원유 세균수 검사장비의 정도관리를 위한 표준시료의 개발 및 평가

Development and evaluation of standard samples for quality control of automated total bacterial counter in raw milk

1

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류수의학

초록

Standard samples were prepared for this study, and applied for BactoScan TM and BactoCount TM in order for quality control evaluation for total bacterial count in raw milk. Accordingly, the preparation of two lots of standard samples for quality control were lyophilized, which contain Lactobacillus lactis. The standard samples were prepared into three different levels of bacterial counts, those were Low 30,000∼ 40,000, Medium 70,000∼90,000, High 150,000∼220,000 CFU/mL, respectively. Then, the proficiency tests were performed in total 19 laboratories for measuring total bacterial counts. The total bacterial counts in the standard samples showed 37∼42, 82∼105, 214∼240 CFU/mL in the first lot, and it showed 30∼36, 67∼75, 131∼163 CFU/mL in the second lot in low, medium and high levels, respectively. Based on these results, the absolute values of z-scores of six standard samples in 18 laboratories were ≤2, which means the samples are satisfactory. However, z-score in one laboratory was ≤3, which means the sample is questionable. Using two standard samples, the correlation between BactoScan TM and BactoCount TM was 0.9982, which means the results of total bacterial count measurement of both equipment were equivalent. Therefore, the standard samples manufactured in this study for quality control of total bacterial count using BactoScan TM and BactoCount TM in the raw milk could be applied to proficiency tests.

Abstract

Standard samples were prepared for this study, and applied for BactoScan TM and BactoCount TM in order for quality control evaluation for total bacterial count in raw milk. Accordingly, the preparation of two lots of standard samples for quality control were lyophilized, which contain Lactobacillus lactis. The standard samples were prepared into three different levels of bacterial counts, those were Low 30,000∼ 40,000, Medium 70,000∼90,000, High 150,000∼220,000 CFU/mL, respectively. Then, the proficiency tests were performed in total 19 laboratories for measuring total bacterial counts. The total bacterial counts in the standard samples showed 37∼42, 82∼105, 214∼240 CFU/mL in the first lot, and it showed 30∼36, 67∼75, 131∼163 CFU/mL in the second lot in low, medium and high levels, respectively. Based on these results, the absolute values of z-scores of six standard samples in 18 laboratories were ≤2, which means the samples are satisfactory. However, z-score in one laboratory was ≤3, which means the sample is questionable. Using two standard samples, the correlation between BactoScan TM and BactoCount TM was 0.9982, which means the results of total bacterial count measurement of both equipment were equivalent. Therefore, the standard samples manufactured in this study for quality control of total bacterial count using BactoScan TM and BactoCount TM in the raw milk could be applied to proficiency tests.

참고문헌 (16)

  1. 공정시험법 세균항목의 정도관리 방안, "국립환경과학원", 49~ (2004)
  2. 원유검사 표준화 운영 실무 가이드라인, "농림축산검역본부", 43~52 (2019)
  3. 농림축산식품부, "낙농통계연감", pp.261~ (2016)
  4. 농림축산식품부, "원유검사 공영화 실시요령", 농림축산식품부 고시 제2016-146호, ~ (2016)
  5. Bolzoni G, "BactoScan FC : conversion system for results at the national level in Italy and reproducibility of total bacterial counting four years after implementation", Euroreference, 2, pp.46~54 (2017)
  6. Boucher SN, "Production and viablity of coccoid forms of Campylobacter jejuni", J Appl Bacteriol, 77, pp.303~307 (1994)
  7. Bulletin of the IDF, "Milk payment systems for ex-farm milk. No 3331", 1998, 1998
  8. NATHALIE DESMASURES, "Monitoring the microbiology of high quality milk by monthly sampling over 2 years", Journal of Dairy Research, 64(2), pp.271~280 (1997), Cambridge University Press (CUP)[DOI]
  9. Gunasekera TS, "A flow cytometry method for rapid detection and enumeration of total bacterial bacteria in milk", App Env Microbiol, 66, pp.1228~1232 (2000)[DOI]
  10. 김주영, "총대장균군 측정의 정도관리에 적합한 균주의 조제", 미생물학회지, 44(1), pp.43~48 (2008), 한국미생물학회
  11. Madden RH, "Determination of the BactoScan conversion Factor for the United Kingdom", Dairy microbiology, 70, pp.15~17 (2017)
  12. "REGULATION (EC) No 853", 2004 of THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 29 April 2004 laying down specific hygiene rules for the hygiene of foodstuffs
  13. Rollins DM, "Viable but nonculturable stage of Campylobacter jejuni. and its role in survival in the natural aquatic environment", Appl Environ Microbiol., 52, pp.531~538 (1986)
  14. Suhren G, "First experiences with automatic flow cytometric determination of total bacterial count in raw milk", Int Dairy Fed Bull, 50, pp.249~275 (2000)
  15. Walter S, "Application of a unified European conversion factor for the transformation of Bactoscan FC values", Food Agroscope Science, 36, pp.1~11 (2016)

관련 논문AI추천

국내 원유 세균수 검사장비 박토스캔 및 박토카운트의 보정식 재평가Dasom Park, ESeul Kim, Ha-Young Kim, Jin-Hwan Kim, Jin-San Moon, Soon-Seek Yoon, Yeong-Seob Byun한국동물위생학회 · 한국동물위생학회지2022 · 제45권 제1호Development of a Proficiency Test Specimen for Enumerating Escherichia coli in Molluscan Bivalve Shellfish권지영, 김영목, 김풍호, 목종수, 손광태, 송기철, 신순범, 오은경, 유홍식, 윤호동, 이가정, 임근식한국수산과학회 · Fisheries and Aquatic Sciences2011 · 제14권 제3호Comparative Evaluation of Culture Media for Quantification of Lactic Acid Bacteria in Various Dairy Products김해영, Gwak Yoon-Soo, Kim Eiseul, Kim Hyun-Jae, Kim Ik-Seon, Kwak Hyo-Sun, Lee Shin-Young한국미생물·생명공학회 · 한국미생물·생명공학회지2023 · 제51권 제1호Biodiversity and identification of spore-forming bacteria in raw milk from Korean dairy farm environment using the culturomics approachDaye Mun, Hyun Woo Choi, Ja Yeon Yoo, Jun-Sang Ham, Sangnam Oh, Won Seo Park, Younghoon Kim한국축산학회 · 한국축산학회지2025 · 제67권 제5호경기 남부지역 HACCP 인증 아이스크림 업체와 미인증 업체의 미생물 관리 수준 비교고태오, 박준조, 박혜원, 배진규, 복민순, 양호열, 전오숙한국동물위생학회 · 한국동물위생학회지2013 · 제36권 제2호농가생산 원유의 유질 및 위생진단에 관한 연구고문석, 김은주, 이조언, 이현종, 진신흠한국축산학회 · 한국축산학회지2002 · 제44권 제1호

이 간행물의 다른 논문 (10건)