반치환 슬관절의 설계 평가를 위한 굴곡각도에 따른곡률반경과 접촉응력의 상관도 분석
The Correlation Analysis through the Contact Stress and the Curvature Radius according to Flexion Angle for the Design of Unicompartment Knee Replacement
2
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Conformity between the femoral component and tibial insert within the knee replacement may be measured in frontal or sagittal view, and shows differences in the curvature radius of the femoral component depending on the flexion angle, i.e., curvature radius has a complex effect on contact stress. Therefore, it is essential to confirm how the curvature radius effects contact stress, and provide an important variable to reduce contact stress. This study correlated contact stress with curvature radius measured in frontal and the sagittal views and confirmed the effect of curvature radius for assessment of the Newly Designed Unicompartment Knee Replacement (NDUKR). Finite element models were constructed for NDUKR and Zimmer® Unicompartment High Flex Knee Replacement system (ZUKR), incorporating the curvature radius as measured in either frontal or sagittal view. The femoral component had 1200N of compressive load applied approximately 1.65xbody weight. Contact stress was predicted at flexion angles 0o, 30o, 60o, 90o and 120o, for NDUKR: 42, 47.7, 47.7, 51.2, and 54.1 MPa, and ZUKR: 41.2, 49.5, 53.2, 54.3, and 57.4 MPa, respectively. Correlation analysis showed the influence of curvature radius measured from the sagittal view was larger than for frontal view.
Abstract
Conformity between the femoral component and tibial insert within the knee replacement may be measured in frontal or sagittal view, and shows differences in the curvature radius of the femoral component depending on the flexion angle, i.e., curvature radius has a complex effect on contact stress. Therefore, it is essential to confirm how the curvature radius effects contact stress, and provide an important variable to reduce contact stress. This study correlated contact stress with curvature radius measured in frontal and the sagittal views and confirmed the effect of curvature radius for assessment of the Newly Designed Unicompartment Knee Replacement (NDUKR). Finite element models were constructed for NDUKR and Zimmer® Unicompartment High Flex Knee Replacement system (ZUKR), incorporating the curvature radius as measured in either frontal or sagittal view. The femoral component had 1200N of compressive load applied approximately 1.65xbody weight. Contact stress was predicted at flexion angles 0o, 30o, 60o, 90o and 120o, for NDUKR: 42, 47.7, 47.7, 51.2, and 54.1 MPa, and ZUKR: 41.2, 49.5, 53.2, 54.3, and 57.4 MPa, respectively. Correlation analysis showed the influence of curvature radius measured from the sagittal view was larger than for frontal view.
참고문헌 (16)
- A. J. Bailey, "The role of bone in osteoarthritis", Age Ageing, 30(5), pp.374~378 (2001)
- 한바울, "경골 내 변형률 및 응력 분포 특성 분석을 통한 새로이 개발된 재치환용 인공슬관절의 생체역학적 안정성 평가: 유한요소해석", 의공학회지, 34(1), pp.14~23 (2013), 대한의용생체공학회[DOI]
- Woo Shin Cho, "CHO’s Knee Joint Arthroplasty", pp.263~284 (2015)
- H. D. Clarke, "Component fracture in total knee arthroplasty", Knee, 6(4), pp.261~267 (1999)
- M. S. Austin, "Knee failure mechanisms after total knee arthroplasty", Techniques in Knee Surgery, 3(1), pp.55~59 (2004)
- C. H. Cho, "Finite Element Analysis for the Contact Stress of Ultra-high Molecular Weight Polyethylene in Total Knee Arthroplasty", J. Biomed. Eng. Res, 20(1), pp.37~44 (1999)
- H.G. Willert, "Osteolysis in Alloarthroplasty of the Hip: The Role of Ultra-High Molecular Weight Polyethylene Wear Particles", Clinical Orthopaedics and Related Research, 258, pp.95~107
- J. P. Collier, "Analysis of the Failure of 122 Polyethylene Inserts From Uncemented Tibial Knee Components", Clinical Orthopaedics, 273, pp.232~242 (1991)
- D. L. Bartel, "The Effect of Conformity, Thickness, and Material on Stresses in Ultra-High Molecular Weight Components for Total Joint Replacement", J, Bone and Joint Surgery, 68(7), pp.1041~1051 (1986)
- S. Sathasivam, "The effect of contact area on wear in relation to fixed bearing and mobile bearing knee replacements", J. Biomedical materials research, 58(3), pp.282~290
- A.L. Galvin, "Effect of conformity and contact stress on wear in fixed-bearing total knee prostheses", Journal of biomechanics, 42(12), pp.1898~1902
- Shaw Akizuki, "In Vivo Determination of Kinematics for Subjects Having a Zimmer Unicompartmental High Flex Knee System", Journal of Arthroplasty, 24(6) (2009)
- Liu. C. W., "Evaluation of Implant Positioning Alignment in Medial Unicompartmental Knee Arthroplasty", 2007. 1st International Conference on Bioinformatics and Biomedical Engineering, pp.1324~1327 (2007), IEEE
- "ASTM F2083-12 Standard Specification for Knee Replacement Prosthesis", ~
- I. J. Harrington, "A bioengineering analysis of force actions at the knee in normal and pathological gait", Biomed Eng, 11(5), pp.167~172 (1976)
- 장영웅, "Alterations in Stress Distribution and Micromotion Characteristics due to an Artificial Defect within a Composite Tibia used for Mechanical", Biomechanical Evaluation of Total Knee Arthroplasty, 2015, pp.16(10), 2213~2218[DOI]