Development of nitrogen and oxygen certified reference materials in 10 μmol/mol for the purity evaluation
2
조회
0
다운로드
1
인용
0
참고문헌
키워드
초록
We have developed 10 μmol/mol nitrogen and oxygen certified reference materials (CRMs) in helium, as a SI-traceable gas standard for a quantifying of impurities in pure gases for the first time in Korea. The standard gas mixtures of nitrogen and oxygen were prepared in 5000 μmol/mol and sequentially were diluted to 250 μmol/mol and 10 μmol/mol according to the gravimetric preparation. In each dilution step, two cylinders of CRMs were prepared. The verification of internal consistency among the prepared gas mixtures was performed by using GC-TCD. The amount fractions and those expanded uncertainties (k = 2) of nitrogen and oxygen in the standard gas mixtures were (10.12 ± 0.08) μmol/mol and (10.18 ± 0.08) μmol/mol for nitrogen, and (9.88 ± 0.06) μmol/mol and (9.94 ± 0.06) μmol/mol for oxygen, respectively. We have conducted a purity assessment of two commercial helium gases using developed CRMs. As the results of the purity assessment, nitrogen and oxygen were detected by (1.66 ± 0.03) μmol/mol and (0.31 ± 0.02) μmol/mol, respectively, as the impurities in one of the pure helium.
Abstract
We have developed 10 μmol/mol nitrogen and oxygen certified reference materials (CRMs) in helium, as a SI-traceable gas standard for a quantifying of impurities in pure gases for the first time in Korea. The standard gas mixtures of nitrogen and oxygen were prepared in 5000 μmol/mol and sequentially were diluted to 250 μmol/mol and 10 μmol/mol according to the gravimetric preparation. In each dilution step, two cylinders of CRMs were prepared. The verification of internal consistency among the prepared gas mixtures was performed by using GC-TCD. The amount fractions and those expanded uncertainties (k = 2) of nitrogen and oxygen in the standard gas mixtures were (10.12 ± 0.08) μmol/mol and (10.18 ± 0.08) μmol/mol for nitrogen, and (9.88 ± 0.06) μmol/mol and (9.94 ± 0.06) μmol/mol for oxygen, respectively. We have conducted a purity assessment of two commercial helium gases using developed CRMs. As the results of the purity assessment, nitrogen and oxygen were detected by (1.66 ± 0.03) μmol/mol and (0.31 ± 0.02) μmol/mol, respectively, as the impurities in one of the pure helium.
참고문헌 (23)
- J. T. Shipman, "An Introduction to Physical Science" (2014), Cengage Learning
- H. Ogino, , Anal. Chem, 63(14), pp.1376~1379 (1991)
- J. D. Young, , Intensive Care Med, 22(1), pp.77~86 (1996)
- Y. -F. Liew, , Appl. Phys. Lett, 77(17), pp.2650~2652 (2000)
- S. N. Ketkar, , Anal. Chem, 73(11), pp.2554~2557 (2001)
- M. Schaer, , Adv. Funct. Mater, 11(2), pp.116~121 (2001)
- T. Martens, , Appl. Phys. Lett, 92(4), pp.041504~ (2008)
- F. Pavese, , Metrologia, 46(1), pp.47~61 (2009)
- T. Tsujimura, "OLED Display Fundamentals and Applications", pp.61~98 (2017), John Wiley & Sons, Inc
- P. Coning, "A Practical Guide to Gas Analysis by Gas Chromatography", pp.63~ (2019), Elsevier Inc
- A. Turak, "Handbook of Organic Materials for Electronic and Photonic Devices", pp.599~662 (2019), Woodhead Publishing
- F. M. Dunnivant, "Gas Chromatography, Liquid Chromatography, and Capillary Electrophoresis-Mass Spectrometry: A Basic Introduction", pp.22~ (2011), Whitman College
- International Organization for Standardization, "ISO 17034 : 2016 General requirements for the competence of reference material producers", ~ (2016)
- R. Zeleny, , Anal. Chim. Acta, 634(2), pp.237~242 (2009)
- E. Maier, , TrAC, Trends Anal. Chem, 16(9), pp.496~503 (1997)
- S. A. Wise, , Anal. Bioanal. Chem, 410(8), pp.2045~2049 (2018)
- ISO 6142-1 : 2015 Gas analysis-Preparation of calibration gas mixtures-Part 1 : Gravimetric method for Class I mixtures, "International Organization for Standardization", ~ (2015)
- S. H. Oh, , Metrologia, 50(4), pp.318~324 (2013)
- N. Matsumoto, , Metrologia, 41(3), pp.178~188 (2004)
- M. J. T. Milton, , Metrologia, 48(5), pp.R1~R9 (2011)
- J. B. Lee, "Rep. Korea, A Mixed Gas Supply System and Operation Method", Patent No. 10-1000996, ~ (2010)
- W. G. Jennings, "Applications of Glass Capillary Gas Chromatography" (1981), CRC Press
- D. R. Rushneck, , J. Chromatogr. Sci, 10(2), pp.123~125 (1972)