Prosthetic Mitral valve Leaflet Escape
0
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Leaflet escape of prosthetic valve is rare but potentially life threatening. It is essential to make timely diagnosis in order to avoid mortality. Transesophageal echocardiography and cinefluoroscopy is usually diagnostic and the location of the missing leaflet can be identified by computed tomography (CT). Emergent surgical correction is mandatory. We report a case of fractured escape of Edward-Duromedics mitral valve 27 years after the surgery. The patient presented with symptoms of acute decompensated heart failure and cardiogenic shock. She was instantly intubated and mechanically ventilated. After prompt evaluation including transthoracic echocardiography and CT, the escape of the leaflet was confirmed. The patient underwent emergent surgery for replacement of the damaged prosthetic valves immediately. Eleven days after the surgery, the dislodged leaflet in iliac artery was removed safely and the patient recovered well.
Abstract
Leaflet escape of prosthetic valve is rare but potentially life threatening. It is essential to make timely diagnosis in order to avoid mortality. Transesophageal echocardiography and cinefluoroscopy is usually diagnostic and the location of the missing leaflet can be identified by computed tomography (CT). Emergent surgical correction is mandatory. We report a case of fractured escape of Edward-Duromedics mitral valve 27 years after the surgery. The patient presented with symptoms of acute decompensated heart failure and cardiogenic shock. She was instantly intubated and mechanically ventilated. After prompt evaluation including transthoracic echocardiography and CT, the escape of the leaflet was confirmed. The patient underwent emergent surgery for replacement of the damaged prosthetic valves immediately. Eleven days after the surgery, the dislodged leaflet in iliac artery was removed safely and the patient recovered well.
참고문헌 (14)
- Deuvaert FE, "Leaflet escape of a mitral Duromedics prosthesis. Case report", Acta Chir Belg, 89, pp.15~18 (1989)
- Klepetko W, "Leaflet fracture in Edwards-Duromedics bileaflet valves", J Thorac Cardiovasc Surg, 97, pp.90~94 (1989)
- Alvarez J, "Leaflet escape from a Duromedics valve", J Thorac Cardiovasc Surg, 99, pp.372~ (1990)
- Kumar N, "Leaflet embolisation from Duromedics valves: a report of two cases", Thorac Cardiovasc Surg, 39, pp.382~383 (1991)
- Tatou E, "Fracture-embolization of duromedics valve prosthesis and microscopic uncommon lesions", Ann Thorac Surg, 71, pp.1366~1369 (2001)
- Baudet E, "A 8-year follow- up of the Edwards-Duromedics bileaflet prosthesis", J Cardiovasc Surg (Torino), 36, pp.437~442 (1995)
- Eichler MJ, "Mechanical heart valve cavitation: valve specific parameters", Int J Artif Organs, 27, pp.855~867 (2004)
- Fragoulis S, "Fractured prosthetic valve leaflet", Eur J Cardiothorac Surg, 34, pp.907~ (2008)
- Devbhandari MP, "Long-term event-free survival with an embolised prosthetic valve leaflet in the thoracic aorta", J Cardiothorac Surg, 3, pp.34~ (2008)
- Vogel W, "Cineradiography for determination of normal and abnormal function in mechanical heart valves", Am J Cardiol, 71, pp.225~232 (1993)
- Bottio T, "Leaflet escape in a new bileaflet mechanical valve: TRI technologies", Circulation, 107, pp.2303~2306 (2003)
- 이삼윤, "기계판막엽의 탈출 -1예 보고-", The Korean Journal of Thoracic and Cardiovascular Surgery, 40(1), pp.63~65 (2007), 대한흉부외과학회
- 김재현, "Edwards Duromedics 기계판막의 판엽 이탈 -1예 보고-", The Korean Journal of Thoracic and Cardiovascular Surgery, 37(4), pp.373~375 (2004), 대한흉부외과학회
- 윤영남, "Edwards Duromedics 기계판막엽 탈출 - 1례 보고 -", The Korean Journal of Thoracic and Cardiovascular Surgery, 35(1), pp.60~63 (2002), 대한흉부외과학회