의료용 흡입기 및 분무기의 성능 평가 연구
Efficiency test Evaluation Method for Nebulizer & Medicinal Nonventilatory Nebulizer
2
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Nebulizer is designed to atomize medicinal fluid for patient with small particles(0.5-5㎛) and also able to deliver particles from devices to the lungs when patient inhales air. Several particle size measurements are currently used to size aerosol particles. The most commonly used test is the cascade impactor method in as a standard. But, other methods for comparative particle size distribution data such as the particle size range and reproducibility are acceptable. Therefore, in this study a new test methode is suggested for nonventilatory nebulizer evaluation equipment.
Abstract
Nebulizer is designed to atomize medicinal fluid for patient with small particles(0.5-5㎛) and also able to deliver particles from devices to the lungs when patient inhales air. Several particle size measurements are currently used to size aerosol particles. The most commonly used test is the cascade impactor method in as a standard. But, other methods for comparative particle size distribution data such as the particle size range and reproducibility are acceptable. Therefore, in this study a new test methode is suggested for nonventilatory nebulizer evaluation equipment.
참고문헌 (13)
- Dieter Kὄ.hler, "Wolfgang Fleischer Established Facts in Inhalation Therapy", pp.8~11, Arcis Verlag
- Ryan G, "Influence of nebulizer output, particle size, and method of inhalation", J Allergy Clin Immunol, 67, pp.156~161 (1981)
- Barry PW, "An in vitro analysis of the output of budesonide from different nebulizer", J Allergy Clin Immunol, 104, pp.1168~1173 (1999)
- Keum-Shil Cho, "Measurement of Particle Size and Particle Size Distribution", Polymer Science and Technology, 15(2) (2004)
- "Anaesthetic and respiratory equipment - Nebulizing systems and components, ISO", DIS 27427, 2007
- Kwong WT, "Comparison of nebulized particle size distribution with Malvern laser diffraction analyzer versus Andersen cascade impactor and low flow marple ersonal cascade impactor", J Aerosol Med, 13, pp.303~314 (2001)
- "Reviewer Guidance for nebulizers, metered dose inhalers, spacers and actuators, CDRH final guidance Cover Sheet", ~ (1993)
- F. Abroug, "A Controlled trial of nebulized salbutamol and adrenaline acute severe asthma", Intensive Care Med, 21, pp.18~23 (1995)
- Heyder J, "Deposition of particles in the human respiratory tract in the size range 0.005-15 μm", J Aerosol Sci, 17, pp.811~825 (1986)
- H.S. Mangat, "Nebulized magnesium sulphate versus nebulized salbutamol in acute bronchial asthma: a clinical trial", Eur Respir J, 12, pp.341~344 (1998)
- Bennett P. deBoisblanc, "Randomized trial of the use of heliox as a driving gas for updraft nebulization of bronchodilators in the emergent treatment of acute exacerbations of chronic obstructive pulmonary disease", Crit Care Med, 28(9) (2000), No.9
- Richard N. Channick, "Pulsed Delivery of Inhaled Nitric Oxide to Patients With Primary Pulmonary Hypertension", An Ambulatory Delivery System and Initial Clinical Tests, 109, pp.1545~1549 (1996)
- Kyung-Jung Lee, "A study on the microcomputer based infusion pump", Journal of Biomedical Engineering Research, 005(01), pp.63~68 (1984)