Susceptance를 이용한 피부수화도 측정 장비의 개발 및 평가

Development and Evaluation of skin Hydration Measure System using the Suscepance Method

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

In this paper, a novel system is proposed to measure skin hydration using the susceptance method. This system largely consists of a low-voltage (±2.6 V) driven circuit and minimized electrodes of size (5 × 5 mm2). To evaluate the accuracy of the novel system in measuring skin hydration, skin hydration values from105 subjects are measured by the proposed system. The measurements are then compared to those obtained by the golden reference device based on the capacitance method in terms of Intraclass Correlation Coefficient(ICC) and correlation coefficient. All measurements are performed on 7 sites, which are forehead, Crow’s foot, cheek, chin, volar forearm, dorsal forearm, and back of the hand, in a room where the temperature and humidity are maintained at an uniform level of 22 ± 2℃ and 50 ± 5%, respectively. ICC values are above 0.9 (p = 0.001), signifying that the skin hydration values measured by the two methods show a good level of reliability. Correlation coefficient between the two methods is also 0.562 (p = 0.001). Based on these results, it is expected that the proposed system may be applicable in a variety of clinical or cosmetic areas.

Abstract

In this paper, a novel system is proposed to measure skin hydration using the susceptance method. This system largely consists of a low-voltage (±2.6 V) driven circuit and minimized electrodes of size (5 × 5 mm2). To evaluate the accuracy of the novel system in measuring skin hydration, skin hydration values from105 subjects are measured by the proposed system. The measurements are then compared to those obtained by the golden reference device based on the capacitance method in terms of Intraclass Correlation Coefficient(ICC) and correlation coefficient. All measurements are performed on 7 sites, which are forehead, Crow’s foot, cheek, chin, volar forearm, dorsal forearm, and back of the hand, in a room where the temperature and humidity are maintained at an uniform level of 22 ± 2℃ and 50 ± 5%, respectively. ICC values are above 0.9 (p = 0.001), signifying that the skin hydration values measured by the two methods show a good level of reliability. Correlation coefficient between the two methods is also 0.562 (p = 0.001). Based on these results, it is expected that the proposed system may be applicable in a variety of clinical or cosmetic areas.

참고문헌 (15)

  1. John A. McGrath, "The filaggrin story : novel insights into skin-barrier function and disease", TRENDS in Molecular Medicine, 14(1), pp.20~27 (2007), novel insights into skin-barrier function and disease,TRENDS in Molecular Medicine,vol.14 No.1,pp.20-27
  2. Anthony J. Mancini, "The Socioeconomic Impact of Atopic Dermatitis in the United States: A Systematic Review", Pediatric Dermatology, 25(1), pp.1~6 (2008)
  3. Severre Grimnes, "Bioimpedance and Bioelectricity Basics", pp.294~296 (2000), Academic Press
  4. C.W.Blichmann, "Assessment of skin moisture,measurement of electrical conductance,capacitance and trans-epidermal water loss", Acta. Derm. Venereol(Stockh), 68, pp.284~290 (1998)
  5. D. C. Slater, "Instrumental methods for asscssing skin moistu- rization", Cosmet. Toiletries, 102, pp.103~109 (1987)
  6. D. Black, "Seasonal variability in the biophysical properties of stratum corneum from different anatomical sites", Skin Res. Technol, 6, pp.70~76 (2000)
  7. T. Yamamoto, "Analysis for the change of skin impedance", Med. Biol. Eng. Comput, 15, pp.219~227 (1977)
  8. A. O. Barel, "Non-invasive electrical measurements for evaluating the water content of the horny layer: comparison between capacitance and conductance measurements", in: Prediction of Percutaneous Penetration: Methods Measurements Modeling Scott R.C. Guy R.H. Had- graft J. and Boode H.E. (eds) IBC Technical Services Ltd London, 2, pp.46~ (1991)
  9. J. W. Fluhr, "Impact of anatomical location on barrier recovery, surface pH and stratum cornuem hydration after acute barrier disruption", Br. J. Dermatol, 146, pp.770~776 (2002)
  10. Z. Ya Xian, "Number of cell layers of the stratum corneum in normal skin relationship to the anatomical location on the body, age, sex and physical parameters", Arch. Dermatol. Res, 291, pp.555~559 (1999)
  11. O.,G., Martinsen, "On Using Single Frequency Electrical Measurements For Skin Hydration Assessment", Innov. Techn. Biol. Med, 19(5), pp.395~399 (1998)
  12. O.,G., Martinsen, "Measurement of the effect of topical liposome preparations by low frequency electrical susceptance", Innov. Techn. Biol. Med, 17(3), pp.217~222 (1996)
  13. T. Agner, "Comparison of Two Electrical Methods for Measurement of Skin Hydration. An Experimental Study on Irritant Patch Test Reactions", Bioeng. Skin, 4, pp.263~269 (1988)
  14. D.Van Neste, "Comparative Study of Normal and Rough Human Skin Hydration in vivo:Evaluation with Four Different Instruments", J. of Dermatol. Sci, 2, pp.119~124 (1991)
  15. Boyce M. Morrison, "Comparison of Instru- mental Measurements of Skin Hydration", J. Toxicol.-Cut.& Ocular Toxicol, 15(4), pp.305~314 (1996)

관련 논문AI추천

이 간행물의 다른 논문 (10건)