Comparison of the Automated Fluorescent Immunoassay System With Roche Elecsys and Beckman Coulter Access 2 Assays for Anti-Müllerian Hormone Measurement

2

조회

0

다운로드

2

인용

0

참고문헌

키워드

주제분류병리학

초록

Background: Since 2017, automated assays have been used in most clinical laboratories for anti-Müllerian hormone (AMH) level measurement. We evaluated the analytical performance of the newly developed automated fluorescent immunoassay system (AFIAS) AMH assay (Boditech Med, Gangwon-do, Korea) in comparison with the Roche Elecsys and Beckman Coulter Access 2 AMH assays. Methods: Analytical performance of the AFIAS AMH assay was assessed in terms of linearity, repeatability, and within-laboratory precision (CV%) using human recombinant AMH samples according to the Clinical and Laboratory Standards Institute (CLSI) guidelines EP05 and EP06. Using 293 serum samples collected from an infertility clinic, the AMH levels were compared across AFIAS, Elecsys, and Access 2 AMH assays according to the CLSI EP09 guidelines. Results: The AFIAS AMH assay results were linear across the measurement range of 0.420–72.386 pmol/L AMH, with repeatability of 6.341%. CV% of the AFIAS AMH assay for three levels of control, 1.786, 7.143, and 56.857 pmol/L, were 5.801%, 5.714%, and 6.228%, respectively. The results of the three AMH assays showed strong correlation: AFIAS and Elecsys [slope, 1.055 (95% confidence interval (CI), 1.022–1.088) and Spearman’s rho, 0.978 (95% CI, 0.973–0.983)], Elecsys and Access 2 [slope, 0.813 (95% CI, 0.791–0.834) and Spearman’s rho, 0.986 (95% CI, 0.983–0.989)], and AFIAS and Access 2 [slope, 0.836 (95% CI, 0.821–0.853) and Spearman’s rho, 0.984 (95% CI, 0.980– 0.988)]. Conclusions: The AFIAS AMH assay may be an alternative to the Roche Elecsys and Beckman Coulter Access 2 AMH assays.

Abstract

Background: Since 2017, automated assays have been used in most clinical laboratories for anti-Müllerian hormone (AMH) level measurement. We evaluated the analytical performance of the newly developed automated fluorescent immunoassay system (AFIAS) AMH assay (Boditech Med, Gangwon-do, Korea) in comparison with the Roche Elecsys and Beckman Coulter Access 2 AMH assays. Methods: Analytical performance of the AFIAS AMH assay was assessed in terms of linearity, repeatability, and within-laboratory precision (CV%) using human recombinant AMH samples according to the Clinical and Laboratory Standards Institute (CLSI) guidelines EP05 and EP06. Using 293 serum samples collected from an infertility clinic, the AMH levels were compared across AFIAS, Elecsys, and Access 2 AMH assays according to the CLSI EP09 guidelines. Results: The AFIAS AMH assay results were linear across the measurement range of 0.420–72.386 pmol/L AMH, with repeatability of 6.341%. CV% of the AFIAS AMH assay for three levels of control, 1.786, 7.143, and 56.857 pmol/L, were 5.801%, 5.714%, and 6.228%, respectively. The results of the three AMH assays showed strong correlation: AFIAS and Elecsys [slope, 1.055 (95% confidence interval (CI), 1.022–1.088) and Spearman’s rho, 0.978 (95% CI, 0.973–0.983)], Elecsys and Access 2 [slope, 0.813 (95% CI, 0.791–0.834) and Spearman’s rho, 0.986 (95% CI, 0.983–0.989)], and AFIAS and Access 2 [slope, 0.836 (95% CI, 0.821–0.853) and Spearman’s rho, 0.984 (95% CI, 0.980– 0.988)]. Conclusions: The AFIAS AMH assay may be an alternative to the Roche Elecsys and Beckman Coulter Access 2 AMH assays.

참고문헌 (18)

  1. Carmina E, "Yen and Jaffe’s reproductive endocrinology", pp.900~901 (2018), Elsevier
  2. Lie Fong S, "Serum anti-Müllerian hormone levels in healthy females : a nomogram ranging from infancy to adulthood", J Clin Endocrinol Metab, 97, pp.4650~4655 (2012)
  3. de Vet A, "Antimüllerian hormone serum levels : a putative marker for ovarian aging", Fertil Steril, 77, pp.357~362 (2002)
  4. van Rooij IA, "Serum anti-Müllerian hormone levels : a novel measure of ovarian reserve", Hum Reprod, 17, pp.3065~3071 (2002)
  5. Lie Fong S, "Anti-Müllerian hormone: a marker for oocyte quantity, oocyte quality and embryo quality?", Reprod Biomed Online, 16, pp.664~670 (2008)
  6. Fanchin R, "Serum anti-Mullerian hormone is more strongly related to ovarian follicular status than serum inhibin B, estradiol, FSH and LH on day 3", Hum Reprod, 18, pp.323~327 (2003)
  7. van Rooij IA, "Serum antimullerian hormone levels best reflect the reproductive decline with age in normal women with proven fertility : a longitudinal study", Fertil Steril, 83, pp.979~987 (2005)
  8. Broer SL, "AMH and AFC as predictors of excessive response in controlled ovarian hyperstimulation : a meta-analysis", Hum Reprod Update, 17, pp.46~54 (2011)
  9. Kumar A, "Development of a second generation anti-Müllerian hormone(AMH)ELISA", J Immunol Methods, 362, pp.51~59 (2010)
  10. Gassner D, "First fully automated immunoassay for anti-Müllerian hormone", Clin Chem Lab Med, 52, pp.1143~1152 (2014)
  11. Pearson K, "Assessment of the access AMH assay as an automated, high-performance replacement for the AMH generation II manual ELISA", Reprod Biol Endocrinol, 14, pp.8~ (2016)
  12. Pastuszek E, "New AMH assay allows rapid point of care measurements of ovarian reserve", Gynecol Endocrinol, 33, pp.638~643 (2017)
  13. "Guideline on bioanalytical method validation. EMEA/CHMP/EWP/192217/2009 Rev. 1 Corr. 2**" (2011)[URL]
  14. CLSI, "Evaluation of the linearity of quantitative measurement procedures: A statistical approach; Approved guideline" (2020), Clinical and Laboratory Standards Institute
  15. CLSI, "Evaluation of precision of quantitative measurement procedures; Approved guideline" (2014), Clinical and Laboratory Standards Institute
  16. CLSI, "Measurement procedure comparison and bias estimation using patient samples; Approved guideline" (2018), Clinical and Laboratory Standards Institute
  17. Dewailly D, "The physiology and clinical utility of anti-Müllerian hormone in women", Hum Reprod Update, 20, pp.370~385 (2014)
  18. Nelson SM, "Two new automated, compared with two enzyme-linked immunosorbent, antimüllerian hormone assays", Fertil Steril, 104, pp.1016~1021 (2015)

관련 논문AI추천

항뮬러관 호르몬검사 시약 Elecsys AMH plus의 성능 및 비오틴 간섭평가김솔잎, 전사일, Ji Misuk, Lee Seunghoo, Lee Woochang, Ryu Hyeseong, Yu Joonsang대한진단검사의학회 · Laboratory Medicine Online2024 · 제14권 제2호Age-Specific Reference Values of Anti-Müllerian Hormone in Korean Women with Normal Menstruation: Time Trend and Clinical Suitability (2015–2021)김희정, 박진영, 한미순대한진단검사의학회 · Laboratory Medicine Online2022 · 제12권 제1호Age Group-specific Reference Intervals for the Elecsys Anti-Müllerian Hormone Assay in Healthy Korean Women: a Nationwide Population-based StudyChun Sail, Ji Misuk, Kim Hyun-Ki, Kim Kwang-Rae, Lee Woochang, Min Won-Ki, Yun Yeo-Min대한진단검사의학회 · Annals of Laboratory Medicine2022 · 제42권 제6호규칙적인 생리주기를 갖는 2,879명의 한국 여성에서의 나이에 따른 anti-Müllerian hormone 수치 분포에 관한 연구김미경, 김미라, 김영주, 박동수, 성석주, 송태종, 윤보성, 이지은대한산부인과학회 · Obstetrics & Gynecology Science2012 · 제55권 제12호Differentiation between polycystic ovary syndrome and polycystic ovarian morphology by means of an anti-Müllerian hormone cutoff value성연아, 송도경, 오지영, 이혜진대한내과학회 · The Korean Journal of Internal Medicine2017 · 제32권 제4호Preoperative Serum Anti-Mullerian Hormone Level in Women with Ovarian Endometrioma and Mature Cystic Teratoma김석현, 김주영, 서창석, 지병철연세대학교의과대학 · Yonsei Medical Journal2013 · 제54권 제4호

이 간행물의 다른 논문 (15건)