6 MV X-band linear accelerator for stereotactic body radiation therapy
2
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
This paper presents the development process of a high-dose X-band linear accelerator (LINAC) for radiotherapy machines, specifically focusing on stereotactic body radiation therapy (SBRT). The development process includes a multi-physics design, approach for the radio frequency cavity in the LINAC, with an emphasis on minimizing beam size and maximizing shunt impedance of the LINAC. Commissioning tests were conducted to evaluate the performance of the LINAC, which involved measurements of beam energy, beam size, beam current, and X-ray dose rate.
Abstract
This paper presents the development process of a high-dose X-band linear accelerator (LINAC) for radiotherapy machines, specifically focusing on stereotactic body radiation therapy (SBRT). The development process includes a multi-physics design, approach for the radio frequency cavity in the LINAC, with an emphasis on minimizing beam size and maximizing shunt impedance of the LINAC. Commissioning tests were conducted to evaluate the performance of the LINAC, which involved measurements of beam energy, beam size, beam current, and X-ray dose rate.
참고문헌 (29)
- S. Soderstr ¨ om, "Aspects on the optimal photon beam energy for radiation therapy", Acta Oncol, 38(2), pp.179~187 (1999)
- G. Solaiappan, "Influence of photon beam energy on IMRT plan quality for radiotherapy of prostate cancer", Rep. Practical Oncol. Radiother, 14(1), pp.18~31 (2009)
- L. Wang, "Dosimetric advantage of using 6 MV over 15 MV photons in conformal therapy of lung cancer : Monte Carlo studies in patient geometries", J. Appl. Clin. Med. Phys, 3(1), pp.51~59 (2002)
- S. W. Shin, "Design of 6 MeV X-band electron linac for dual-head gantry radiotherapy system", J. Kor. Phys. Soc, 71, pp.1048~1055 (2017)
- S. W. Shin, "Measurement of characteristic of X-band RF cavity for 6 MeV electron linac", J. Kor. Phys. Soc, 72, pp.818~825 (2018)
- S. H. Lee, "X-band Linac for a 6 MeV dual-head radiation therapy gantry", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 852, pp.40~45 (2017)
- J. H. Kim, "Novel biological strategies to enhance the radiation therapeutic ratio", Radiation Oncology Journal, 36(3), pp.172~ (2018)
- J. Y. Luh, "ACR–ASTRO practice parameter for image-guided radiation therapy(IGRT)", Am. J. Clin. Oncol, 43(7), pp.459~468 (2020)
- M. S. Kim, "The first step towards a respiratory motion prediction for natural-breathing by using a motion generator", J. Kor. Phys. Soc, 70, pp.621~628 (2017)
- J. M. Brown, "The tumor radiobiology of SRS and SBRT:are more than the 5 Rs involved?", Int. J. Radiat. Oncol. Biol. Phys, 88(2), pp.254~262 (2014)
- C. L. Tsai, "How to improve therapeutic ratio in radiotherapy of HCC", Liver Cancer, 5(3), pp.210~220 (2016)
- J.D. Zindler, "Increasing the therapeutic ratio of stereotactic ablative radiotherapy by individualized isotoxic dose prescription", J. Natl. Cancer Inst, 108(2), pp.djv305~ (2016)
- H. M. Petroccia, "Spine SBRT with Halcyon™: plan quality, modulation complexity, delivery accuracy, and speed", Front. Oncol, 9, pp.319~ (2019)
- P. Bonomo, "Stereotactic body radiotherapy with CyberKnife for cardiac malignancies", Tumori Journal, 101(3), pp.294~297 (2015)
- B. T. Huang, "Monitor unit optimization in stereotactic body radiotherapy for small peripheral non-small cell lung cancer patients", Sci. Rep, 5(1), pp.18453~ (2015)
- N. T. Sebastian, "Stereotactic body radiation therapy (SBRT) for early stage non-small cell lung cancer (NSCLC): contemporary insights and advances", J. Thorac. Dis, 10(Suppl 21), pp.S2451~ (2018)
- M. Shi, "First-year experience of SBRT", IMRT treatment using a novel biology guided radiotherapy machine, 2023, 101300~
- C. Tang, "Electron Linacs for cargo inspection and other industrial applications", Power, 10(44kV), pp.11~ (2009)
- H. B. Zhao, "The improvement of the output performance of a low energy medical standing wave accelerator", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 356(2–3), pp.552~557 (1995)
- "Poisson/Superfish"[URL]
- "A Space Charge Tracking Algorithm (ASTRA)"[URL]
- "CST Microwave Studio"[URL]
- Y. Kamino, "Development of an ultrasmall-band linear accelerator guide for a four-dimensional image-guided radiotherapy system with a gimbaled x-ray head", Med. Phys, 34(5), pp.1797~1808 (2007)
- J. Mondal, "1 MeV standing wave linac", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 795, pp.343~350 (2015)
- H. Yang, "Transverse RF focusing in bunching cells for standing-wave linac", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 703, pp.145~151 (2013)
- J. Gao, "Analytical formula for the coupling coefficient β of a cavity-waveguide coupling system", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 309(1–2), pp.5~10 (1991)