D, "The fit of cast and premachined implant abutments", J Prosthet Dent, 80, pp.184~192 (1998)
  22. Vigolo P, "Measurement of the dimensions and abutment rotational freedom of gold-machined 3i UCLA-type abutments in the as-received condition, after casting with a noble metal alloy and porcelain firing", J Prosthet Dent, 84, pp.548~553 (2000)

관련 논문AI추천

이 간행물의 다른 논문 (9건)