초록
Purpose: Energy devices, such as Harmonic Focus Plus (Ethicon; ultrasonic scalpel; Harmonic) and Ligasure Small Jaw (Covidien; bipolar sealer; Ligasure) are widely used in thyroid surgery as substitutes for Bovie (electrocautery) and sutures. They reduce operation time and peri-operative complications. Thunderbeat Open Fine Jaw (Olympus; Thunderbeat) is a recently developed energy device, which combines the functions of both ultrasonic scalpel and bipolar sealer. We prospectively evaluated the safety and feasibility of Thunderbeat during open thyroidectomy compared with Harmonic and Ligasure. Methods: From September to November 2016, patients with papillary thyroid cancer or benign thyroid disease who underwent unilateral or bilateral thyroidectomy at Samsung Medical Center were enrolled. Thyroidectomies were performed by a single surgeon who specializes in endocrine surgery and is experienced in using energy devices. The patients were randomly allocated to one of the 3 groups according to the energy device used during the operation: Harmonic, Ligasure, and Thunderbeat group. Results: No significant differences were detected among the 3 groups in intra-operative bleeding, drainage amount, incidence of transient hypocalcemia, and transient vocal cord palsy. The surgical time for thyroidectomy in the T group was shorter than in the other 2 groups (P=0.031). However, the H group showed reduced surgical time for central lymph node dissection. Conclusion: Among the 3 energy devices, Thunderbeat and Harmonic shortened the surgical time for thyroidectomy and central lymph node dissection, respectively. However, each energy device has its own strengths and weaknesses, and we recommend the selection of devices based on the respective features.
Abstract
Purpose: Energy devices, such as Harmonic Focus Plus (Ethicon; ultrasonic scalpel; Harmonic) and Ligasure Small Jaw (Covidien; bipolar sealer; Ligasure) are widely used in thyroid surgery as substitutes for Bovie (electrocautery) and sutures. They reduce operation time and peri-operative complications. Thunderbeat Open Fine Jaw (Olympus; Thunderbeat) is a recently developed energy device, which combines the functions of both ultrasonic scalpel and bipolar sealer. We prospectively evaluated the safety and feasibility of Thunderbeat during open thyroidectomy compared with Harmonic and Ligasure. Methods: From September to November 2016, patients with papillary thyroid cancer or benign thyroid disease who underwent unilateral or bilateral thyroidectomy at Samsung Medical Center were enrolled. Thyroidectomies were performed by a single surgeon who specializes in endocrine surgery and is experienced in using energy devices. The patients were randomly allocated to one of the 3 groups according to the energy device used during the operation: Harmonic, Ligasure, and Thunderbeat group. Results: No significant differences were detected among the 3 groups in intra-operative bleeding, drainage amount, incidence of transient hypocalcemia, and transient vocal cord palsy. The surgical time for thyroidectomy in the T group was shorter than in the other 2 groups (P=0.031). However, the H group showed reduced surgical time for central lymph node dissection. Conclusion: Among the 3 energy devices, Thunderbeat and Harmonic shortened the surgical time for thyroidectomy and central lymph node dissection, respectively. However, each energy device has its own strengths and weaknesses, and we recommend the selection of devices based on the respective features.
참고문헌 (16)
- Fagotti A, "Randomized study comparing use of THUNDERBEAT technology vs standard electrosurgery during laparoscopic radical hysterectomy and pelvic lymphadenectomy for gynecologic cancer", J Minim Invasive Gynecol, 21, pp.447~453 (2014)
- Obonna GC, "Differences between Thunderbeat, Ligasure and Harmonic scalpel energy system in minimally invasive surgery", World J Laparos Surg, 7, pp.41~44 (2014)
- Seehofer D, "Safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5-mm sealing and cutting instruments", Surg Endosc, 26, pp.2541~2549 (2012)
- Edge SB, "AJCC Cancer Staging Manual" (2010), Springer
- Lang BH, "Does using an energized device in open thyroidectomy reduce complications?", J Surg Res, 181, pp.e23~e29 (2013)
- Garas G, "Which hemostatic device in thyroid surgery? A network meta-analysis of surgical technologies", Thyroid, 23, pp.1138~1150 (2013)
- Ruggiero R, "Sutureless thyroidectomy", Int J Surg, 12(Suppl 1), pp.S189~S193 (2014)
- Alexiou K, "The use of new energy-based haemostatic methods in thyroid surgery : sutureless total thyroidectomy(ultrasonic coagulation and", or electro thermal bipolar vessel sealing system)versus classic suture ligation(clamp-and-tie)technique, 2015, pp.87, 92~96
- Teksoz S, "Sutureless thyroidectomy with energy-based devices : cerrahpasa experience", Updates Surg, 65, pp.301~307 (2013)
- Kwak HY, "Comparison of surgical outcomes between papillary thyroid cancer patients treated with the Harmonic ACE scalpel and LigaSure Precise instrument during conventional thyroidectomy : a single-blind prospective randomized controlled trial", J Surg Res, 187, pp.484~489 (2014)
- Dionigi G, "The safety of energy-based devices in open thyroidectomy : a prospective, randomised study comparing the LigaSure™(LF1212)and the Harmonic® FOCUS", Langenbecks Arch Surg, 397, pp.817~823 (2012)
- Van Slycke S, "Comparison of the Harmonic focus and the Thunderbeat for open thyroidectomy", Langenbecks Arch Surg, 401, pp.851~859 (2016)
- Smith CT, "Infrared thermographic profiles of vessel sealing devices on thyroid parenchyma", J Surg Res, 170, pp.64~68 (2011)
- Applewhite MK, "Ultrasonic, bipolar, and integrated energy devices : comparing heat spread in collateral tissues", J Surg Res, 207, pp.249~254 (2017)
- Kwak HY, "Thermal injury of the recurrent laryngeal nerve by THUNDERBEAT during thyroid surgery : findings from continuous intraoperative neuromonitoring in a porcine model", J Surg Res, 200, pp.177~182 (2016)
- Park MW, "Long-term voice outcome after thyroidectomy using energy based devices", Auris Nasus Larynx, 45, pp.527~532 (2018)