본문 바로가기
KERD
논문KCI 등재

Development of a prototype TL/OSL reader for on-site use in a large-scale radiological accident

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류원자력공학

초록

This study presents the development and characterization of a prototype TL/OSL reader for the retrospective dose assessment of individuals in radiological emergencies. The reader is portable, semi-automatic, and capable of accurate measurements. The dimension of the reader is 25 × 25 × 37 cm3 and the weight is about 15 kg. The reader consists of a sample moving stage, a heating module, an optical stimulation module, a detection module, a data acquisition (DAQ) unit, a nitrogen gas control module, and a PC with a GUI program. The reader has three measurement modes: TL, CW_OSL, and custom mode. The reader was characterized using commercial thermal luminescence dosimeters (TLD, LiF:Mg,Cu,Si) and optically stimulated dosimeters (OSLD, Al2O3:C), as well as fortuitous materials, such as display glasses and resistors of mobile phone. The results showed that the reader is capable of measuring signals with a detection limit of up to 0.02 mGy using a commercial dosimeter. In the dose recovery test using fortuitous materials, the reconstructed doses obtained three days post-irradiation closely aligned with the initially administered doses. As a result, this study suggests that the developed TL/OSL reader is a promising instrument for emergency dose assessment at accident sites

Abstract

This study presents the development and characterization of a prototype TL/OSL reader for the retrospective dose assessment of individuals in radiological emergencies. The reader is portable, semi-automatic, and capable of accurate measurements. The dimension of the reader is 25 × 25 × 37 cm3 and the weight is about 15 kg. The reader consists of a sample moving stage, a heating module, an optical stimulation module, a detection module, a data acquisition (DAQ) unit, a nitrogen gas control module, and a PC with a GUI program. The reader has three measurement modes: TL, CW_OSL, and custom mode. The reader was characterized using commercial thermal luminescence dosimeters (TLD, LiF:Mg,Cu,Si) and optically stimulated dosimeters (OSLD, Al2O3:C), as well as fortuitous materials, such as display glasses and resistors of mobile phone. The results showed that the reader is capable of measuring signals with a detection limit of up to 0.02 mGy using a commercial dosimeter. In the dose recovery test using fortuitous materials, the reconstructed doses obtained three days post-irradiation closely aligned with the initially administered doses. As a result, this study suggests that the developed TL/OSL reader is a promising instrument for emergency dose assessment at accident sites

참고문헌 (20)

  1. Methods for Initial-phase Assessment of Individual Doses Following Acute Exposure to Ionizing Radiation, "ICRU", ~, pp.ICRU Report 94 (2019)
  2. I. Fiedler, "Thermoluminescence of chip inductors from mobile phones for retrospective and accident dosimetry", Radiat. Meas, 46, pp.1862~1865 (2011)
  3. E. L. Inrig, "Optically stimulated luminescence of electronic components for forensic, retrospective, and accident dosimetry", Radiat. Meas, 43, pp.726~730 (2008)
  4. C. Bassinet, "Radiation accident dosimetry : TL properties of mobile phone screen glass", Radiat. Meas, 71, pp.461~465 (2014)
  5. M. Discher, "Thermoluminescence of glass display from mobile phones for retrospective and accident dosimetry", Radiat. Meas, 53–54, pp.12~21 (2013)
  6. A. Kandemir, "Thermoluminescence studies of SIM card chips used in mobile communication providers in Turkey", Radiat. Phys. Chem, 149, pp.84~89 (2018)
  7. C. Woda, "On the use of OSL of wire-bond chip card modules for retrospective and accident dosimetry", Radiat. Meas, 44, pp.548~553 (2009)
  8. E. A. Ainsbury, "Review of retrospective dosimetry techniques for external ionising radiation exposures", Radiat. Protect. Dosim, 147, pp.573~592 (2011)
  9. P. Fattibene, "Reflections on the future developments of research in retrospective physical dosimetry", Physics Open, 14, pp.100132~ (2023)
  10. H. Kim, "Thermoluminescence of AMOLED substrate glasses in recent mobile phones for retrospective dosimetry", Radiat. Meas, 122, pp.53~56 (2019)
  11. C. Bassinet, "Retrospective radiation dosimetry using OSL of electronic components : results of an inter-laboratory comparison", Radiat. Meas, 71, pp.475~479 (2014)
  12. H. Kim, "Dose re-evaluation in personal dosimetry using the phototransferred thermoluminescence method of LiF:Mg,Cu,Si TLD", Radiat. Meas, 118, pp.20~25 (2018)
  13. G. L. Long, "Limit of detection. A closer look at the IUPAC definition", Anal. Chem, 55, pp.712A~724A (2008)
  14. Retrospective dosimetry for the population in emergency situations, Final report, "C. Woda", Bundesamt fuer Strahlenschutz, ~ (2012)
  15. H. Kim, "Thermally assisted optically stimulated luminescence protocol of mobile phone substrate glasses for accident dosimetry", Radiat. Meas, 146, pp.106696~ (2021)
  16. M. Discher, "Evaluation of physical retrospective dosimetry methods in a realistic accident scenario : results of a field test", Radiat. Meas, 142, pp.106544~ (2021)
  17. H. Kim, "A small-scale realistic inter-laboratory accident dosimetry comparison using the TL", OSL from mobile phone components, 2022, pp.150, 106696~
  18. K. Munyikwa, "The potential of portable luminescence readers in geomorphological investigations : a review", Earth Surf. Process. Landforms, 46, pp.131~150 (2021)
  19. D. C. Sanderson, "Using simple portable OSL measurements and laboratory characterisation to help understand complex and heterogeneous sediment sequences for luminescence dating", Quat. Geochronol, 5, pp.299~305 (2010)
  20. C. L. Larsson, "Portable Optically Stimulated Luminescence Detector for Forensics and Retrospective Dosimetry", ~ (2011)

관련 논문AI추천

이 간행물의 다른 논문 (49건)