본문 바로가기
KERD
논문KCI 등재

Determination of the number of 235U target nuclei in the irregular target using a fission time projection chamber

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류원자력공학

초록

Based on multiple measurements of ionization loss, the Time Projection Chamber (TPC) combines strong tracking ability with particle identification ability in a large momentum range, which is an important advantage of TPC detection technology over traditional ionization measurement technology. According to these two characteristics of TPC, applying it to the measurement of fission cross-section can greatly improve the measurement accuracy. During the measurement of the fission cross-section, the number of target nuclei is required to be accurately measured. So this paper introduces a method for measuring the number of 235U target nuclei using a fission TPC system. The measurement result agrees with the reference value, and relative error is around 1 %.

Abstract

Based on multiple measurements of ionization loss, the Time Projection Chamber (TPC) combines strong tracking ability with particle identification ability in a large momentum range, which is an important advantage of TPC detection technology over traditional ionization measurement technology. According to these two characteristics of TPC, applying it to the measurement of fission cross-section can greatly improve the measurement accuracy. During the measurement of the fission cross-section, the number of target nuclei is required to be accurately measured. So this paper introduces a method for measuring the number of 235U target nuclei using a fission TPC system. The measurement result agrees with the reference value, and relative error is around 1 %.

참고문헌 (13)

  1. H. J. Hilke, "Time projection chambers", Rep. Prog. Phys, 73, pp.116201~ (2010)[DOI]
  2. M. Heffher, "A time projection chamber for high accuracy and precision fission cross section measurements", Nucl. Instrum. Methods Phys. Res., Sect. A, 759, pp.50~64 (2014)[DOI]
  3. J.W. Kim, "A preliminary study on time projection chamber simulation for fission cross section measurements with Geant4", Transactions of the Korean Nuclear Society Spring Meeting (2014)
  4. J. C. Kim, "Neutron science facility for neutron time-offlight and fission cross-section measurements at RAON", J. Kor. Phys. Soc, 66, pp.478~485 (2015)[DOI]
  5. G.H. Zhang, "Determination of 238U nucleus number using grid ionization chamber method and small solid angle method", Atomic Energy Sci. Technol, 34, pp.487~492 (2000)
  6. S. O. Troja, "Alpha and beta dose-rate determination using a gas proportional counter", Radiat. Meas, 24, pp.297~308 (1995)[DOI]
  7. M. T. Yang, "The present status and develop current of α particle detect instrument", Nucl. Electron. Detect. Technol, 31, pp.1198~1201 (2011)[DOI]
  8. H. Bai, "Determination of the number of 238U target nuclei and simulation of the neutron induced fission fragment energy spectrum", Appl. Radiat. Isot, 152, pp.180~187 (2019)[DOI]
  9. J. Zhang, "Simulation and data analysis of a fission time projection chamber based on Monte Carlo method", J. Instrum, 18, pp.P03043~ (2023)[DOI]
  10. S. F. Biagi, "Monte Carlo simulation of electron drift and diffusion in counting gases under the influence of electric and magnetic fields", Nucl. Instrum. Methods Phys. Res., Sect. A, 421, pp.234~240 (1999)[DOI]
  11. C. Dalitz, "Iterative Hough transform for line detection in 3D point clouds", Image Process. Line, 7, pp.184~196 (2017)[DOI]
  12. Q. Y. Yu, "Uncertainty analysis of measuring thermal neutron fluence rate by using 235U fission ionization chamber and gold foil-activation methods", Atomic Energy Sci. Technol, 42, pp.233~238 (2008)
  13. Y. Y. Yan, "Development of a Time Projection Chamber (TPC) for High Precision Fission Cross-Section Measurements", pp.2018

관련 논문AI추천