백-밸브-마스크 환기증진을 위한 보조기구 개발 및 효과
Assistant device development and effects for promotion of bag-valve-mask ventilation
1
조회
0
다운로드
2
인용
0
참고문헌
키워드
초록
Purpose: The purpose of this study was to develop an assistant device for the promotion of bag-valve-mask ventilation based on a non-equivalent control group pre-test and post-test design. Methods: The experimental tool was a mask assistance device developed by the researchers. Data were analyzed using SPSS 21.0 with the cardiopulmonary resuscitation (CPR) evaluation program from August 18 to 30, 2016. The research tools included general, hand-related, and ventilation-related characteristics. Results: Before and after using the mask assistance device, the tidal volume increased by 64 mL (p<.001) from 461.76 mL to 525.86 mL. The tidal volume for control was 477.86 mL, and there was a statistical difference (p<.05). The ventilation frequency in device users was 10 times per minute for a total of 20 ventilations with before 10.65 after 10 times, and that of the control group was before 10.36 times after 10 times; there was no difference in both groups(p>.05). The accuracy of the assistance device was 81.72±30.86%, which was a very high value. However, the accuracy of ventilation in the control group with no assistance device was 18.97±32.44%, which was a very low accuracy rate. Conclusion: This study's results suggested utilizing the newly-developed mask assistance device in CPR, and showed increases in tidal volume and accuracy of ventilation using the bag-valve-mask ventilation equipment. The general and hand-related characteristics did not have any effect, so the use of the device proved to increase the efficacy in all users.
Abstract
Purpose: The purpose of this study was to develop an assistant device for the promotion of bag-valve-mask ventilation based on a non-equivalent control group pre-test and post-test design. Methods: The experimental tool was a mask assistance device developed by the researchers. Data were analyzed using SPSS 21.0 with the cardiopulmonary resuscitation (CPR) evaluation program from August 18 to 30, 2016. The research tools included general, hand-related, and ventilation-related characteristics. Results: Before and after using the mask assistance device, the tidal volume increased by 64 mL (p<.001) from 461.76 mL to 525.86 mL. The tidal volume for control was 477.86 mL, and there was a statistical difference (p<.05). The ventilation frequency in device users was 10 times per minute for a total of 20 ventilations with before 10.65 after 10 times, and that of the control group was before 10.36 times after 10 times; there was no difference in both groups(p>.05). The accuracy of the assistance device was 81.72±30.86%, which was a very high value. However, the accuracy of ventilation in the control group with no assistance device was 18.97±32.44%, which was a very low accuracy rate. Conclusion: This study's results suggested utilizing the newly-developed mask assistance device in CPR, and showed increases in tidal volume and accuracy of ventilation using the bag-valve-mask ventilation equipment. The general and hand-related characteristics did not have any effect, so the use of the device proved to increase the efficacy in all users.
참고문헌 (10)
- Robert WN, "Part 1: Executive summary, 2015 american heart association guidelines update for cardiopulmonary resuscitation and emergency cardiovascular care", Circulation, 132(18), pp.315~367 (2015)[DOI]
- Odegaard S, "Time used for ventilation in two-rescuer CPR with a bag-valve-mask device during out-of hospital cardiac arrest", Resuscitation, 77(1), pp.57~62 (2007)[DOI]
- Jin YH, "The Effects of Ventilation Rate and Characteristics of the Hand on Inspiratory Oxygen Concentration and Tidal Volume During Bag-Valve Ventilation", Korean Soc Emerg Med, 9(1), pp.7~13 (1998)
- Cho YC, "How can a single rescuer adequately deliver tidal volume with a manual resuscitator? An improved device for delivering regular tidal volume", Emer Med J, 28(1), pp.40~43 (2011)[DOI]
- Park SO, "Evaluation of a novel paediatric self-inflating bag to improve accuracy of tidal volumes delivered during simulated advanced paediatric resuscitation", Resuscitation, 83(1), pp.101~106 (2011)[DOI]
- 한수천, "가슴 압박 시행 중 기존의 수동식 소생기와 새로 고안된 수동식 소생기의 일회 호흡량 비교", 대한응급의학회지, 26(4), pp.269~275 (2015), 대한응급의학회
- Lee GS, "The performances of standard and ResMed masks during bag-valve-mask ventilation", Prehospital Emergency Care, 17(2), pp.235~240 (2012)[DOI]
- Tom PA, "Death by hyperventilation: A common and life-threatening problem during cardiopulmonary resuscitation", Crit Care Med, 32(9), pp.S345~S351 (2004)[DOI]
- 조승묵, "Bag-Valve-Mask의 사용방법에 따른 일회호흡량과 평균기도압의 변화 연구", 한국응급구조학회지, 16(2), pp.67~74 (2012), 사단법인 한국응급구조학회
- 신소연, "RespiTrainer를 활용한 백-밸브마스크 환기에서 일회호흡량과 기도압 비교 연구", 한국화재소방학회 논문지, 28(6), pp.76~81 (2014), 한국화재소방학회[DOI]