User experience of MV2000-MT (SU:M2)® as a Mechanical Ventilator: A Comparative Clinical Study on Usability, Safety, and Medical Staff Satisfaction

0

조회

0

다운로드

1

인용

0

참고문헌

키워드

초록

In the present study, we aimed to demonstrate that MV2000-MT(SU:M2)® (MV, MEK-ICS, Paju, Korea), a domestic ventilator, is not inferior in terms of usability, safety, and medical staff satisfaction as compared to Hamilton G5 (G5, Hamilton Medical AG, Rhäzüns, Switzerland). A total of 39 patients who applied MV (group M) or G5 (group H) were included in the study sample. Usability was evaluated by the following factors: the number of alarm errors, replacement requirement of breathing circuit, replacement requirement of a right-angle connector, and ease of ventilator weaning. For safety evaluation, the number of ventilator replacements due to malfunction of the ventilator was evaluated. Items for medical staff satisfaction survey were as follows: the number of MV and G5 uses, hardware, and software assessment. In the usability evaluation, the replacement requirement of the right-angle connector was lower in Group M than in Group H (mean ± standard deviation, Group M: 7.39 ± 6.72, Group H: 14.19 ± 10.24, p = 0.021); however, the evaluations of other parts were not significantly different between the two groups. The number of ventilator replacements due to a malfunction of the ventilator did not differ between two groups. The number of MV and G5 uses was 3.0 [3.0-4.0] and 10.0 [5.0-10.0] (median [interquartile range], p < 0.001). Overall, the mean medical staff satisfaction score of Hamilton G5 was higher than that of MV2000-MT(SU:M2)®. The usability of MV is comparable to that of G5. However, medical staff satisfaction with Hamilton G5 was higher than that with MV2000-MT(SU:M2)®, and this difference could be due to the difference in the number of uses. In order to improve the penetration rate of the domestic mechanical ventilator, it is necessary to find ways to increase familiarity of medical staff with domestic mechanical ventilators.

Abstract

In the present study, we aimed to demonstrate that MV2000-MT(SU:M2)® (MV, MEK-ICS, Paju, Korea), a domestic ventilator, is not inferior in terms of usability, safety, and medical staff satisfaction as compared to Hamilton G5 (G5, Hamilton Medical AG, Rhäzüns, Switzerland). A total of 39 patients who applied MV (group M) or G5 (group H) were included in the study sample. Usability was evaluated by the following factors: the number of alarm errors, replacement requirement of breathing circuit, replacement requirement of a right-angle connector, and ease of ventilator weaning. For safety evaluation, the number of ventilator replacements due to malfunction of the ventilator was evaluated. Items for medical staff satisfaction survey were as follows: the number of MV and G5 uses, hardware, and software assessment. In the usability evaluation, the replacement requirement of the right-angle connector was lower in Group M than in Group H (mean ± standard deviation, Group M: 7.39 ± 6.72, Group H: 14.19 ± 10.24, p = 0.021); however, the evaluations of other parts were not significantly different between the two groups. The number of ventilator replacements due to a malfunction of the ventilator did not differ between two groups. The number of MV and G5 uses was 3.0 [3.0-4.0] and 10.0 [5.0-10.0] (median [interquartile range], p < 0.001). Overall, the mean medical staff satisfaction score of Hamilton G5 was higher than that of MV2000-MT(SU:M2)®. The usability of MV is comparable to that of G5. However, medical staff satisfaction with Hamilton G5 was higher than that with MV2000-MT(SU:M2)®, and this difference could be due to the difference in the number of uses. In order to improve the penetration rate of the domestic mechanical ventilator, it is necessary to find ways to increase familiarity of medical staff with domestic mechanical ventilators.

참고문헌 (6)

  1. Richard JCM, "ICU mechanical ventilators, technological advances vs. user friendliness: the right picture is worth a thousand numbers", Intensive Care Med, 35(10), pp.1662~1663 (2009)
  2. Soar J, "2019 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science With Treatment Recommendations: Summary From the Basic Life Support; Advanced Life Support;Pediatric Life Support; Neonatal Life Support; Education, Implementation, and Teams; and First Aid Task Forces", Circulation, 140(24), pp.e826~e880 (2019)
  3. Ventzke MM, "Cardiopulmonary resuscitation : when guidelines provide no answers", Anaesthesist, 68(4), pp.239~244 (2019)
  4. Ramsay MA, "Controlled sedation with alphaxalone-alphadolone", Br Med J, 2(5920), pp.656~659 (1974)
  5. Vignaux L, "Evaluation of the user-friendliness of seven new generation intensive care ventilators", Intensive Care Med, 35(10), pp.1687~1691 (2009)

관련 논문AI추천

Change in management and outcome of mechanical ventilation in Korea: a prospective observational study강형구, 김영삼, 김윤성, 리원연, 박성훈, 박주희, 서지영, 심윤수, 이광하, 이연주, 이진화, 임채만, 홍지영, Heung Bum Lee, Jae Kyeom Sim, Kyung Chan Kim, Sang-Min Lee, So Young Park, Won-Il Choi, Won Jun Song대한내과학회 · The Korean Journal of Internal Medicine2022 · 제37권 제3호국내 운반형인공호흡기 안전성 확보를 위한분류체계 및 평가기준 제안김지현, 조성구보건의료산업학회 · 보건의료산업학회지2017 · 제11권 제4호급성 폐 손상 및 급성 호흡곤란증후군 환자에서 Conventional Volume-Controlled Ventilation과 Adaptive Support Ventilation (ASV) 적용에 대한 비교고윤석, 임채만, 최익수, 최정은, 홍상범대한중환자의학회 · Acute and Critical Care2009 · 제24권 제2호Characteristics and management of mechanically ventilated patients in South Korea compared with other high-income Asian countries and regions강성욱, 고윤석, 곽상현, 김덕규, 김성현, 김진형, 남경훈, 박성훈, 박소정, 박소희, 박지수, 박진경, 박태선, 변기섭, 신윤미, 심윤수, 안지환, 이광하, 이보영, 이상민, 이송이, 임아영, 장항제, 전경만, 전상범, 정원재, 정윤정, 조영재, 조우현, 홍석경대한중환자의학회 · Acute and Critical Care2025 · 제40권 제3호사용적합성 평가를 적용한 국산 범용인공호흡기의 위험요인 분석권경민, 김승희, 김유림, 장원석대한의용생체공학회 · 의공학회지2024 · 제45권 제1호Association between mechanical power and intensive care unit mortality in Korean patients under pressure-controlled ventilation강형구, 김경찬, 김영삼, 김윤성, 리원연, 박성훈, 박소영, 박주희, 서지영, 송원준, 심재겸, 이광하, 이상민, 이연주, 이진화, 이흥범, 임채만, 최원일, 홍지영, Sim Yun Su대한중환자의학회 · Acute and Critical Care2024 · 제39권 제1호

이 간행물의 다른 논문 (5건)