본문 바로가기
KERD
논문KCI 등재

Comparison of Endoscopic Ultrasound Biopsy Needles for Endoscopic Ultrasound-Guided Liver Biopsy

1

조회

0

다운로드

3

인용

0

참고문헌

키워드

주제분류내과학

초록

Background/Aims: To compare the performance of latest commercially available endoscopic ultrasound biopsy needles. Methods: Six latest commercially available needles were tested on a freshly harvested bovine liver; the tested needles included three19 G, one 20 G, and two 22 G needles. Five biopsies were performed per needle with 10 mL of wet suction. The primary outcome wasthe number of complete portal tracts (CPTs) per needle aspirate. The secondary outcomes were the mean specimen length and meanfragment length. Analysis of variance and Tukey’s test were applied. Results: All 19 G needles and the 20 G needle yielded similar mean CPTs and were superior to the SharkCore 22 G needle (p<0.001adjusted for multiplicity). There was no statistically significant difference in total specimen length among the three 19 G needles and the20 G needle tested. The two 22 G needles performed similarly with respect to the number of CPTs, mean fragment length, and meanspecimen length (adjusted p=0.07, p=0.59, and p=0.10, respectively). Conclusions: The specimen adequacy was similar among the 3 latest commercially available 19 G needles. The endoscopist may choosea larger-bore needle based on availability without concerns of specimen adequacy. Further studies are needed to assess the ease ofneedle use in various anatomical locations and to confirm the optimal needle design.

Abstract

Background/Aims: To compare the performance of latest commercially available endoscopic ultrasound biopsy needles. Methods: Six latest commercially available needles were tested on a freshly harvested bovine liver; the tested needles included three19 G, one 20 G, and two 22 G needles. Five biopsies were performed per needle with 10 mL of wet suction. The primary outcome wasthe number of complete portal tracts (CPTs) per needle aspirate. The secondary outcomes were the mean specimen length and meanfragment length. Analysis of variance and Tukey’s test were applied. Results: All 19 G needles and the 20 G needle yielded similar mean CPTs and were superior to the SharkCore 22 G needle (p<0.001adjusted for multiplicity). There was no statistically significant difference in total specimen length among the three 19 G needles and the20 G needle tested. The two 22 G needles performed similarly with respect to the number of CPTs, mean fragment length, and meanspecimen length (adjusted p=0.07, p=0.59, and p=0.10, respectively). Conclusions: The specimen adequacy was similar among the 3 latest commercially available 19 G needles. The endoscopist may choosea larger-bore needle based on availability without concerns of specimen adequacy. Further studies are needed to assess the ease ofneedle use in various anatomical locations and to confirm the optimal needle design.

참고문헌 (14)

  1. Kalambokis G, "Transjugular liver biopsy--indications, adequacy, quality of specimens, and complications--a systematic review", J Hepatol, 47, pp.284~294 (2007)
  2. Rockey DC, "Liver biopsy", Hepatology, 49, pp.1017~1044 (2009)
  3. Pineda JJ, "EUS-guided liver biopsy provides diagnostic samples comparable with those via the percutaneous or transjugular route", Gastrointest Endosc, 83, pp.360~365 (2016)
  4. Gleeson FC, "Adequacy of endoscopic ultrasound core needle biopsy specimen of nonmalignant hepatic parenchymal disease", Clin Gastroenterol Hepatol, 6, pp.1437~1440 (2008)
  5. Gor N, "Histological adequacy of EUS-guided liver biopsy when using a 19-gauge non-Tru-cut FNA needle", Gastrointest Endosc, 79, pp.170~172 (2014)
  6. Dewitt J, "Initial experience with EUS-guided tru-cut biopsy of benign liver disease", Gastrointest Endosc, 69(3 Pt 1), pp.535~542 (2009)
  7. Sey MS, "EUS-guided liver biopsy for parenchymal disease: a comparison of diagnostic yield between two core biopsy needles", Gastrointest Endosc, 83, pp.347~352 (2016)
  8. Bravo AA, "Liver biopsy", N Engl J Med, 344, pp.495~500 (2001)
  9. Stavropoulos SN, "High yield of same-session EUS-guided liver biopsy by 19-gauge FNA needle in patients undergoing EUS to exclude biliary obstruction", Gastrointest Endosc, 75, pp.310~318 (2012)
  10. Diehl DL, "Endoscopic ultrasound-guided liver biopsy : a multicenter experience", Endosc Int Open, 3, pp.E210~E215 (2015)
  11. Schulman AR, "Optimizing EUS-guided liver biopsy sampling : comprehensive assessment of needle types and tissue acquisition techniques", Gastrointest Endosc, 85, pp.419~426 (2017)
  12. Lee WJ, "Comparison of the diagnostic yield of EUS needles for liver biopsy : ex vivo study", Diagn Ther Endosc, 2017, pp.1497831~ (2017)
  13. Bang JY, "Randomized trial comparing the Franseen and fork-tip needles for EUS-guided fine-needle biopsy sampling of solid pancreatic mass lesions", Gastrointest Endosc, 87, pp.1432~1438 (2018)
  14. Kemper AR, "Biomechanical response of human liver in tensile loading", Ann Adv Automot Med, 54, pp.15~26 (2010)

관련 논문AI추천

Superior Specimen and Diagnostic Accuracy with Endoscopic Ultrasound-Guided Liver Biopsies Using 19-Gauge versus 22-Gauge Core NeedlesElizabeth John, Jason Schmidt, Prashant Kedia, Priyanka Acharya, Rucha M. Shah, Sheila Rastegari대한소화기내시경학회 · Clinical Endoscopy2021 · 제54권 제5호Comparison of the Diagnostic Yield of the Standard 22-Gauge Needle and the New 20-Gauge Forward-Bevel Core Biopsy Needle for Endoscopic Ultrasound-Guided Tissue Acquisition from Pancreatic LesionsHirotoshi Ishiwatari, Hiroyuki Matsubayashi, Hiroyuki Ono, Junya Sato, Keiko Sasaki, Kenichiro Imai, Kinichi Hotta, Kohei Takizawa, Masao Yoshida, Naomi Kakushima, Noboru Kawata, Sayo Ito, Shinya Fujie거트앤리버 소화기연관학회협의회 · Gut and Liver2019 · 제13권 제3호Endoscopic Ultrasound-Guided, Percutaneous, and Transjugular Liver Biopsy: A Comparative Systematic Review and Meta-AnalysisAhmad Najdat Bazarbashi, Basile Njei, Harry R. Aslanian, Marvin Ryou, Thiruvengadam Muniraj, Thomas R. McCarty대한소화기내시경학회 · Clinical Endoscopy2020 · 제53권 제5호Endoscopic Ultrasound-Guided Liver Biopsy Using a Core Needle for Hepatic Solid Mass김태현, 전형구, 최금하, Hee Chan Yang대한소화기내시경학회 · Clinical Endoscopy2019 · 제52권 제4호Comparison of Histologic Core Portions Acquired from a Core Biopsy Needle and a Conventional Needle in Solid Mass Lesions: A Prospective Randomized Trial배한익, 이반석, 장정식, 정민규, 조창민거트앤리버 소화기연관학회협의회 · Gut and Liver2017 · 제11권 제4호Confirming Whether Fine Needle Biopsy Device Shortens the Learning Curve of Endoscopic Ultrasound-Guided Tissue Acquisition Without Rapid Onsite EvaluationBor-Shyang Sheu, Cheng-Lin Wu, Meng-Ying Lin, Mitsuhiro Kida, Wei-Lun Chang대한소화기내시경학회 · Clinical Endoscopy2021 · 제54권 제3호