Echocardiographic Classification of Perimembranous Ventricular Septal Defect Guides Selection of the Occluder Design for Their Transcatheter Device Closure

1

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류내과학

초록

BACKGROUND: Perimembranous ventricular septal defects (VSDs) has proximate relation to the aortic and tricuspid valves as well as the conduction tissues. Transcatheter closure utilizes various off-label device designs. METHODS: Perimembranous VSD without aortic margin were classified as group A, with thick aortic margin as group B, with membranous septal aneurysm as group C and defects restricted by tricuspid valve attachments as group D. The proposed ideal design was asymmetric device in group A; duct occluder I (ADOI) and muscular ventricular septal occluder (MVSO) in group B; thin profile duct occluder II (ADOII) in group C and ADOI in group D. Device was 0–2 mm larger than the defect. RESULTS: Eighty patients with VSD measuring 6.83 ± 2.87mm underwent successful closure. Device was retrieved before release in one group A and one group C patient due to aortic regurgitation. Asymmetric device was used in 16 group A defects. Among group B defects, ADOI was used in 5, ADOII in 5, MVSO in one and asymmetric device in 3. Group C defects were closed with ADOI in 7, ADOII in 10 and asymmetric device in 3. Three patients with multiple exits had 2 ADOII devices. Group D defects were closed using ADOI in 20 and ADOII in 10 patients. There was no late aortic regurgitation or heart block on a follow-up exceeding 7 years. CONCLUSIONS: This echocardiographic classification helps device selection in every single patient. While asymmetric device is uniquely suited for group A defects, different designs are appropriate in the other groups.

Abstract

BACKGROUND: Perimembranous ventricular septal defects (VSDs) has proximate relation to the aortic and tricuspid valves as well as the conduction tissues. Transcatheter closure utilizes various off-label device designs. METHODS: Perimembranous VSD without aortic margin were classified as group A, with thick aortic margin as group B, with membranous septal aneurysm as group C and defects restricted by tricuspid valve attachments as group D. The proposed ideal design was asymmetric device in group A; duct occluder I (ADOI) and muscular ventricular septal occluder (MVSO) in group B; thin profile duct occluder II (ADOII) in group C and ADOI in group D. Device was 0–2 mm larger than the defect. RESULTS: Eighty patients with VSD measuring 6.83 ± 2.87mm underwent successful closure. Device was retrieved before release in one group A and one group C patient due to aortic regurgitation. Asymmetric device was used in 16 group A defects. Among group B defects, ADOI was used in 5, ADOII in 5, MVSO in one and asymmetric device in 3. Group C defects were closed with ADOI in 7, ADOII in 10 and asymmetric device in 3. Three patients with multiple exits had 2 ADOII devices. Group D defects were closed using ADOI in 20 and ADOII in 10 patients. There was no late aortic regurgitation or heart block on a follow-up exceeding 7 years. CONCLUSIONS: This echocardiographic classification helps device selection in every single patient. While asymmetric device is uniquely suited for group A defects, different designs are appropriate in the other groups.

참고문헌 (18)

  1. Penny DJ, "Ventricular septal defect", Lancet, 377, pp.1103~1112 (2011)
  2. Ho SY, "Morphology of perimembranous ventricular septal defects : implications for transcatheter device closure", J Interv Cardiol, 17, pp.99~108 (2004)
  3. Butera G, "Transcatheter closure of perimembranous ventricular septal defects : early and long-term results", J Am Coll Cardiol, 50, pp.1189~1195 (2007)
  4. Yang J, "Transcatheter versus surgical closure of perimembranous ventricular septal defects in children : a randomized controlled trial", J Am Coll Cardiol, 63, pp.1159~1168 (2014)
  5. Chen Q, "Comparison of transthoracic device closure and surgical repair with right submammary or right infra-axillary thoracotomy for perimembranous VSD", Thorac Cardiovasc Surg, 67, pp.8~13 (2019)
  6. Dai XF, "A comparative study of minimal lower-sternal incision device closure, minimal right thoracic incision device closure, and midsternal open repair of isolated perimembranous VSD, a retrospective cohort study", Int J Cardiol, 306, pp.15~19 (2020)
  7. Predescu D, "Complete heart block associated with device closure of perimembranous ventricular septal defects", J Thorac Cardiovasc Surg, 136, pp.1223~1228 (2008)
  8. Sivakumar K, "Textbook of Echocardiography", pp.474~503 (2018), Jaypee Publishers
  9. Carminati M, "Transcatheter closure of congenital ventricular septal defects : results of the European Registry", Eur Heart J, 28, pp.2361~2368 (2007)
  10. Lee SM, "Transcatheter closure of perimembranous ventricular septal defect using Amplatzer ductal occluder", Catheter Cardiovasc Interv, 82, pp.1141~1146 (2013)
  11. Kanaan M, "Follow-up of patients with interventional closure of ventricular septal defects with Amplatzer duct occluder II", Pediatr Cardiol, 36, pp.379~385 (2015)
  12. Lytrivi ID, "Normal values for left ventricular volume in infants and young children by the echocardiographic subxiphoid five-sixth area by length(bullet)method", J Am Soc Echocardiogr, 24, pp.214~218 (2011)
  13. Bergmann M, "Short-and long-term outcome after interventional VSD closure : a single-center experience in pediatric and adult patients", Pediatr Cardiol, 42, pp.78~88 (2020)
  14. Saurav A, "Comparison of percutaneous device closure versus surgical closure of peri-membranous ventricular septal defects : a systematic review and meta-analysis", Catheter Cardiovasc Interv, 86, pp.1048~1056 (2015)
  15. Li G, "Re-evaluation of the criteria for asymmetric amplatzer occluders in the closure of perimembranous ventricular septal defects: a case series report", Medicine (Baltimore), 99, pp.e21356~ (2020)
  16. Li H, "Short-and medium-term follow-up of transcatheter closure of perimembranous ventricular septal defects", BMC Cardiovasc Disord, 19, pp.222~ (2019)
  17. Li D, "Comparisons of perventricular device closure, conventional surgical repair, and transcatheter device closure in patients with perimembranous ventricular septal defects: a network meta-analysis", BMC Surg, 20, pp.115~ (2020)
  18. Haddad RN, "Device closure of perimembranous ventricular septal defect : Choosing between Amplatzer occluders", Front Pediatr, 7, pp.300~ (2019)

관련 논문AI추천

Echocardiographic Classification and Guidance for Transcatheter Closure of Ventricular Septal DefectSeong Ho Kim, So-Ick Jang한국심초음파학회 · Journal of Cardiovascular Imaging2021 · 제29권 제4호Early Experiences Using Cocoon Occluders for Closure of a Ventricular Septal Defect허준, 강이석, 김이슬, 박효정, 송진영한국심초음파학회 · Journal of Cardiovascular Imaging2018 · 제26권 제3호Outcome of Transcatheter Closure of Oval Shaped Atrial Septal Defect with Amplatzer Septal Occluder백재숙, 송진영, 심우섭, 이상윤, 최은영연세대학교의과대학 · Yonsei Medical Journal2013 · 제54권 제5호A Case of Perimembranous Ventricular Septal Defect Associated with Sinus of Valsalva Aneurysm Mimicking Membranous Septal Aneurysm김신재, 김용균, 김종민, 김형래, 이상곤, 이준호, 임경훈, 정종필한국심초음파학회 · Journal of Cardiovascular Imaging2015 · 제23권 제2호Segmental Analysis of Right Ventricular Longitudinal Deformation in Children before and after Percutaneous Closure of Atrial Septal Defect강소연, 고재곤, 고홍기, 김영휘, 백재숙, 서창덕, 유정진, 조은경한국심초음파학회 · Journal of Cardiovascular Imaging2014 · 제22권 제4호Three-dimensional Echocardiography for Decision Making in a Complex Ventricular Septal DefectAnil Kumar Singhi, Aritra Mukherji, Soumya Kanti Mohapatra한국심초음파학회 · Journal of Cardiovascular Imaging2021 · 제29권 제2호