본문 바로가기
KERD
논문KCI 등재

근육 강성도 힘 피드백을 이용한 하지 보조기의 무릎 신전 운동 보조 특성 분석

Analysis on the Assist Characteristics for the Knee Extension Motion of Lower Limb Orthosis Using Muscular Stiffness Force Feedback

0

조회

0

다운로드

2

인용

0

참고문헌

키워드

초록

The lower limb orthosis with a pneumatic rubber actuator, which is intended for the assistance and the enhancement of muscular activities of lower limbs was developed in this study. Compared to other knee extension assistive devices being developed by other researchers, our device is designed especially for the elderly people and intended only for slight assistance so that the subjects can keep their muscular strength. For the effectiveness of system, muscular activities of major muscles in lower limbs during sit-to-stand (STS) and squat motion were measured and analyzed. Subjects were performed the STS and squat motion with and without lower limb orthosis. We made comparison muscular activities between with and without lower limb orthosis. Lower limb orthosis was controlled using muscular stiffness force feedback that is controlled by muscular activities of the measured muscle from force sensor. For analysis of muscular activities, electromyography of the subjects was measured during STS and squat motion, and these were measured using MP 150(BIOPAC Systems, Inc.). Muscles of interest were rectus femoris(RF), vastus lateralis(VL), vastus medialis(VM) and vastus intermedius(VI) muscles in lower limbs of the right side. A biodex dynamometer was used to measure the maximal concentric isokinetic strength of the knee extensors of wearing and not wearing orthosis on right side. The test were performed using the concentric isokinetic mode of test with the velocity set at 60°/s for muscles around the knee joints. The experimental result showed that muscular activities in lower limbs wearing orthosis using muscular stiffness force of a vastus medialis muscle was reduced and knee extension torque of an knee joint wearing lower limb orthosis was increased. With this, we confirmed the effectiveness of the developed lower limb orthosis.

Abstract

The lower limb orthosis with a pneumatic rubber actuator, which is intended for the assistance and the enhancement of muscular activities of lower limbs was developed in this study. Compared to other knee extension assistive devices being developed by other researchers, our device is designed especially for the elderly people and intended only for slight assistance so that the subjects can keep their muscular strength. For the effectiveness of system, muscular activities of major muscles in lower limbs during sit-to-stand (STS) and squat motion were measured and analyzed. Subjects were performed the STS and squat motion with and without lower limb orthosis. We made comparison muscular activities between with and without lower limb orthosis. Lower limb orthosis was controlled using muscular stiffness force feedback that is controlled by muscular activities of the measured muscle from force sensor. For analysis of muscular activities, electromyography of the subjects was measured during STS and squat motion, and these were measured using MP 150(BIOPAC Systems, Inc.). Muscles of interest were rectus femoris(RF), vastus lateralis(VL), vastus medialis(VM) and vastus intermedius(VI) muscles in lower limbs of the right side. A biodex dynamometer was used to measure the maximal concentric isokinetic strength of the knee extensors of wearing and not wearing orthosis on right side. The test were performed using the concentric isokinetic mode of test with the velocity set at 60°/s for muscles around the knee joints. The experimental result showed that muscular activities in lower limbs wearing orthosis using muscular stiffness force of a vastus medialis muscle was reduced and knee extension torque of an knee joint wearing lower limb orthosis was increased. With this, we confirmed the effectiveness of the developed lower limb orthosis.

참고문헌 (15)

  1. K. Kim, "Assist of the Extension Motion of Knee joint on with and without Active Knee Orthosis", Conference on The Korean Society Medical and Biological Engineering (2009)
  2. J. A. Norris, "Effect of augmented plantarflexion power on preferred walking speed and economy in young and older adults", Gait & Posture, 25, pp.620~627 (2007)
  3. K. E. Gorden, "Mechanical performance of artificial pneumatic muscles to power an ankle-foot orthosis", Journal of Biomechanics, 39, pp.1832~1841 (2006)
  4. S. Moromugi, "Muscle stiffness sensor to control an assistance device for the disabled", Artificial Life and Robotics, 8(1), pp.42~45 (2004)
  5. D. P. Ferris, "An Ankle-Foot Orthosis Powered by Artificial Pneumatic Muscles", Journal of Applied Biomechanics, 189~197 (2005)
  6. D. P. Ferris, "An improved powered ankle-foot orthosis using proportional myoelectric control", Gait and Posture, 23(4), pp.425~426 (2006)
  7. K. E. Gordon, "Learning to walk with a robotic ankle exoskeleton", Journal of Biomechanics, 40(12), pp.2636~2644 (2007)
  8. H. Kawamoto, "Power assist method based on Phase Sequence and muscle force condition for HAL", Advanced Robotics, 19(7), pp.717~734 (2005)
  9. S. Lee, "Power assist control for walking aid with HAL-3 based on EMG and impedance adjustment around knee joint", in Proceeding International Conference of Intelligent Robots and Systems (2002)
  10. H. Kawamoto, "Power assist method for HAL-3 using EMG-based feedback controller", Proceeding International Conference on Systems, Man and Cybernetics, pp.1648~1653 (2003)
  11. Yamamoto, K, "Development of pneumatic actuator for powered arm", Proceeding JHPS Autumn Meeting, pp.85~88 (1991)
  12. Yamamoto, K., "Development of Powered Suit for Assisting Nurse Labor", Research Reports of Kanagawa Institute of Technology,Part B, 20, pp.29~43 (1996)
  13. Yamamoto, K., "Development of Power Assisting Suit for Assisting Nurse Labor-Miniaturization of supply system to realize wearable suit-", JSME International Journal,Series C, 46(3), pp.923~930 (2003)
  14. Ishii, M., "A Stand-Alone Wearable Power Assist Suit-Development and Availability-", Journal of Robotics and Mechatronics, 17(5), pp.575~583 (2005)

관련 논문AI추천