Evaluation of the mechanical properties of current biliary self-expandable metallic stents: axial and radial force, and axial force zero border
2
조회
0
다운로드
2
인용
0
참고문헌
키워드
초록
Background/Aims: Mechanical properties (MPs) and axial and radial force (AF and RF) may influence the efficacy and complicationsof self-expandable metallic stent (SEMS) placement. We measured the MPs of various SEMSs and examined their influence on theSEMS clinical ability. Methods: We evaluated the MPs of 29 types of 10-mm SEMSs. RF was measured using a conventional measurement device. AF wasmeasured using the conventional and new methods, and the correlation between the methods was evaluated. Results: A high correlation in AFs was observed, as measured by the new and conventional manual methods. AF and RF scatterplotsdivided the SEMSs into three subgroups according to structure: hook-and-cross-type (low AF and RF), cross-type (high AF and lowRF), and laser-cut-type (intermediate AF and high RF). The hook-and-cross-type had the largest axial force zero border (>20°), followedby the laser-cut and cross types. Conclusions: MPs were related to stent structure. Hook-and-cross-type SEMSs had a low AF and high axial force zero border and wereconsidered safest because they caused minimal stress on the biliary wall. However, the increase in RF must be overcome.
Abstract
Background/Aims: Mechanical properties (MPs) and axial and radial force (AF and RF) may influence the efficacy and complicationsof self-expandable metallic stent (SEMS) placement. We measured the MPs of various SEMSs and examined their influence on theSEMS clinical ability. Methods: We evaluated the MPs of 29 types of 10-mm SEMSs. RF was measured using a conventional measurement device. AF wasmeasured using the conventional and new methods, and the correlation between the methods was evaluated. Results: A high correlation in AFs was observed, as measured by the new and conventional manual methods. AF and RF scatterplotsdivided the SEMSs into three subgroups according to structure: hook-and-cross-type (low AF and RF), cross-type (high AF and lowRF), and laser-cut-type (intermediate AF and high RF). The hook-and-cross-type had the largest axial force zero border (>20°), followedby the laser-cut and cross types. Conclusions: MPs were related to stent structure. Hook-and-cross-type SEMSs had a low AF and high axial force zero border and wereconsidered safest because they caused minimal stress on the biliary wall. However, the increase in RF must be overcome.
참고문헌 (21)
- Conio M, "Covered versus uncovered self-expandable metal stent for palliation of primary malignant extrahepatic biliary strictures : a randomized multicenter study", Gastrointest Endosc, 88, pp.283~291 (2018)
- Sawas T, "Self-expandable metal stents versus plastic stents for malignant biliary obstruction : a meta-analysis", Gastrointest Endosc, 82, pp.256~267 (2015)
- Lam R, "Fully covered metal biliary stents : a review of the literature", World J Gastroenterol, 27, pp.6357~6373 (2021)
- Koh Kitagawa, "Laser-cut-type versus braided-type covered self-expandable metallic stents for distal biliary obstruction caused by pancreatic carcinoma: a retrospective comparative cohort study", Clinical Endoscopy, 55(3), pp.434~442 (2022), 대한소화기내시경학회[DOI]
- Isayama H, "Comparison of partially covered nitinol stents with partially covered stainless stents as a historical control in a multicenter study of distal malignant biliary obstruction : the WATCH study", Gastrointest Endosc, 76, pp.84~92 (2012)
- Saito H, "Biliary endoprosthesis using Gore-Tex covered expandable metallic stents : preliminary clinical evaluation", Nihon Igaku Hoshasen Gakkai Zasshi, 54, pp.180~182 (1994)
- Hori Y, "Novel characteristics of traction force in biliary self-expandable metallic stents", Dig Endosc, 29, pp.347~352 (2017)
- U.S. Department of Health and Human Services, Food and Drug Administration, "Center for Devices and Radiological Health. Metal expandable biliary stents: premarket notification (510(k)) submissions", Center for Drug Evaluation and Research, ~ (2019)
- Isayama H, "Measurement of radial and axial forces of biliary self-expandable metallic stents", Gastrointest Endosc, 70, pp.37~44 (2009)
- Katsinelos P, "A fully covered self-expandable metal stent anchored by a 10-Fr double pigtail plastic stent : an effective anti-migration technique", Ann Gastroenterol, 30, pp.114~117 (2017)
- Isayama H, "Understanding the mechanical forces of self-expandable metal stents in the biliary ducts", Curr Gastroenterol Rep, 18, pp.64~ (2016)
- Nakai Y, "Metallic stent with high axial force as a risk factor for cholecystitis in distal malignant biliary obstruction", J Gastroenterol Hepatol, 29, pp.1557~1562 (2014)
- Jang S, "Association of covered metallic stents with cholecystitis and stent migration in malignant biliary stricture", Gastrointest Endosc, 87, pp.1061~1070 (2018)
- Kawakubo K, "Risk factors for pancreatitis following transpapillary self-expandable metal stent placement", Surg Endosc, 26, pp.771~776 (2012)
- Takeda T, "Novel risk factors for recurrent biliary obstruction and pancreatitis after metallic stent placement in pancreatic cancer", Endosc Int Open, 8, pp.E1603~ (2020)
- Nakai Y, "Risk factors for covered metallic stent migration in patients with distal malignant biliary obstruction due to pancreatic cancer", J Gastroenterol Hepatol, 29, pp.1744~1749 (2014)
- Sasaki T, "Comparing the mechanical properties of a self-expandable metallic stent for colorectal obstruction : proposed measurement method of axial force using a new measurement machine", Dig Endosc, 33, pp.170~178 (2021)
- Rodrigues-Pinto E, "Development of an online app to predict post-endoscopic retrograde cholangiopancreatography adverse events using a single-center retrospective cohort", Dig Dis, 39, pp.283~293 (2021)
- Ida Bagus B, "A rare clinical presentation of third part duodenal perforation due to post-endoscopic retrograde cholangiopancreatography stent migration on advanced stage peri-ampullary tumor", JGH Open, 5, pp.968~970 (2021)
- Mukai T, "Comparison of axial force and cell width of self-expandable metallic stents : which type of stent is better suited for hilar biliary strictures?", J Hepatobiliary Pancreat Sci, 18, pp.646~652 (2011)
- Weaver M, "Metal biliary stent erosion through the common bile duct leading to obstructive jaundice and cholangitis", Am J Gastroenterol, 116, pp.1372~ (2021)