내측연결 시스템에서 임플란트 고정체의 경부 형태에 따른 지지골에서의 유한요소 응력분석
Finite Element Stress Analysis in Supporting Bone according to Crest Module Shape of Fixture in Internal ction System
2
조회
0
다운로드
0
인용
0
참고문헌
초록
Abstract
Finite Element Stress Analysis in Supporting Bone according to Crest Module Shape of Fixture in Internal Connection System Young-Nam Park, Hee-Jung Kim, Sang-Ho Oh, Chae-Heon Chung Dept of Prosthodontics, College of Dentistry, Chosun University The external contour of an implant can have significant effects on the load transfer characteristics and may result in different bone failure rates for different implant system. The purpose of this study was to investigate the effects of crest module shape and occlusal load direction on bone failure modes of five commercially available dental implant systems. Five different implant systems with internal connection; ITI (Model 1), Astra (Model 2), Bicon (Model 3), Friadent (Model 4), and Paragon (Model 5), comparable in size, but different in thread profile and cest module shapes, were compared using the finite element method. Conclusively, in the internal connection system of the implant-abutment connection methods, the stress-induced pattern at the supporting bone according to the abutment connection form had differenence among them, and implants with narrowing crestal module cross-sections at the top of the cortical bone created more favorable load transfer characteristics in this region. But it is considered that the future study is necessary about how this difference in the magnitude of the stress have an effect on the practical clinic.
참고문헌 (36)
- "Marginal tissue reactions at osseointegrated titanium fixtures", pp.39~52 (1986), Branemark PI
- "Comparative study of antirotational single tooth abutments", 73, pp.36~43 (1995)
- "Finite element analysis of non axial vs. axial loading of oral implants in the mandible of dog.", 25, pp.847~858 (1998)
- "Mechanics of the implant-abutment connection", 15, pp.519~526 (2000)
- "Evaluation of load transfer characte- ristics of five implant in compact bone at different load levels by finite element analysis", 92, pp.523~530 (2004)
- "Biomaterials and biomechanics of oral and maxillofacial implants", 15, pp.15~46 (2000)
- "Finite element analysis of implant prostheis according to platform width of fixture", 41, pp.674~688 (2003)
- "Use of an axisym- metric finite element method to compare maxillary bone variables for a loaded implant"
- "The influence of static and dynamic loading on marginal bone reactions around osseointegrated implants", 12, pp.207~218 (2001)
- "The implant neck", 10, pp.394~405 (1999)
- "A 10-year follow-up study of the masticatory system after treatment with osseointegrated implant bridges", 96, pp.243~252 (1988)
- "Comparison of stress transmission in the IMZ implant system with polyoxymethylene or titanium intramobile element:A finite element stress analysis", 7, pp.450~458 (1992)
- "Evaluating parameters of osseointegrated dental implants using finite element analysis-a two- dimentional comparative study examining the effects of implant diameter,implant shape,and load direction", 24, pp.80~88 (1998)
- "Fatigue strength of dental implants with different types of connection between fixture and abutment cylinder", 18, pp.1~19 (2002)
- "Osseointegrated implants for single tooth replacement", 6, pp.29~36 (1991)
- "A study on the fit of the implant- abutment-screw interface", 41, pp.503~518 (2003)
- "Biomechanical aspects of marginal bone resorption around osseointegrated implants", 15, pp.401~412 (2004)
- "A multicenter retrospective analysis of ITI implant system used for single-tooth replacements", 12, pp.237~242 (1997)
- "A biomechanical rationale for the use of short implants", 17, pp.126~131 (1991)
- "Load transfer from endosteal implants to supporting bone", pp.343~348
- "Two- dimentional FEM analysis of hydroxyapatite implants:diameter effects on stress distribution", 16, pp.6~11 (1990)
- "A scientific rationale for dental implant design", pp.329~343 (1999), Mosby
- "An experimental study on the tissue compatibility of the titanium blade-vent implant coated with HAP-alumina in the semi-functional state", 13, pp.387~401 (1987)
- "An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design", 8, pp.290~298 (1997)
- "Assessment of cold welding properties of internal conical interface two commercially available implant system", 81, pp.159~166 (1999)
- "In vitro evaluation of the strength of the conical implant- to- abutment joint in two commercially available implant systems", 83, pp.567~571 (2000)
- "Guidlines for analysis and redesign of dental implants", 7, pp.355~368 (1998)
- "Forces and moments on Branemark implants", 4, pp.241~247 (1989)
- "Bending overload and implant fracture", 10, pp.326~334 (1995)
- "Load factor control for implants in the posterior partially edentulous segment", 12, pp.360~370 (1997)
- "Numerical investigations of the influence of implant shape on stress distribution in the jaw bone", 4, pp.333~340 (1989)
- "The new concept of ITI hollow- cylinder and hollow- screw implants:Part 1", 3, pp.161~172
- "The new restorative concept of the ITI dental implant system:design and engineering", 13, pp.409~431 (1993)
- "Force distribution in splinted anterior teeth College of Dentistry", pp.484~494 825, Chosun University 421
- "Force distribution in splinted posterior teeth", 10, pp.1268~1276 (1957)
- "The biomechanics of force distribution in implant-supported prostheses", 8, pp.19~31 (1993)