본문 바로가기
KERD
논문KCI 등재

Determination of counting efficiency considering the biodistribution of 131I activity in the whole-body counting measurement

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류원자력공학

초록

Whole-body counters are widely used to assess internal contamination after a nuclear accident. However, it is difficult to determine radioiodine activity due to limitations in conventional calibration phantoms. Inhaled or ingested radioiodine is heterogeneously distributed in the human body, necessitating time-dependent biodistribution for the assessment of the internal contamination caused by the radioiodine intake. This study aims at calculating counting efficiencies considering the biodistribution of 131I in whole-body counting measurement. Monte Carlo simulations with computational human phantoms were performed to calculate the whole-body counting efficiency for a realistic radioiodine distribution after its intake. The biodistributions of 131I for different age groups were computed based on biokinetic models and applied to age- and gender-specific computational phantoms to estimate counting efficiency. After calculating the whole-body counting efficiencies, the efficiency correction factors were derived as the ratio of the counting efficiencies obtained by considering a heterogeneous biodistribution of 131I over time to those obtained using the BOMAB phantom assuming a homogeneous distribution. Based on the correction factors, the internal contamination caused by 131I can be assessed using wholebody counters. These correction factors can minimize the influence of the biodistribution of 131I in whole-body counting measurement and improve the accuracy of internal dose assessment.

Abstract

Whole-body counters are widely used to assess internal contamination after a nuclear accident. However, it is difficult to determine radioiodine activity due to limitations in conventional calibration phantoms. Inhaled or ingested radioiodine is heterogeneously distributed in the human body, necessitating time-dependent biodistribution for the assessment of the internal contamination caused by the radioiodine intake. This study aims at calculating counting efficiencies considering the biodistribution of 131I in whole-body counting measurement. Monte Carlo simulations with computational human phantoms were performed to calculate the whole-body counting efficiency for a realistic radioiodine distribution after its intake. The biodistributions of 131I for different age groups were computed based on biokinetic models and applied to age- and gender-specific computational phantoms to estimate counting efficiency. After calculating the whole-body counting efficiencies, the efficiency correction factors were derived as the ratio of the counting efficiencies obtained by considering a heterogeneous biodistribution of 131I over time to those obtained using the BOMAB phantom assuming a homogeneous distribution. Based on the correction factors, the internal contamination caused by 131I can be assessed using wholebody counters. These correction factors can minimize the influence of the biodistribution of 131I in whole-body counting measurement and improve the accuracy of internal dose assessment.

참고문헌 (31)

  1. O. Kurihara, "Effective population monitoring for determination of the thyroidal radioiodine content of the public following a nuclear accident in Japan", Environmental Advances, 8, pp.100206~ (2022)
  2. O. Kurihara, "Experiences of population monitoring using whole-body counters in response to the Fukushima nuclear accident", Health Phys, 115(2), pp.259~274 (2018)
  3. E. Kim, "Early intake of radiocesium by residents living near the TEPCO Fukushima Dai-ichi nuclear power plant after the accident. Part 1 : internal doses based on whole-body measurements by NIRS", Health Phys, 111(5), pp.451~464 (2016)
  4. American National Standard: specifications for the bottle manikin absorption phantom, "Health Physics Society", 35~, pp.ANSI (1999), HPS N13
  5. 박민석, "Virtual calibration of whole-body counters to consider the size dependency of counting efficiency using Monte Carlo simulations", Nuclear Engineering and Technology, 53(12), pp.4122~4129 (2021), 한국원자력학회[DOI]
  6. B. Zhang, "An analysis of dependency of counting efficiency on worker anatomy for in vivo measurements: whole-body counting", Phys. Med. Biol, 53(13), pp.3463~3475 (2008)
  7. J. Bento, "Efficiency correction factors of an ACCUSCAN whole-body counter due to the biodistribution of 134Cs, 137Cs and 60℃", Radiat. Protect. Dosim, 155(1), pp.16~24 (2013)
  8. W. E. Bolch, "Guidance on the use of handheld survey meters for radiological triage: timedependent detector count rates corresponding to 50, 250, and 500 mSv effective dose for adult male and females", Health Phys, 102(3), pp.305~325 (2012)
  9. R. Anigstein, "Use of transportable radiation detection instruments to assess internal contamination from intakes of radionuclides Part II: calibration factors and ICAT computer program", Health Phys, 111(6), pp.542~558 (2016)
  10. E. Carinou, "The calculation of a size correction factor for a whole-body counter", Nucl. Instrum. Methods Phys. Res, 580(1), pp.197~200 (2007)
  11. D. Krstic, "Application of MCNP5 software for efficiency calculation of a whole body counter", Health Phys, 102(6), pp.657~663 (2012)
  12. D. Krstic, "Efficiency of whole-body counter for various body size calculated by MCNP5 software", Radiat. Protect. Dosim, 152, pp.179~183 (2012)
  13. M. Park, "Measurement and simulation of the counting efficiency of a whole-body counter using a BOMAB phantom inserted with rod sources containing mixed radionuclides", Health Phys, 114(3), pp.282~287 (2018)
  14. Y. Chen, "Construction of Chinese adult male phantom library and its application in the virtual calibration of in vivo measurement", Phys. Med. Biol, 61(5), pp.2124~2144 (2016)
  15. T. C. F. Fonseca, "Development of a 3D human body library based on polygonal mesh surface for whole body counter set-up calibration", Prog. Nucl. Sci. Technol, 4, pp.614~618 (2014)
  16. S. Nagataki, "Thyroid consequences of the Fukushima nuclear reactor accident", Eur Thyroid J, 1, pp.148~159 (2012)
  17. K. Tani, "Monte-Carlo simulations with mathematical phantoms to investigate the effectiveness of a whole-body counter for thyroid measurement", Radiat. Meas, 135 (2020)
  18. MCNP User's Manual Code Version 6.2, "C.J. Werner", Los Alamos National Laboratory, ~, pp.LA-UR-17-29981 (2017)
  19. F.L. Bronson, "Fastscan e a computerized, anthropometrically designed, high throughput, whole body counter for the nuclear industry", Proceedings of the Seventeenth Midyear Topical Symposium of the Health Physics Society (1984)
  20. F.L. Bronson, "The ACCUSCAN-II vertical scanning germanium whole body counter", Proceedings of the International Congress of the International Radiation Protection Association (IRPA) (1988)
  21. G. H. Kramer, "Monte Carlo simulation of a scanning detector whole body counter and the effect of BOMAB phantom size on the calibration", Health Phys, 83(4), pp.526~533 (2002)
  22. "ICRP 110, Adult reference computational phantoms" (2009), ICRP
  23. "ICRP 143, Paediatric reference computational phantoms" (2020), ICRP
  24. T. Momose, "Whole-body counting of Fukushima residents after the TEPCO Fukushima Daiichi nuclear power station accident", Proceedings of the First NIRS Symposium on the Reconstruction of Early Internal Dose in the TEPCO Fukushima Daiichi Nuclear Power Station Accident, pp.67~82 (2012), National Institute of Radiological Sciences
  25. R. S. Hayano, "BABSCAN : a whole-body counter for small children in Fukushima", J. Radiol. Prot, 34, pp.645~653 (2014)
  26. Dose and Risk Calculation Software, "K.F. Eckerman", ~, pp.ORNL (2006), TM-2001[DOI]
  27. ICRP 119, "Compendium of dose coefficients based on ICRP publication 60, ICRP publication 119", Ann. ICRP, 41(Suppl) (2012)
  28. ICRP 72, "Age-dependent doses to the members of the public from intake of radionuclides. Part 5: compilation of ingestion and inhalation coefficients, ICRP Publication 72", Ann. ICRP, 26(1) (1995)
  29. ICRP 66, "Human respiratory tract model for radiological protection, ICRP publication 66", Ann. ICRP, 24(1-3) (1994)
  30. T. Ishikawa, "Radiation doses and associated risk from the Fukushima nuclear accident: a review of recent publications", Asia Pac. J. Publ. Health, 29(2S), pp.18S~28S (2017)
  31. ICRP 145, "Adult mesh-type reference computational phantoms", ICRP publication, 145 (2020), Ann. ICRP49(3)

관련 논문AI추천