본문 바로가기
KERD
논문KCI 등재

Evaluation of the mechanical properties of current biliary self-expandable metallic stents: axial and radial force, and axial force zero border

2

조회

0

다운로드

2

인용

0

참고문헌

키워드

주제분류내과학

초록

Background/Aims: Mechanical properties (MPs) and axial and radial force (AF and RF) may influence the efficacy and complicationsof self-expandable metallic stent (SEMS) placement. We measured the MPs of various SEMSs and examined their influence on theSEMS clinical ability. Methods: We evaluated the MPs of 29 types of 10-mm SEMSs. RF was measured using a conventional measurement device. AF wasmeasured using the conventional and new methods, and the correlation between the methods was evaluated. Results: A high correlation in AFs was observed, as measured by the new and conventional manual methods. AF and RF scatterplotsdivided the SEMSs into three subgroups according to structure: hook-and-cross-type (low AF and RF), cross-type (high AF and lowRF), and laser-cut-type (intermediate AF and high RF). The hook-and-cross-type had the largest axial force zero border (>20°), followedby the laser-cut and cross types. Conclusions: MPs were related to stent structure. Hook-and-cross-type SEMSs had a low AF and high axial force zero border and wereconsidered safest because they caused minimal stress on the biliary wall. However, the increase in RF must be overcome.

Abstract

Background/Aims: Mechanical properties (MPs) and axial and radial force (AF and RF) may influence the efficacy and complicationsof self-expandable metallic stent (SEMS) placement. We measured the MPs of various SEMSs and examined their influence on theSEMS clinical ability. Methods: We evaluated the MPs of 29 types of 10-mm SEMSs. RF was measured using a conventional measurement device. AF wasmeasured using the conventional and new methods, and the correlation between the methods was evaluated. Results: A high correlation in AFs was observed, as measured by the new and conventional manual methods. AF and RF scatterplotsdivided the SEMSs into three subgroups according to structure: hook-and-cross-type (low AF and RF), cross-type (high AF and lowRF), and laser-cut-type (intermediate AF and high RF). The hook-and-cross-type had the largest axial force zero border (>20°), followedby the laser-cut and cross types. Conclusions: MPs were related to stent structure. Hook-and-cross-type SEMSs had a low AF and high axial force zero border and wereconsidered safest because they caused minimal stress on the biliary wall. However, the increase in RF must be overcome.

참고문헌 (21)

  1. Conio M, "Covered versus uncovered self-expandable metal stent for palliation of primary malignant extrahepatic biliary strictures : a randomized multicenter study", Gastrointest Endosc, 88, pp.283~291 (2018)
  2. Sawas T, "Self-expandable metal stents versus plastic stents for malignant biliary obstruction : a meta-analysis", Gastrointest Endosc, 82, pp.256~267 (2015)
  3. Lam R, "Fully covered metal biliary stents : a review of the literature", World J Gastroenterol, 27, pp.6357~6373 (2021)
  4. Koh Kitagawa, "Laser-cut-type versus braided-type covered self-expandable metallic stents for distal biliary obstruction caused by pancreatic carcinoma: a retrospective comparative cohort study", Clinical Endoscopy, 55(3), pp.434~442 (2022), 대한소화기내시경학회[DOI]
  5. Isayama H, "Comparison of partially covered nitinol stents with partially covered stainless stents as a historical control in a multicenter study of distal malignant biliary obstruction : the WATCH study", Gastrointest Endosc, 76, pp.84~92 (2012)
  6. Saito H, "Biliary endoprosthesis using Gore-Tex covered expandable metallic stents : preliminary clinical evaluation", Nihon Igaku Hoshasen Gakkai Zasshi, 54, pp.180~182 (1994)
  7. Hori Y, "Novel characteristics of traction force in biliary self-expandable metallic stents", Dig Endosc, 29, pp.347~352 (2017)
  8. U.S. Department of Health and Human Services, Food and Drug Administration, "Center for Devices and Radiological Health. Metal expandable biliary stents: premarket notification (510(k)) submissions", Center for Drug Evaluation and Research, ~ (2019)
  9. Isayama H, "Measurement of radial and axial forces of biliary self-expandable metallic stents", Gastrointest Endosc, 70, pp.37~44 (2009)
  10. Katsinelos P, "A fully covered self-expandable metal stent anchored by a 10-Fr double pigtail plastic stent : an effective anti-migration technique", Ann Gastroenterol, 30, pp.114~117 (2017)
  11. Isayama H, "Understanding the mechanical forces of self-expandable metal stents in the biliary ducts", Curr Gastroenterol Rep, 18, pp.64~ (2016)
  12. Nakai Y, "Metallic stent with high axial force as a risk factor for cholecystitis in distal malignant biliary obstruction", J Gastroenterol Hepatol, 29, pp.1557~1562 (2014)
  13. Jang S, "Association of covered metallic stents with cholecystitis and stent migration in malignant biliary stricture", Gastrointest Endosc, 87, pp.1061~1070 (2018)
  14. Kawakubo K, "Risk factors for pancreatitis following transpapillary self-expandable metal stent placement", Surg Endosc, 26, pp.771~776 (2012)
  15. Takeda T, "Novel risk factors for recurrent biliary obstruction and pancreatitis after metallic stent placement in pancreatic cancer", Endosc Int Open, 8, pp.E1603~ (2020)
  16. Nakai Y, "Risk factors for covered metallic stent migration in patients with distal malignant biliary obstruction due to pancreatic cancer", J Gastroenterol Hepatol, 29, pp.1744~1749 (2014)
  17. Sasaki T, "Comparing the mechanical properties of a self-expandable metallic stent for colorectal obstruction : proposed measurement method of axial force using a new measurement machine", Dig Endosc, 33, pp.170~178 (2021)
  18. Rodrigues-Pinto E, "Development of an online app to predict post-endoscopic retrograde cholangiopancreatography adverse events using a single-center retrospective cohort", Dig Dis, 39, pp.283~293 (2021)
  19. Ida Bagus B, "A rare clinical presentation of third part duodenal perforation due to post-endoscopic retrograde cholangiopancreatography stent migration on advanced stage peri-ampullary tumor", JGH Open, 5, pp.968~970 (2021)
  20. Mukai T, "Comparison of axial force and cell width of self-expandable metallic stents : which type of stent is better suited for hilar biliary strictures?", J Hepatobiliary Pancreat Sci, 18, pp.646~652 (2011)
  21. Weaver M, "Metal biliary stent erosion through the common bile duct leading to obstructive jaundice and cholangitis", Am J Gastroenterol, 116, pp.1372~ (2021)

관련 논문AI추천

Functional Self-Expandable Metal Stents in Biliary Obstruction강대환, 고광현, 권창일, 함기백대한소화기내시경학회 · Clinical Endoscopy2013 · 제46권 제5호‘Radial force’ of colonic stents: A parameter without consistency, definition or standardDerek W. Edwards, Graham D. Milward, Hans-Ulrich Laasch소화기인터벤션의학회 · International Journal of Gastrointestinal Intervention2020 · 제9권 제3호Understanding mechanical properties of biliary metal stents for wise stent selection정석대한소화기내시경학회 · Clinical Endoscopy2023 · 제56권 제5호Basic Knowledge about Metal Stent Development정석대한소화기내시경학회 · Clinical Endoscopy2016 · 제49권 제2호A Novel, Fully Covered Laser-Cut Nitinol Stent with Antimigration Properties for Nonresectable Distal Malignant Biliary Obstruction: A Multicenter Feasibility StudyChimyon Gon, Hirofum Kogure, Hiroyuki Isayama, Kazuhiko Koike, Kazumichi Kawakubo, Keisuke Yamamoto, Kenji Hirano, Koji Miyabayashi, Kouta Inoue, Minoru Tada, Naoki Sasahira, Natsuyo Yamamoto, Rie Uchino, Saburo Matsubara, Suguru Mizuno, Takashi Sasaki, Takeshi Tsujino, Tsuyoshi Hamada, Yousuke Nakai, Yukiko Ito거트앤리버 소화기연관학회협의회 · Gut and Liver2013 · 제7권 제6호A novel fully covered metal stent for unresectable malignant distal biliary obstruction: results of a multicenter prospective studyArata Sakai, Atsuhiro Masuda, Daisuke Shirasaka, Hideyuki Shiomi, Ikuya Miki, Keisuke Furumatsu, Kodai Yamanaka, Saori Kakuyama, Shiei Yoshida, Shinya Kohashi, Takaaki Eguchi, Takao Iemoto, Takashi Kobayashi, Yoshihiro Okabe, Yosuke Yagi, Yuzo Kodama대한소화기내시경학회 · Clinical Endoscopy2024 · 제57권 제3호

이 간행물의 다른 논문 (17건)