Performance Evaluation of the AllCheck hCG Card Point-of-Care Device and Method Comparison with the Alere hCG Cassette Assay for Pregnancy Test

1

조회

0

다운로드

1

인용

0

참고문헌

키워드

초록

Background: Immunochromatographic point-of-care (POC) devices ae widely used by laboratories and lay users for urinary human chorionicgonadotropin (hCG) detection. Performance evaluation of pregnancy POC devices is rarely published. We performed an analytical and clinical vali dation of the newly introduced AllCheck hCG Card assay and compared it with the Alere hCG Cassette comparative assay. Methods: The analytical performance of the assay was evaluated using an international standard material for hCG, as per the protocol recom mended in the Clinical and Laboratory Standards Institute (CLSI) guideline. Clinical validation and comparison study with the comparative method were performed with remnant urine samples from pregnant and non-pregnant women. Results: Probit analysis showed an analytical sensitivity of 15.82 mIU/mL. The precision of the assay was validated at a threshold of 30%. Cross-reactivity with luteinizing hormone, follicle-stimulating hormone, and thyroid-stimulating hormone was not observed. Comparison with the comparative assay showed a negative percent agreement of 100.0% (95% confidence interval [CI]: 92.9%-100.0%) and a positive percent agreement of 96.4% (95% CI: 89.9%-98.8%). Cohen’s kappa value was 0.952 (95% CI: 0.899-1.000). Conclusions: Overall, we validated the performance of the urine hCG POC device and suggest that probit regression is suitable for qualitative tests other than molecular tests. The AllCheck hCG Card device satisfied the demanding standards suggested by the CLSI guideline and was suitable for clinical use.

Abstract

Background: Immunochromatographic point-of-care (POC) devices ae widely used by laboratories and lay users for urinary human chorionicgonadotropin (hCG) detection. Performance evaluation of pregnancy POC devices is rarely published. We performed an analytical and clinical vali dation of the newly introduced AllCheck hCG Card assay and compared it with the Alere hCG Cassette comparative assay. Methods: The analytical performance of the assay was evaluated using an international standard material for hCG, as per the protocol recom mended in the Clinical and Laboratory Standards Institute (CLSI) guideline. Clinical validation and comparison study with the comparative method were performed with remnant urine samples from pregnant and non-pregnant women. Results: Probit analysis showed an analytical sensitivity of 15.82 mIU/mL. The precision of the assay was validated at a threshold of 30%. Cross-reactivity with luteinizing hormone, follicle-stimulating hormone, and thyroid-stimulating hormone was not observed. Comparison with the comparative assay showed a negative percent agreement of 100.0% (95% confidence interval [CI]: 92.9%-100.0%) and a positive percent agreement of 96.4% (95% CI: 89.9%-98.8%). Cohen’s kappa value was 0.952 (95% CI: 0.899-1.000). Conclusions: Overall, we validated the performance of the urine hCG POC device and suggest that probit regression is suitable for qualitative tests other than molecular tests. The AllCheck hCG Card device satisfied the demanding standards suggested by the CLSI guideline and was suitable for clinical use.

참고문헌 (12)

  1. Stenman UH, "The classification, functions and clinical use of different isoforms of HCG", Hum Reprod Update, 12, pp.769~784 (2006)
  2. Birken S, "Structure of the human chorionic gonadotropin beta-subunit fragment from pregnancy urine", Endocrinology, 123, pp.572~583 (1988)
  3. Kovalevskaya G, "Hyperglycosylated HCG expression in pregnancy : cellular origin and clinical applications", Mol Cell Endocrinol, 260-262, pp.237~243 (2007)
  4. Norman RJ, "Relationship between blood and urine concentrations of intact human chorionic gonadotropin and its free subunits in early pregnancy", Obstet Gynecol, 69, pp.590~593 (1987)
  5. Kato Y, "Beta-core fragment is a major form of immunoreactive urinary chorionic gonadotropin in human pregnancy", J Clin Endocrinol Metab, 66, pp.1197~1201 (1988)
  6. Herskovits AZ, "False-negative urine human chorionic gonadotropin testing in the clinical laboratory", Lab Med, 51, pp.86~93 (2020)
  7. Cervinski MA, "Qualitative point-of-care and over-the-counter urine hCG devices differentially detect the hCG variants of early pregnancy", Clin Chim Acta, 406, pp.81~85 (2009)
  8. Griffey RT, ""Hook-like effect"causes false-negative point-of-care urine pregnancy testing in emergency patients", J Emerg Med, 44, pp.155~160 (2013)
  9. Sturgeon C, "Recommendations for validation testing of home pregnancy tests(HPTs)in Europe", Clin Chem Lab Med, 59, pp.823~835 (2021)
  10. User protocol for evaluation of qualitative test performance; approved guideline—second edition, "Clinical and Laboratory Standards Institute", Clinical and Laboratory Standards Institute, ~, pp.CLSI document EP12-A2 (2008)
  11. Kamer SM, "Analytical sensitivity of four commonly used hCG point of care devices", Clin Biochem, 48, pp.448~452 (2015)
  12. Evaluation of detection capability for clinical laboratory measurement procedures; approved guideline—second edition, "Clinical and Laboratory Standards Institute", Clinical and Laboratory Standards Institute, ~, pp.CLSI document EP17-A2 (2012)

관련 논문AI추천

Evaluation of Leukocyte and Bacterial Interference in Point-of-Care Human Chorionic Gonadotropin TestsAnn M. Gronowski, Carey-Ann D. Burnham, John V. Mitsios대한진단검사의학회 · Annals of Laboratory Medicine2013 · 제33권 제6호Performance Evaluation of a Point of Care SelexOnTM B-Type Natriuretic Peptide Immunoassay박해일, 이승옥, 조세운, 최현유, 제갈동욱대한진단검사의학회 · Laboratory Medicine Online2020 · 제10권 제3호임신 초기 임신양상에 따른 혈청 β-hCG의 결과 예측에 의한 희석배수 참고치 설정김윤식, 서영미, 신장용, 유신수대한임상검사과학회 · 대한임상검사과학회지2004 · 제36권 제2호Performance Characteristics of the UniCel DxI 800 Immunoassay for the Maternal Serum Quadruple Test, Including Median Values for Each Week of Gestation, in Korean Women권오헌, 김정호, 박용정, 서보름, 송선미, 이정훈대한진단검사의학회 · Annals of Laboratory Medicine2010 · 제30권 제2호형광 면역크로마토그래피 분석법을 이용한 간편용 혈청 β-hCG 정량 방법의 개발조용, 김광중, 최의열대한산부인과학회 · Obstetrics & Gynecology Science2003 · 제46권 제6호Comparison of the Automated Fluorescent Immunoassay System With Roche Elecsys and Beckman Coulter Access 2 Assays for Anti-Müllerian Hormone MeasurementHan Aera, Kim Sung Eun, Lee Yoo Jin, Suh Borum, Yi Gwang대한진단검사의학회 · Annals of Laboratory Medicine2022 · 제42권 제1호

이 간행물의 다른 논문 (9건)