For the patients with bladder dysfunction, measurement of urine volume inside the bladder is very critical to avoid bladder failure. In measuring urine volume inside a bladder, low-resolution 3D ultrasound images are widely used. However, urine volume estimation from 3D ultrasound images is prone to big errors and inconsistency because of low spatial resolution and low signal-to-noise ratio of ultrasound images. We developed a new robust volume estimation algorithm which is not computationally expensive. We tested the algorithm on a lab-built ultrasound bladder phantom and volunteers. The average error rate of the human bladder volume estimation was 5.9% which was better than the commercial machine.

Abstract

For the patients with bladder dysfunction, measurement of urine volume inside the bladder is very critical to avoid bladder failure. In measuring urine volume inside a bladder, low-resolution 3D ultrasound images are widely used. However, urine volume estimation from 3D ultrasound images is prone to big errors and inconsistency because of low spatial resolution and low signal-to-noise ratio of ultrasound images. We developed a new robust volume estimation algorithm which is not computationally expensive. We tested the algorithm on a lab-built ultrasound bladder phantom and volunteers. The average error rate of the human bladder volume estimation was 5.9% which was better than the commercial machine.

참고문헌 (12)

  1. Holmes JH, "Ultrasonic studies of bladder", J Urol, 97, pp.654~663 (1967)
  2. J.M. Walton, "Comparison of real-time ultrasonography and magnetic resonance imaging in the assessment of urinary bladder volume", British J Urol, 78, pp.856~861 (1996)
  3. KA Ung, "Comparison study of portable bladder scanner versus conebeam CT scan for measuring bladder volumes in post-prostatectomy patients undergoing radiotherapy", J Medical Imaging and Radiation Oncology, 58, pp.377~383 (2014)
  4. Bakary Diarra, "Optimized 2D array design for ultrasound imaging", IEEE EURASIP 2012, 2076-1465, pp.2718~2722 (2012)
  5. Pedersen JF, "Residual urine determination by ultrasonic scanning", AJR Am J Roentgenol Ther Nucl Med, 125, pp.474~478 (1975)
  6. J. Kjeldsen-Kragh MD, "Measurement of Residual Urine Volume by Means of Ultrasonic Scanning : A Comparative Study", Paraplegia, 26, pp.192~199 (1988)
  7. Poston GJ, "The accuracy of ultrasound in the measurement of changes in bladder volume", Br J Urol, 55, pp.361~363 (1986)
  8. E A Kiely, "Measurement of bladder volume by real-time ultrasound", British J Urol, 60, pp.33~35 (1987)
  9. ViKram Chalana, "Image enhancement and segmentation of fluid-filled structures in 3D ultrasound images", Proc. of SPIE, 5032, pp.500~506 (2003)
  10. Barbara Levienaise-Obadia, "Adaptive segmentation of ultrasound images", Image and Vision Computing, 17, pp.583~588 (1999)
  11. Chitkasaem Suwanrath, "Three-Dimensional Ultrasonographic Bladder Volume Measurement", J Ultrasound Med, 28, pp.847~854 (2009)
  12. Donka Angelova, "Contour segmentation in 2D ultrasound medical images with particle filtering", Machine Vision and Applications, 22, pp.551~561 (2011)