본문 바로가기
KERD
논문KCI 등재

온도 조절형 고주파 시스템 및 식염수 분사를 고려한 전극도자절제술용 전극의 수치 모델 개발

Development of Numerical Model of Electrode for Radiofrequency Catheter Ablation Considering Saline Irrigation and Temperature-controlled Radiofrequency System

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Radiofrequency catheter ablation is the interventional therapy that be employed to eliminate cardiac tissue caused by arrhythmias. During radiofrequency catheter ablation, The thrombus can occur at electrode tip if the temperature of tissue and electrode is excess 100oC. To prevent this phenomenon, we investigated numerical model of electrode for radiofrequency catheter ablation considering saline irrigation and temperature-controlled radiofrequency system. The numerical model is based on coupled electric-thermal-flow problem and solved by COMSOL Multiphysics software. The results of the models show that the dimensions of the thermal lesion are increased if the flow rate of the saline irrigation and the set temperature are increased. The surface width characterized to determine the thermal lesion isn’t need to measure in temperature-controlled radiofrequency system due to convective heat transfer by saline irrigation at tissue-electrode interface.

Abstract

Radiofrequency catheter ablation is the interventional therapy that be employed to eliminate cardiac tissue caused by arrhythmias. During radiofrequency catheter ablation, The thrombus can occur at electrode tip if the temperature of tissue and electrode is excess 100oC. To prevent this phenomenon, we investigated numerical model of electrode for radiofrequency catheter ablation considering saline irrigation and temperature-controlled radiofrequency system. The numerical model is based on coupled electric-thermal-flow problem and solved by COMSOL Multiphysics software. The results of the models show that the dimensions of the thermal lesion are increased if the flow rate of the saline irrigation and the set temperature are increased. The surface width characterized to determine the thermal lesion isn’t need to measure in temperature-controlled radiofrequency system due to convective heat transfer by saline irrigation at tissue-electrode interface.

참고문헌 (23)

  1. McRury, I.D, "Ablation for the treatment of arrhythmias", Proceedings of the IEEE, 84(3), pp.404~416 (1996)
  2. Haines, D.E, "The biophysics of radiofrequency catheter ablation in the heart: the importance of temperature monitoring", Pacing and Clinical Electrophysiology, 16(3), pp.586~591 (1993)
  3. Jain, M.K, "Temperature-controlled and constant-power radio-frequency ablation : what affects lesion growth?", IEEE Transactions on Biomedical Engineering, 46(12), pp.1405~1412 (1999)
  4. GUERRA, J. M, "Effects of Open-Irrigated Radiofrequency Ablation Catheter Design on Lesion Formation and Complications : In Vitro Comparison of 6 Different Devices", Journal of cardiovascular electrophysiology, 24(10), pp.1157~1162 (2004)
  5. WOOD, M. A, "Effect of electrode orientation on lesion sizes produced by irrigated radiofrequency ablation catheters", Journal of cardiovascular electrophysiology, 20(11), pp.1262~1268 (2009)
  6. EVERETT IV, T. H, "Safety profiles and lesion size of different radiofrequency ablation technologies : a comparison of large tip, open and closed irrigation catheters", Journal of cardiovascular electrophysiology, 20(3), pp.325~335 (2009)
  7. Demazumder, D, "Biophysics of radiofrequency ablation using an irrigated electrode", Journal of interventional cardiac electrophysiology, 5(4), pp.377~389 (2001)
  8. Yokoyama, K, "Comparison of electrode cooling between internal and open irrigation in radiofrequency ablation lesion depth and incidence of thrombus and steam pop", Circulation, 113(1), pp.11~19 (2006)
  9. Weiss, C, "Radiofrequency catheter ablation using cooled electrodes : impact of irrigation flow rate and catheter contact pressure on lesion dimensions", Pacing and Clinical Electrophysiology, 25(4), pp.463~469 (2002)
  10. González-Suárez, A, "Computational modeling of open-irrigated electrodes for radiofrequency cardiac ablation including blood motion-saline flow interaction", PloS one, 11(3), pp.e0150356~ (2016)
  11. Pérez, J. J, "Electrical and thermal effects of esophageal temperature probes on radiofrequency catheter ablation of atrial fibrillation : results from a computational modeling study", Journal of cardiovascular electrophysiology, 26(5), pp.556~564 (2015)
  12. Jain, M.K, "A three-dimensional finite element model of radiofrequency ablation with blood flow and its experimental validation", Annals of biomedical engineering, 28(9), pp.1075~1084 (2000)
  13. Gopalakrishnan, J, "A mathematical model for irrigated epicardial radiofrequency ablation", Annals of biomedical engineering, 30(7), pp.884~893 (2002)
  14. Gallagher, N, "Contact geometry affects lesion formation in radio-frequency cardiac catheter ablation", PloS one, 8(9), pp.e73242~ (2013)
  15. Tungjitkusolmun, S, "Thermal-electrical finite element modelling for radio frequency cardiac ablation : effects of changes in myocardial properties", Medical and Biological Engineering and Computing, 38(5), pp.562~568 (2000)
  16. González-Suarez, A, "Comparative analysis of different methods of modeling the thermal effect of circulating blood flow during RF cardiac ablation", IEEE Transactions on Biomedical Engineering, 63(2), pp.250~259 (2016)
  17. Pennes, H.H, "Analysis of tissue and arterial blood temperatures in the resting human forearm", Journal of applied physiology, 1(2), pp.93~122 (1948)
  18. Obradovi, M, "Finite element modeling of the endocardial radiofrequency ablation", J Serbian Soc Comput Mech, 4(2), pp.43~53 (2010)
  19. Haines, D.E, "Tissue heating during radiofrequency catheter ablation : a thermodynamic model and observations in isolated perfused and superfused canine right ventricular free wall", Pacing and Clinical Electrophysiology, 12(6), pp.962~976 (1989)
  20. Doss, J. D, "Calculation of electric fields in conductive media", Medical physics, 9(4), pp.566~573 (1982)
  21. Plonsey, R, "Considerations of quasi-stationarity in electrophysiological systems", Bulletin of mathematical Biology, 29(4), pp.657~664 (1967)
  22. MORENO, J, "Morphological and Thermodynamic Comparison of the Lesions Created by 4 Open-Irrigated Catheters in 2 Experimental Models", Journal of cardiovascular electrophysiology, 25(12), pp.1391~1399 (2014)
  23. Nguyen, D. T, "Effect of irrigant characteristics on lesion formation after radiofrequency energy delivery using ablation catheters with actively cooled tips", Journal of cardiovascular electrophysiology, 26(7), pp.792~798 (2015)

관련 논문AI추천

이 간행물의 다른 논문 (10건)