초록

In this paper, we proposed a laser-based non-contact temperature measuring method for high speed rotating polycarbonate (PC) disk using transparency change of thermocolor. The thermocolor has abilities to change color and transparency due to a change in temperature. The thermocolor is applied on one side of polyvinylidene fluoride (PVDF) membrane. The thermocolor applied membrane is attached to inside of reaction chamber in disk. An optical system consisted of a laser beam radiator and a laser photometer is installed. Laser is irradiated at the bottom side of disk and the transmitted laser beam is detected by the laser photometer at the opposite side of disk. During the disk is rotating, laser is irradiated and detected simultaneously. The laser photometer senses the transmitted laser power and generates voltage as output. The temperature of disk can be detected during the disk is rotating up to 3000 RPM.

Abstract

In this paper, we proposed a laser-based non-contact temperature measuring method for high speed rotating polycarbonate (PC) disk using transparency change of thermocolor. The thermocolor has abilities to change color and transparency due to a change in temperature. The thermocolor is applied on one side of polyvinylidene fluoride (PVDF) membrane. The thermocolor applied membrane is attached to inside of reaction chamber in disk. An optical system consisted of a laser beam radiator and a laser photometer is installed. Laser is irradiated at the bottom side of disk and the transmitted laser beam is detected by the laser photometer at the opposite side of disk. During the disk is rotating, laser is irradiated and detected simultaneously. The laser photometer senses the transmitted laser power and generates voltage as output. The temperature of disk can be detected during the disk is rotating up to 3000 RPM.

참고문헌 (12)

  1. S. Haeberle, "Microfluidic platforms for lab-on-a-chip applications", Lab Chip, 1037(9), pp.1037~1110 (2007)
  2. D. C. Leslie, "Platinum nanoparticlefacilitated reflective surfaces for non-contact temperature control in microfluidic devices for PCR amplification", Lab Chip, 12(1), pp.127~132 (2012)
  3. P. J. Pinter, "Infrared thermometry: A remote sensing technique for predicting yield in water-stressed cotton", Agricultural Water Management, 6(4), pp.385~395 (1983)
  4. H. T. Lam, "A high-resolution non-contact fluorescencebased temperature sensor for neonatal care", Measurement Science and Technology, 23(23) (2012)
  5. M. Seaver, "Noncontact fluorescence thermometry of acoustically levitated waterdrops", Applied Optics, 29(33), pp.4956~4961 (1990)
  6. P. Neuzil, "Non-contact fluorescent bleachingindependent method for temperature measurement in microfluidic systems based on DNA melting curves", Lab Chip, 10(20), pp.2818~2821 (2010)
  7. D. R. Sabatino, "A high-accuracy calibration technique for thermochromic liquid crystal temperature measurements", Experiments in Fluids, 28(6), pp.479~505 (2000)
  8. T. Hieta, "High-precision diode-laser-based temperature measurement for air refractive index compensation", Applied Optics, 50(31), pp.5990~5998 (2011)
  9. J. P. Longtin, "Laser-based temperature measurement at a liquid surface", pp.119~123 (1998), IHTC
  10. I. P. Nagy, "Measurement of temperature profiles of chemical waves by thermocolor imaging", Chemical Physics Letters, 200(1), pp.147~152 (1992)
  11. K. Tashiro, "Structure and thermochromic solid-state phase transition of poly (3-alkyl-thiophene)", Journal of Polymer Science Part B: Polymer Physics, 29(10), pp.1223~1233 (1991)