좌우 비대칭형 난청 환자에서 CROS 보청기와 이식형 골도보청기의 선택 과정과 착용 전후 청력역치의 비교
Comparison of Aided and Unaided Thresholds and Selection Processes of Contralateral Routing of Signal Hearing Aids and Implantable Bone Conduction Devices in Patients with Asymmetric Hearing Loss .
2
조회
0
다운로드
1
인용
0
참고문헌
초록
Background and Objectives We reviewed the selection processes of contralateral routing of signal (CROS) hearing aids (HAs) and bone-conduction (BC) Has, and compared aided and unaided hearing thresholds. Subjects and Method Twenty-four patients with asymmetrical hearing loss who used BC HAs (n=12) and CROS HAs (n=12) were enrolled. The choice of two different HAs were compared with respect to the degree of hearing loss, the unaided hearing thresholds and functional gains. Results When the hearing thresholds of the better hearing ears were >30 dB HL, most (92%, 11 of 12) chose CROS rather than BC HAs, with significant difference (p=0.001). Both CROS and BC HAs groups showed significantly improved functional gains (46.6 dB and 53.4 dB, respectively). Aided air-conduction (AC) thresholds (40.2 dB HL) in the CROS group were similar to the AC thresholds (43.1 dB HL) of better hearing ears. However, the hearing threshold of Aided AC thresholds (35.8 dB HL) in BC HAs group were less than the BC thresholds (17.3 dB HL) of better hearing ears by 19 dB (p<0.001). Conclusion Both groups showed significantly increased functional gains. CROS HAs were preferred when hearing thresholds in better hearing ears were >30 dB HL. The CROS group showed aided thresholds similar to the thresholds of better hearing ears, but the BC HAs group showed poorer aided thresholds than the thresholds of better hearing ears. For patients with asymmetric hearing loss, HAs should be selected based on the degree and types of hearing loss and the maximum output level of the selected device.
Abstract
Background and Objectives We reviewed the selection processes of contralateral routing of signal (CROS) hearing aids (HAs) and bone-conduction (BC) Has, and compared aided and unaided hearing thresholds. Subjects and Method Twenty-four patients with asymmetrical hearing loss who used BC HAs (n=12) and CROS HAs (n=12) were enrolled. The choice of two different HAs were compared with respect to the degree of hearing loss, the unaided hearing thresholds and functional gains. Results When the hearing thresholds of the better hearing ears were >30 dB HL, most (92%, 11 of 12) chose CROS rather than BC HAs, with significant difference (p=0.001). Both CROS and BC HAs groups showed significantly improved functional gains (46.6 dB and 53.4 dB, respectively). Aided air-conduction (AC) thresholds (40.2 dB HL) in the CROS group were similar to the AC thresholds (43.1 dB HL) of better hearing ears. However, the hearing threshold of Aided AC thresholds (35.8 dB HL) in BC HAs group were less than the BC thresholds (17.3 dB HL) of better hearing ears by 19 dB (p<0.001). Conclusion Both groups showed significantly increased functional gains. CROS HAs were preferred when hearing thresholds in better hearing ears were >30 dB HL. The CROS group showed aided thresholds similar to the thresholds of better hearing ears, but the BC HAs group showed poorer aided thresholds than the thresholds of better hearing ears. For patients with asymmetric hearing loss, HAs should be selected based on the degree and types of hearing loss and the maximum output level of the selected device.
참고문헌 (22)
- Noble W, "Interaural asymmetry of hearing loss, Speech, Spatial and Qualities of Hearing Scale(SSQ)disabilities, and handicap", Int J Audiol, 43(2), pp.100~114 (2004)
- McLeod B, "Self reported hearing difficulties following excision of vestibular schwannoma", Int J Audiol, 47(7), pp.420~430 (2008)
- Niparko JK, "Comparison of the bone anchored hearing aid implantable hearing device with contralateral routing of offside signal amplification in the rehabilitation of unilateral deafness", Otol Neurotol, 24(1), pp.73~78 (2003)
- Baguley DM, "The evidence base for the application of contralateral bone anchored hearing aids in acquired unilateral sensorineural hearing loss in adults", Clin Otolaryngol, 31(1), pp.6~14 (2006)
- Vermeire K, "Binaural hearing after cochlear implantation in subjects with unilateral sensorineural deafness and tinnitus", Audiol Neurootol, 14(3), pp.163~171 (2009)
- van Zon A, "Cochlear implantation for patients with single-sided deafness or asymmetrical hearing loss : a systematic review of the evidence", Otol Neurotol, 36(2), pp.209~219 (2015)
- Kitterick PT, "Hearing instruments for unilateral severe-to-profound sensorineural hearing loss in adults : a systematic review and meta-analysis", Ear Hear, 37(5), pp.495~507 (2016)
- 한규희, "Bone Anchored Hearing Aid(BAHA)를 이용한 청각 재활", 대한이비인후-두경부외과학회지, 53(12), pp.755~760 (2010), 대한이비인후과학회
- 박미나, "일측성 전농 환자에서의 청력 재활: 바하와 크로스 보청기의 이득과 선택 기준", 대한이비인후-두경부외과학회지, 56(6), pp.339~345 (2013), 대한이비인후과학회[DOI]
- 이지혜, "Bone Anchored Hearing Aids(BAHA)의 주관적 만족도 및 청각학적 결과", 대한이비인후-두경부외과학회지, 56(7), pp.418~424 (2013), 대한이비인후과학회[DOI]
- 김세아, "이식형 골도 보청기의 발전과 앞으로의 발전", 대한이비인후-두경부외과학회지, 61(2), pp.67~75 (2018), 대한이비인후과학회[DOI]
- Zwartenkot JW, "Amplification options for patients with mixed hearing loss", Otol Neurotol, 35(2), pp.221~226 (2014)
- Mertens G, "An experimental objective method to determine maximum output and dynamic range of an active bone conduction implant : the Bonebridge", Otol Neurotol, 35(7), pp.1126~1130 (2014)
- Hol MK, "Comparison between a new implantable transcutaneous bone conductor and percutaneous bone-conduction hearing implant", Otol Neurotol, 34(6), pp.1071~1075 (2013)
- Mojallal H, "Retrospective audiological analysis of bone conduction versus round window vibratory stimulation in patients with mixed hearing loss", Int J Audiol, 54(6), pp.391~400 (2015)
- Park MJ, "Amplification of transcutaneous and percutaneous bone-conduction devices with a test-band in an induced model of conductive hearing loss", Int J Audiol, 55(11), pp.653~657 (2016)
- Finbow J, "A comparison between wireless CROS and bone-anchored hearing devices for single-sided deafness : a pilot study", Otol Neurotol, 36(5), pp.819~825 (2015)
- Snik AF, "Consensus statements on the BAHA system: where do we stand at present?", Ann Otol Rhinol Laryngol Suppl, 195, pp.2~12 (2005)
- Leterme G, "Contralateral routing of signal hearing aid versus transcutaneous bone conduction in single-sided deafness", Audiol Neurootol, 20(4), pp.251~260 (2015)
- Siegert R, "A new semi-implantable transcutaneous bone conduction device : clinical, surgical, and audiologic outcomes in patients with congenital ear canal atresia", Otol Neurotol, 34(5), pp.927~934 (2013)
- Håkansson B, "Hearing thresholds with direct bone conduction versus conventional bone conduction", Scand Audiol, 13(1), pp.1~3 (1984)
- 신중욱, "Surgical and Audiologic Comparison Between Sophono and Bone-Anchored Hearing Aids Implantation", Clinical and Experimental Otorhinolaryngology, 9(1), pp.21~26 (2016), 대한이비인후과학회