Harmonic versus LigaSure for Flap Creation in Bilateral Axillary Breast Approach Thyroid Surgery
2
조회
0
다운로드
1
인용
0
참고문헌
키워드
초록
Purpose: The bilateral axillary breast approach (BABA) is frequently utilized for remote access thyroid surgery. The first step in this surgical approach is to make a subcutaneous flap from the axilla to the neck, usually using a Harmonic ACE. Our institution has utilized LigaSure to create these flaps. This study reports comparative results with these 2 energy devices. Methods: Medical data and video clips from 73 patients who underwent robotic BABA thyroid surgery were retrospectively reviewed. Flaps for BABA were created by first and second year endocrine fellows under the guidance of an experienced endocrine surgeon. There were no criteria for allocation to energy devices. Results: Of the 73 patients, 31 and 42 underwent flap creation with the Harmonic ACE and LigaSure devices, respectively. Demographic characteristics were similar in the 2 groups. Mean time for flap creation was significantly shorter (33.46±8.33 vs. 38.27±7.14 min, P=0.012), and mean number of camera cleanings significantly lower (4.21±2.53 vs. 10.45±4.95, P<0.001), in the LigaSure group. Surgical site pain on day 2 was significantly lower in the LigaSure group. Postoperative bleeding and skin burn occurred only in patients in the Harmonic group. Conclusion: LigaSure showed better performance than the Harmonic ACE device for BABA flap creation, as shown by time for flap creation, postoperative pain scores, and complications. Because this study was a small sized retrospective comparison, large randomized controlled trials are needed to confirm these results.
Abstract
Purpose: The bilateral axillary breast approach (BABA) is frequently utilized for remote access thyroid surgery. The first step in this surgical approach is to make a subcutaneous flap from the axilla to the neck, usually using a Harmonic ACE. Our institution has utilized LigaSure to create these flaps. This study reports comparative results with these 2 energy devices. Methods: Medical data and video clips from 73 patients who underwent robotic BABA thyroid surgery were retrospectively reviewed. Flaps for BABA were created by first and second year endocrine fellows under the guidance of an experienced endocrine surgeon. There were no criteria for allocation to energy devices. Results: Of the 73 patients, 31 and 42 underwent flap creation with the Harmonic ACE and LigaSure devices, respectively. Demographic characteristics were similar in the 2 groups. Mean time for flap creation was significantly shorter (33.46±8.33 vs. 38.27±7.14 min, P=0.012), and mean number of camera cleanings significantly lower (4.21±2.53 vs. 10.45±4.95, P<0.001), in the LigaSure group. Surgical site pain on day 2 was significantly lower in the LigaSure group. Postoperative bleeding and skin burn occurred only in patients in the Harmonic group. Conclusion: LigaSure showed better performance than the Harmonic ACE device for BABA flap creation, as shown by time for flap creation, postoperative pain scores, and complications. Because this study was a small sized retrospective comparison, large randomized controlled trials are needed to confirm these results.
참고문헌 (24)
- 정규원, "Cancer Statistics in Korea: Incidence, Mortality, Survival, and Prevalence in 2016", Cancer Research and Treatment, 51(2), pp.417~430 (2019), 대한암학회
- Choe JH, "Endoscopic thyroidectomy using a new bilateral axillo-breast approach", World J Surg, 31, pp.601~606 (2007)
- Lee KE, "Endoscopic thyroidectomy with the da Vinci robot system using the bilateral axillary breast approach(BABA)technique : our initial experience", Surg Laparosc Endosc Percutan Tech, 19, pp.e71~e75 (2009)
- 태경, "Robotic and Endoscopic Thyroid Surgery: Evolution and Advances", Clinical and Experimental Otorhinolaryngology, 12(1), pp.1~11 (2019), 대한이비인후과학회[DOI]
- Lee KE, "Outcomes of 109 patients with papillary thyroid carcinoma who underwent robotic total thyroidectomy with central node dissection via the bilateral axillo-breast approach", Surgery, 148, pp.1207~1213 (2010)
- Lee J, "Differences in postoperative outcomes, function, and cosmesis : open versus robotic thyroidectomy", Surg Endosc, 24, pp.3186~3194 (2010)
- Lee J, "Perioperative clinical outcomes after robotic thyroidectomy for thyroid carcinoma : a multicenter study", Surg Endosc, 25, pp.906~912 (2011)
- Song CM, "Factors affecting operative time in robotic thyroidectomy", Head Neck, 40, pp.893~903 (2018)
- Choi JY, "Endoscopic thyroidectomy via bilateral axillo-breast approach(BABA) : review of 512 cases in a single institute", Surg Endosc, 26, pp.948~955 (2012)
- Kim SJ, "Recovery of sensation in the anterior chest area after bilateral axillo-breast approach endoscopic", robotic thyroidectomy, 2011, pp.21, 366~371
- Parker DJ, "Use of the harmonic scalpel in thyroidectomy", ANZ J Surg, 79, pp.476~480 (2009)
- Kim FJ, "Temperature safety profile of laparoscopic devices : harmonic ACE(ACE), Ligasure V(LV), and plasma trisector(PT)", Surg Endosc, 22, pp.1464~1469 (2008)
- Koch C, "Determination of temperature elevation in tissue during the application of the harmonic scalpel", Ultrasound Med Biol, 29, pp.301~309 (2003)
- 윤혜정, "Initial Experience of BABA Robotic Thyroidectomy Using the Da Vinci Xi System in Incheon, Korea", Journal of Endocrine Surgery, 19(3), pp.59~67 (2019), 대한내분비외과학회[DOI]
- Youssef T, "Thyroid surgery: use of the LigaSure Vessel Sealing System versus conventional knot tying", Int J Surg, 6, pp.323~327 (2008)
- 채영준, "Comparative study of bilateral axillo-breast approach endoscopic and robotic thyroidectomy: propensity score matching analysis of large multi-institutional data", Annals of Surgical Treatment and Research, 98(6), pp.307~314 (2020), 대한외과학회
- Liu SY, "Bilateral axillo-breast approach robotic thyroidectomy : review of evidences", Gland Surg, 6, pp.250~257 (2017)
- Tae K, "Oncologic outcomes of robotic thyroidectomy : 5-year experience with propensity score matching", Surg Endosc, 30, pp.4785~4792 (2016)
- Sung TY, "Oncologic safety of robot thyroid surgery for papillary thyroid carcinoma : a comparative study of robot versus open thyroid surgery using inverse probability of treatment weighting", PLoS One, 11, pp.e0157345~ (2016)
- Tae K, "Robotic thyroidectomy by a gasless unilateral axillobreast or axillary approach : our early experiences", Surg Endosc, 25, pp.221~228 (2011)
- Lee JH, "Preoperative flap-site injection with ropivacaine and epinephrine in BABA robotic and endoscopic thyroidectomy safely reduces postoperative pain: A CONSORT-compliant double-blinded randomized controlled study (PAIN-BREKOR trial)", Medicine (Baltimore), 96, pp.e6896~ (2017)
- Bae DS, "Prospective, randomized controlled trial on use of ropivacaine after robotic thyroid surgery: Effects on postoperative pain", Head Neck, 38(Suppl 1), pp.E588~E593 (2016)
- Rahbari R, "Prospective randomized trial of ligasure versus harmonic hemostasis technique in thyroidectomy", Ann Surg Oncol, 18, pp.1023~1027 (2011)
- Zarebczan B, "A Comparison of the LigaSure and harmonic scalpel in thyroid surgery : a single institution review", Ann Surg Oncol, 18, pp.214~218 (2011)