Effects of Coronal Thread Pitch in Scalloped Implant with 2 Different Connections on Loading Stress using 3 Dimensional Finite Element Analysis
2
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Purpose of present study is to investigate the effects of thread pitch in coronal portion in scalloped implant with 2 different connections on loading stress using 3 dimensional finite element analysis. Scalloped implant with 4 different thread pitches (0.4mm, 0.5mm, 0.6, and 0.7mm) in the coronal part was modeled with 2 different implant-abutment connections. Platform matching connection had the same implant and abutment diameter so that they were in flush contact at the periphery while platform mismatching connection had smaller abutment diameter than implant so that their connection was made away from periphery of implant-bone interface. Occlusal loading of 100N force was applied vertically and 30 degree obliquely to all 8 models and the maximum von Mises bone stress was identified. Loading stress as highly concentrated in cortical bone. Platform mismatching scalloped implant with small thread pitch (0.4mm) model had consistently lowest maximum von Mises bone stress in vertical and oblique loads. Platform matching model had lowest maximum von Mises bone stress with 0.6mm thread pitch in vertical load and with 0.4mm thread pitch in oblique load. Platform mismatching connection had important roles in reducing maximum von Mises bone stress. Scalloped implant with smaller coronal thread pitch showed trend of reducing maximum von Mises bone stress under load.
Abstract
Purpose of present study is to investigate the effects of thread pitch in coronal portion in scalloped implant with 2 different connections on loading stress using 3 dimensional finite element analysis. Scalloped implant with 4 different thread pitches (0.4mm, 0.5mm, 0.6, and 0.7mm) in the coronal part was modeled with 2 different implant-abutment connections. Platform matching connection had the same implant and abutment diameter so that they were in flush contact at the periphery while platform mismatching connection had smaller abutment diameter than implant so that their connection was made away from periphery of implant-bone interface. Occlusal loading of 100N force was applied vertically and 30 degree obliquely to all 8 models and the maximum von Mises bone stress was identified. Loading stress as highly concentrated in cortical bone. Platform mismatching scalloped implant with small thread pitch (0.4mm) model had consistently lowest maximum von Mises bone stress in vertical and oblique loads. Platform matching model had lowest maximum von Mises bone stress with 0.6mm thread pitch in vertical load and with 0.4mm thread pitch in oblique load. Platform mismatching connection had important roles in reducing maximum von Mises bone stress. Scalloped implant with smaller coronal thread pitch showed trend of reducing maximum von Mises bone stress under load.
참고문헌 (22)
- Lindquist LW, "A prospective 15-year follow-up study of mandibular fixed prosthesis supported by osseointegrated implants Clinical results and marginal bone loss", Clin Oral Implants Res, 7, pp.329~336 (1996)
- Jemt T, "Osseointegration in the treatment of partially edentulous patients: A preliminary study of 876 consecutively placed fixtures", Int J Oral Maxillofac Implants, 4, pp.211~217 (1989)
- Adell R, "A 15-year study of osseointegrated implants in the treatment of the edentulous jaw", Int J Oral Surg, 10, pp.387~416 (1981)
- Brunski JB, "Biomaterials and biomechanics in dental implant design", Int J Oral Maxillofac Implants, 3(2), pp.85~97 (1988)
- Steigenga JT, "Dental implant design and its relationship to long-term implant success", Implant Dent, 12(4), pp.306~317 (2003)
- Albrektsson T, "State of the art in oral implants", J Clin Periodontol, 18(6), pp.474~481 (1991)
- Whrle, PS, "Nobel perfect esthetic scalloped implant: Rationale for a new design", Clin Implant Dent Relat Res, 5, pp.64~73 (2003)
- Becker WC, "Alveolar bone anatomic profiles as measured from dry skulls. Clinical ramifications", J Clin Periodontol, 24(10), pp.727~731 (1997)
- Nowzari H, "Scalloped dental implants: a retrospective analysis of radiographic and clinical outcomes of 17 Nobel- Perfect implants in 6 patients", Clin Implant Dent Relat Res, 8, pp.1~10 (2006)
- Kan J, "Peri-implant tissue response following immediate provisional restoration of scalloped implants in the esthetic zone: A one-year pilot prospective multi-center study", J Prosthet Dent, 97, pp.109~118 (2007)
- Tymstra N, "Treatment outcome of two adjacent implant crowns with different implant platform designs in the aesthetic zone: a 1-year randomized clinical trial", Clin Periodontol, 38, pp.74~85 (2010)
- Abrahamsson I, "Tissue characteristics at microthreaded implants: An experimental study in dogs", Clin Implant Dent Relat Res, 8, pp.107~113 (2006)
- Lee D, "Effect of microthread on the maintenance of marginal bone level: a 3-year prospective study", Clin Oral Implants Res, 18, pp.465~470 (2007)
- Shin Y, "Radiographic evaluation of marginal bone level around implants with different neck designs after 1 year", The Int J Oral Maxillofac Implants, 21, pp.789~794 (2006)
- Choi KS, "Study of an experimental microthreaded scalloped implant design: proximal bone healing at different interimplant distances in a canine model", Int J Oral Maxillofac Implants, 25, pp.195~202 (2010)
- Chun H, "Influence of implant abutment type on stress distribution in bone under various loading conditions using finite element analysis", Int J Oral Maxillofac Implants, 21, pp.195~202 (2006)
- Merz BR, "Mechanics of the implant-abutment connection: an 8-degree taper compared to a butt joint connection", The Int J Oral Maxillofac Implants, 15, pp.789~794 (2000)
- Abkowitz S, "Titanium in Industry" (1955), Van Nostrand Co Inc
- Katranji A, "Cortical bone thickness in dentate and edentulous human cadavers", J Periodontol, 78, pp.974~978 (2007)
- Misch CE, "A positive correlation between occlusal trauma and peri-implant bone loss: literature support", Implant Dent, 2, pp.108~116 (2005)
- Oh TJ, "The causes of early implant bone loss: myth or science?", J Periodontol, 73, pp.322~333 (2002)
- Choi KS, "Stress distribution of scalloped implant with different microthread and connection configurations using 3 dimensional finite element analysis", Int J Oral Maxillofac Implants