We have developed a multi-channel, multi-frequency EIT system with operating frequency of 10Hz to 500KHz. The number of digital voltmeters using phase-sensitive demodulation can be varied from 8 to 64 and we found that 16 and 32-channels are most practical. This paper describes the design, implementation, and construction of 16 and 32-channel systems. The performance of the system was thoroughly tested and we found that CMRR of the developed voltmeter is about 85dB with 100 unbalancing series resistor. The SNR is greater than 99.6dB and the output impedance of the constant current source is 1 W at least for all frequencies. Imaging experiments using a banana with frequency-dependent conductivity and permittivity show that frequency-difference imaging is possible using the developed system. Future works of animal and human experiments are discussed.
참고문헌 (7)
- D. Holder, "Electrical Impedance Tomography: Methods,History and Applications" (2005), IOP Publishing, Bristol, UK
- K. Boone, "Imaging with electricity report of the European concerted action on impedance tomography", 21, pp.201~232 (1997)
- T. I. Oh, "Development of prototype multi-channel digital EIT system with radially symmetric architecture", 26, pp.215~221 (2005)
- S. Franco, "Design with Operational Amplifiers and Analog Integrated Circuits" (2002), McGraw-Hill
- A. S. Ross, "Current source design for electrical impedance tomography", 24, pp.509~516 (2003)
- B. Baker, "Real Analog Solution for Digital Designers" (2005)
- J. L. Mueller, "A reconstruction algorithm for electrical impedance tomography data collected on rectangular electrode arrays", 46, pp.1379~1386 (1999)