본문 바로가기
KERD
논문KCI 등재

Standardization Status of Total Cholesterol Concentration Measurement: Analysis of Korean External Quality Assessment Data

2

조회

0

다운로드

4

인용

0

참고문헌

키워드

주제분류병리학

초록

Background: Total cholesterol concentration measurement is important in the diagnosis of dyslipidemia and evaluation of cardiovascular disease risk factors. Measurement reliability for obtaining an accurate total cholesterol concentration requires procedure standardization. We evaluated the standardization status for total cholesterol concentration measurement through Korean external quality assessment (EQA) data analysis. Methods: This study involved 1,670 laboratories that participated in the EQA of total cholesterol concentration measurements in 2019 for 32 products from different manufacturers. The target concentrations of three quality control (QC) materials (samples A, B, and C) were measured using the reference method and compared with EQA data. The performance criteria for total cholesterol concentration measurement were based on the National Cholesterol Education Program guidelines, with ±3% inaccuracy. Results: The target values and inaccuracies of the QC material based on the reference method measurements were 254.65±7.64, 108.30±3.25, and 256.29±7.69 mg/dL (6.59±0.20, 2.80±0.08, and 6.63±0.20 mmol/L) for samples A, B, and C, respectively. The performance criteria were not met in 42.7% laboratories for sample A, 68.4% of laboratories for sample B, and 38.0% laboratories for sample C. Conclusions: Despite significant efforts to accurately measure total cholesterol concentrations, further actions are needed for measurement standardization. Manufacturers reporting values that differ from target values should check calibrator traceability; additional efforts to accurately measure total cholesterol concentrations are required for laboratories that use products from these manufacturers.

Abstract

Background: Total cholesterol concentration measurement is important in the diagnosis of dyslipidemia and evaluation of cardiovascular disease risk factors. Measurement reliability for obtaining an accurate total cholesterol concentration requires procedure standardization. We evaluated the standardization status for total cholesterol concentration measurement through Korean external quality assessment (EQA) data analysis. Methods: This study involved 1,670 laboratories that participated in the EQA of total cholesterol concentration measurements in 2019 for 32 products from different manufacturers. The target concentrations of three quality control (QC) materials (samples A, B, and C) were measured using the reference method and compared with EQA data. The performance criteria for total cholesterol concentration measurement were based on the National Cholesterol Education Program guidelines, with ±3% inaccuracy. Results: The target values and inaccuracies of the QC material based on the reference method measurements were 254.65±7.64, 108.30±3.25, and 256.29±7.69 mg/dL (6.59±0.20, 2.80±0.08, and 6.63±0.20 mmol/L) for samples A, B, and C, respectively. The performance criteria were not met in 42.7% laboratories for sample A, 68.4% of laboratories for sample B, and 38.0% laboratories for sample C. Conclusions: Despite significant efforts to accurately measure total cholesterol concentrations, further actions are needed for measurement standardization. Manufacturers reporting values that differ from target values should check calibrator traceability; additional efforts to accurately measure total cholesterol concentrations are required for laboratories that use products from these manufacturers.

참고문헌 (13)

  1. National Cholesterol Education Program (NCEP) Expert Panel on Detection, Evaluation, and Treatment of High Blood Cholesterol in Adults (Adult Treatment Panel III), "Third report of the National Cholesterol Education Program (NCEP) Expert Panel on Detection, Evaluation, and Treatment of High Blood Cholesterol in Adults (Adult Treatment Panel III) final report", Circulation, 106, pp.3143~3421 (2002)
  2. "Total cholesterol certification protocol for manufacturers"[URL]
  3. Miller WG, "The role of proficiency testing in achieving standardization and harmonization between laboratories", Clin Biochem, 42, pp.232~235 (2009)
  4. Kim Sollip, "Schemes and Performance Evaluation Criteria of Korean Association of External Quality Assessment (KEQAS) for Improving Laboratory Testing", Annals of Laboratory Medicine, 41(2), pp.230~239 (2021), 대한진단검사의학회[DOI]
  5. Kim H, "Status of quality control for laboratory tests of medical institutions in Korea: analysis of 10 years of data on external quality assessment participation", Healthcare (Basel), 8, pp.75~ (2020)
  6. Sundvall J, "Serum cholesterol during 27 years: assessment of systematic error and affecting factors and their role in interpreting population trends", Clin Chim Acta, 378, pp.93~98 (2007)
  7. Ross JW, "The accuracy of laboratory measurements in clinical chemistry: a study of 11 routine chemistry analytes in the College of American Pathologist Chemistry Survey with fresh frozen serum, definitive methods, and reference methods", Arch Pathol Lab Med, 122, pp.587~608 (1998)
  8. Ellerbe P, "A comparison of results for cholesterol in human serum obtained by the Reference Method and by the Definitive Method of the National Reference System for cholesterol", Clin Chem, 36, pp.370~375 (1990)
  9. Abel LL, "A simplified method for the estimation of total cholesterol in serum and demonstration of its specificity", J Biol Chem, 195, pp.357~366 (1952)
  10. Bowers GN, Jr, "Isotope dilution mass spectrometry and the National Reference System", Anal Chem, 65, pp.475R~479R (1993)
  11. Edwards SH, "Proposed serum cholesterol reference measurement procedure by gas chromatography-isotope dilution mass spectrometry", Clin Chem, 57, pp.614~622 (2011)
  12. Warnick GR, "Standardization of measurements for cholesterol, triglycerides, and major lipoproteins", Lab Med, 39, pp.481~490 (2008)
  13. Long Qichen, "Commutability Assessment of Candidate External Quality Assessment Materials for Aminotransferase Activity Measurements Based on Different Approaches in China", Annals of Laboratory Medicine, 41(1), pp.68~76 (2021), 대한진단검사의학회

관련 논문AI추천

Can Reference Materials Prepared Following CLSI C37-A Be Utilized Without Commutability Assessment? Perspectives Based on Lipid MeasurementsAn Dongheui, Cho Chan-Ik, Choi Qute, Kim Jeong-Ho, Kim Sollip, Kwon Gye Cheol, Lee Sang-Guk, Park Pil-Whan, Seo Jong Do, Song Junghan, Yun Yeo Min대한진단검사의학회 · Annals of Laboratory Medicine2025 · 제45권 제6호Schemes and Performance Evaluation Criteria of Korean Association of External Quality Assessment (KEQAS) for Improving Laboratory TestingChun Sail, Kim Sollip, Lee Kyunghoon, Lee Yong-Wha, Min Won-Ki, Park Hyung-Doo대한진단검사의학회 · Annals of Laboratory Medicine2021 · 제41권 제2호Performance evaluation of Barozen Lipid Plus for point-of-care testing of lipid profiles: a method comparison study선화연, 유병욱, Soojoung Yu영남대학교 의과대학 · Journal of Yeungnam Medical Science2023 · 제40권 A New Strategy for Evaluating the Quality of Laboratory Results for Big Data Research: Using External Quality Assessment Survey Data (2010–2020)김솔잎, 민원기, 박형두, 전사일, Cho Eun-Jung, Jeong Tae-Dong, Yun Yeo-Min대한진단검사의학회 · Annals of Laboratory Medicine2023 · 제43권 제5호Quantitative Evaluation of the Real-World Harmonization Status of Laboratory Test Items Using External Quality Assessment Data김솔잎, 송정한, 전사일, Cho Eun-Jung, Chung Jae-Woo, Jeong Tae-Dong, Lee Kyunghoon, Lee Seunghoo, Min Won-Ki대한진단검사의학회 · Annals of Laboratory Medicine2024 · 제44권 제6호Practical Considerations for Clinical Laboratories in Top-down Approach for Assessing the Measurement Uncertainty of Clinical Chemistry Analytes전사일, Gu Hyunjung, Hong Jinyoung, Lee Juhee, Lee Woochang, Lee Woo-In, Min Won-Ki, Yun Yeo-Min대한진단검사의학회 · Annals of Laboratory Medicine2022 · 제42권 제6호

이 간행물의 다른 논문 (15건)