본문 바로가기
KERD
논문KCI 등재

Optimization of image reconstruction method for dual-particle time-encode imager through adaptive response correction

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류원자력공학

초록

Time-encoded imagers (TEI) are important class of instruments to search for potential radioactive sources to prevent illicit transportation and trafficking of nuclear materials and other radioactive sources. The energy of the radiation cannot be known in advance due to the type and shielding of source is unknown in practice. However, the response function of the time-encoded imagers is related to the energy of neutrons or gamma-rays. An improved image reconstruction method based on MLEM was proposed to correct for the energy induced response difference. In this method, the count vector versus time was first smoothed. Then, the preset response function was adaptively corrected according to the measured counts. Finally, the smoothed count vector and corrected response were used in MLEM to reconstruct the source distribution. A one-dimensional dual-particle time-encode imager was developed and used to verify the improved method through imaging an AmeBe neutron source. The improvement of this method was demonstrated by the image reconstruction results. For gamma-ray and neutron images, the angular resolution improved by 17.2% and 7.0%; the contrast-to-noise ratio improved by 58.7% and 14.9%; the signal-to-noise ratio improved by 36.3% and 11.7%, respectively.

Abstract

Time-encoded imagers (TEI) are important class of instruments to search for potential radioactive sources to prevent illicit transportation and trafficking of nuclear materials and other radioactive sources. The energy of the radiation cannot be known in advance due to the type and shielding of source is unknown in practice. However, the response function of the time-encoded imagers is related to the energy of neutrons or gamma-rays. An improved image reconstruction method based on MLEM was proposed to correct for the energy induced response difference. In this method, the count vector versus time was first smoothed. Then, the preset response function was adaptively corrected according to the measured counts. Finally, the smoothed count vector and corrected response were used in MLEM to reconstruct the source distribution. A one-dimensional dual-particle time-encode imager was developed and used to verify the improved method through imaging an AmeBe neutron source. The improvement of this method was demonstrated by the image reconstruction results. For gamma-ray and neutron images, the angular resolution improved by 17.2% and 7.0%; the contrast-to-noise ratio improved by 58.7% and 14.9%; the signal-to-noise ratio improved by 36.3% and 11.7%, respectively.

참고문헌 (18)

  1. H. Al Hamrashdi, "Passive gamma-ray and neutron imaging systems for national security and nuclear non-proliferation in controlled and uncontrolled detection areas : review of past and current status", Sensors, 19, pp.2638~ (2019)
  2. R. C. Runkle, "Neutron sensors and their role in nuclear nonproliferation", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 652, pp.37~40 (2011)
  3. R. C. Runkle, "Securing special nuclear material : recent advances in neutron detection and their role in nonproliferation", J. Appl. Phys, 108, pp.13~ (2010)
  4. A. Glick, "Deployment of a double scatter system for directional detection of background neutron radiation", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 992, pp.165029~ (2021)
  5. A. Poitrasson-Riviere, "Dual-particle imaging system based on simultaneous detection of photon and neutron collision events", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 760, pp.40~45 (2014)
  6. R. S. Woolf, "Fast-neutron, coded-aperture imager", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 784, pp.398~404 (2015)
  7. "The deployable fast-neutron coded-aperture imager: demonstration of locating one or more sources in three dimensions" (2013)[URL]
  8. N. P. Shah, "Improved localization precision and angular resolution of a cylindrical, time-encoded imaging system from adaptive detector movements", IEEE Trans. Nucl. Sci, 68, pp.410~425 (2021)
  9. J. Mueller, "Passive one-dimensional self-transmission imaging of subcritical metallic plutonium assemblies", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 903, pp.277~286 (2018)
  10. N. Shah, "Adaptive Imaging with a Cylindrical, Time-Encoded Imaging System" (2020)
  11. N. P. Shah, "Design and construction of a 1-D, cylindrical, dual-particle, time-encoded imaging system", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 954, pp.161785~ (2020)
  12. A. Barzilov, "Time encoded imaging of neutrons and photons using CLYC detector equipped with a dual mode collimator", Sensors & Transducers, 229, pp.78~83 (2019)
  13. J. Brennan, "Demonstration of two-dimensional timeencoded imaging of fast neutrons", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 802, pp.76~81 (2015)
  14. X. Liang, "Self-supporting design of a time-encoded aperture, gamma-neutron imaging system", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 951, pp.162964~ (2020)
  15. J. Brennan, "Source detection at 100 meter standoff with a time-encoded imaging system", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 877, pp.375~383 (2018)
  16. B. Kowash, "A unified near-and far-field imaging model for rotating modulation collimators", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 637, pp.178~184 (2011)
  17. P. Marleau, "Time encoded fast neutron", gamma imager for large standoff SNM detection, 2011 (2011), 591~595[DOI]
  18. R. J. Olesen, "Maximum likelihood reconstructions for rotating scatter mask imaging", Radiat. Meas, 137, pp.106441~ (2020)

관련 논문AI추천

Comparative study of the pulse shape discrimination (PSD) performance of pixelated stilbene and plastic scintillator (EJ-276) arrays for a coded-aperture-based hand-held dual-particle imager부지환, 정만희한국원자력학회 · Nuclear Engineering and Technology2023 · 제55권 제5호Pulse shape discrimination using a stilbene scintillator array coupled to a large-area SiPM array for hand-held dual particle imager applications부지환, 정만희, Hammig Mark D.한국원자력학회 · Nuclear Engineering and Technology2023 · 제55권 제2호Development of a dual-mode energy-resolved neutron imaging detector: High spatial resolution and large field of viewChen Yuanbo, Jiang Xingfen, Li Jiajie, Qiu Yongxiang, Shen Peixun, Song Yushou, Sun Zhijia, Tan Jinhao, Teng Haiyun, Wang Songlin, Wang Xiuku, Wei Yadong, Xia Yuanguang, Yang Jianqing, Yang Wenqin, Yu Li, Zhang Junrong, Zhao Yubin, Zhou Jianrong, Zhou Xiaojuan, Zhuang Jian, Zhu Lin한국원자력학회 · Nuclear Engineering and Technology2024 · 제56권 제7호Image reconstruction algorithm for momentum dependent muon scattering tomographyBae JungHyun, Chatzidakis Stylianos, Montgomery Rose한국원자력학회 · Nuclear Engineering and Technology2024 · 제56권 제5호CdZnTe semiconductor-based dual imager combining collimatorless and Compton imaging: Monte Carlo simulation김영학, 이원호한국원자력학회 · Nuclear Engineering and Technology2024 · 제56권 제10호Super-spatial resolution method combined with the maximum-likelihood expectation maximization (MLEM) algorithm for alpha imaging detector김건아, 김종국, 송강현, 임일한한국원자력학회 · Nuclear Engineering and Technology2022 · 제54권 제6호

이 간행물의 다른 논문 (41건)