형광과 레이저 스펙클 대조도 이미징을 결합한 실시간 의료영상 시스템 개발

Development of a Real-time Medical Imaging System Combined with Laser Speckle Contrast Imaging and Fluorescence Imaging

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

It is important to differentiate between the target tissue (or organ) and the rest of the tissue before incision during surgery. And when it is necessary to preserve the differentiated tissues, the blood vessels connected to the tissue must be preserved together. Various non-invasive medical imaging methods have been developed for this pur�pose. We aimed to develop a medical imaging system that can simultaneously apply fluorescence imaging using indo�cyanine green (ICG) and laser speckle contrast imaging (LSCI) using laser speckle patterns. We designed to collect images directed to the two cameras on a co-axial optical path and to compensate equal optical path length for two optical designs. The light source used for fluorescence and LSCI the same 785 nm wavelength. This system outputs real-time images and is designed to intuitively distinguish target tissues or blood vessels. This system outputs LSCI images up to 37 fps through parallel processing. Fluorescence for ICG and blood flow in animal models were observed throughout the experiment.

Abstract

It is important to differentiate between the target tissue (or organ) and the rest of the tissue before incision during surgery. And when it is necessary to preserve the differentiated tissues, the blood vessels connected to the tissue must be preserved together. Various non-invasive medical imaging methods have been developed for this pur�pose. We aimed to develop a medical imaging system that can simultaneously apply fluorescence imaging using indo�cyanine green (ICG) and laser speckle contrast imaging (LSCI) using laser speckle patterns. We designed to collect images directed to the two cameras on a co-axial optical path and to compensate equal optical path length for two optical designs. The light source used for fluorescence and LSCI the same 785 nm wavelength. This system outputs real-time images and is designed to intuitively distinguish target tissues or blood vessels. This system outputs LSCI images up to 37 fps through parallel processing. Fluorescence for ICG and blood flow in animal models were observed throughout the experiment.

참고문헌 (13)

  1. Draijer M, "Review of laser speckle contrast techniques for visualizing tissue perfusion", Lasers in Medical Science, 24(4), pp.639~651 (2009)
  2. Briers JD, "Laser speckle contrast analysis (LASCA): a nonscanning, full-field technique for monitoring capillary blood flow", Journal of Biomedical Optics, 1(2), pp.174~179 (1996)
  3. Fercher AF, "Flow visualization by means of single-exposure speckle photography", Optics Communications, 37(5), pp.326~330 (1981)
  4. Briers JD, "Laser Doppler and time-varying speckle: a reconciliation", JOSA A, 13(2), pp.345~350 (1996)
  5. Rege A, "Multiexposure laser speckle contrast imaging of the angiogenic microenvironment", Journal of Biomedical Optics, 16(5), pp.056006~ (2011)
  6. Alander JT, "A review of indocyanine green fluorescent imaging in surgery", International Journal of Biomedical Imaging, 2012, pp.26~ (2012)
  7. Frangioni JV, "In vivo near-infrared fluorescence imaging", Current Opinion in Chemical Biology, 7(5), pp.626~634 (2003)
  8. Nguyen JM, "The use of indocyanine green fluorescence angiography to assess anastomotic perfusion following bowel resection in surgery for gynecologic malignancies-A report of 100 consecutive anastomoses", Gynecologic Oncology, 158(2), pp.402~406 (2020)
  9. Griffiths M, "Indocyanine greenbased fluorescent angiography in breast reconstruction", Gland Surgery, 5(2), pp.133~ (2016)
  10. Hultman M, "A 15.6 frames per second 1-megapixel multiple exposure laser speckle contrast imaging setup", Journal of Biophotonics, 11(2), pp.e201700069~ (2018)
  11. Zölei-Szénási D, "Enhancements on multi-exposure LASCA to reveal information of speed distribution", Journal of the European Optical Society-Rapid publications, 10, pp.15033~ (2015)
  12. Yang O, "Real-time blood flow visualization using the graphics processing unit", Journal of Biomedical Optics, 16(1), pp.016009~ (2011)